Alternating minimization algorithms for transmission tomography using energy detectors

Expectation-minimization algorithms for transmission tomography are extended to account for detectors that are modeled as integrating the energy of the incident X-ray photons. The resulting algorithm is guaranteed to monotonically increase the log likelihood functional. Both the expectation step and...

Celý popis

Uložené v:
Podrobná bibliografia
Vydané v:Conference Record of the Thirty-Sixth Asilomar Conference on Signals, Systems and Computers, 2002 Ročník 1; s. 144 - 147 vol.1
Hlavní autori: O'Sullivan, J.A., Snyder, D.L., Whiting, B.R.
Médium: Konferenčný príspevok..
Jazyk:English
Vydavateľské údaje: IEEE 2002
Predmet:
ISBN:0780375769, 9780780375765
ISSN:1058-6393
On-line prístup:Získať plný text
Tagy: Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
Popis
Shrnutí:Expectation-minimization algorithms for transmission tomography are extended to account for detectors that are modeled as integrating the energy of the incident X-ray photons. The resulting algorithm is guaranteed to monotonically increase the log likelihood functional. Both the expectation step and the maximization step are given in closed form. The formulation is general enough that it can be extended to other energy-dependent models for the detectors.
ISBN:0780375769
9780780375765
ISSN:1058-6393
DOI:10.1109/ACSSC.2002.1197165