Impact of planning CT slice thickness on the accuracy of automatic target registration using the on-board cone-beam CT

We have evaluated relationship between planning CT slice thickness and the accuracy of automatic target registration using cone-beam CT (CBCT). Planning CT images were acquired with reconstructed slice thickness of 1, 2, 3, 5, and 10mm for three different phantoms: Penta-Guide phantom, acrylic ball...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Igaku butsuri : Nihon Igaku Butsuri Gakkai kikanshi = Japanese journal of medical physics : an official journal of Japan Society of Medical Physics Jg. 31; H. 1; S. 2
Hauptverfasser: Inoue, Hiroyuki, Tanooka, Masao, Doi, Hiroshi, Miura, Hideharu, Nakagawa, Hideo, Sakai, Toshiyuki, Oda, Masahiko, Yasumasa, Katsumi, Sakamoto, Kiyoshi, Kamikonya, Norihiko, Hirota, Shozo
Format: Journal Article
Sprache:Japanisch
Veröffentlicht: Japan 2011
Schlagworte:
ISSN:1345-5354
Online-Zugang:Weitere Angaben
Tags: Tag hinzufügen