Characterization of Compton-scatter imaging with an analytical simulation method

By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and...

Full description

Saved in:
Bibliographic Details
Published in:Physics in medicine & biology Vol. 63; no. 2; p. 025016
Main Authors: Jones, Kevin C, Redler, Gage, Templeton, Alistair, Bernard, Damian, Turian, Julius V, Chu, James C H
Format: Journal Article
Language:English
Published: England 11.01.2018
Subjects:
ISSN:1361-6560, 1361-6560
Online Access:Get more information
Tags: Add Tag
No Tags, Be the first to tag this record!
Abstract By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min treatment delivery) is expected to generate an average 1000 photons per mm at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.
AbstractList By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min-1 treatment delivery) is expected to generate an average 1000 photons per mm2 at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min-1 treatment delivery) is expected to generate an average 1000 photons per mm2 at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.
By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min treatment delivery) is expected to generate an average 1000 photons per mm at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.
Author Bernard, Damian
Chu, James C H
Redler, Gage
Turian, Julius V
Templeton, Alistair
Jones, Kevin C
Author_xml – sequence: 1
  givenname: Kevin C
  surname: Jones
  fullname: Jones, Kevin C
  organization: Department of Radiation Oncology, Rush University Medical Center, Chicago, Illinois 60612, United States of America
– sequence: 2
  givenname: Gage
  surname: Redler
  fullname: Redler, Gage
– sequence: 3
  givenname: Alistair
  surname: Templeton
  fullname: Templeton, Alistair
– sequence: 4
  givenname: Damian
  surname: Bernard
  fullname: Bernard, Damian
– sequence: 5
  givenname: Julius V
  surname: Turian
  fullname: Turian, Julius V
– sequence: 6
  givenname: James C H
  surname: Chu
  fullname: Chu, James C H
BackLink https://www.ncbi.nlm.nih.gov/pubmed/29243663$$D View this record in MEDLINE/PubMed
BookMark eNpNUEtLxDAQDqK4D717kh691M2rSXqU4gsW9KDnMk3S3UibrE2KrL_ewq7i8MEM8z1gZoFOffAWoSuCbwlWakWYILkoBF4BAMX4BM3_Vqf_5hlaxPiBMSGK8nM0oyXlTAg2R6_VFgbQyQ7uG5ILPgttVoV-l4LPo4Y0MZnrYeP8JvtyaZuBnwDdPjkNXRZdP3YHY2_TNpgLdNZCF-3lsS_R-8P9W_WUr18en6u7da65KFKuBZNMFq21ZYNbwmVLZVnwouQclNDYGCM1yIlitqENodYQ0ZjGFFwrVrZ0iW4OubshfI42prp3UduuA2_DGGtSyqlKpegkvT5Kx6a3pt4N00HDvv79Av0Bcg5h_w
CitedBy_id crossref_primary_10_1002_acm2_12615
crossref_primary_10_1088_1361_6560_abd66c
crossref_primary_10_1016_j_radphyschem_2022_110712
crossref_primary_10_1016_j_radphyschem_2025_112611
crossref_primary_10_1109_TRPMS_2019_2943555
crossref_primary_10_1515_jiip_2020_0052
crossref_primary_10_1002_mp_14365
crossref_primary_10_1088_1361_6420_abf0f0
crossref_primary_10_1002_mp_12755
crossref_primary_10_1109_TCI_2020_2999672
ContentType Journal Article
DBID CGR
CUY
CVF
ECM
EIF
NPM
7X8
DOI 10.1088/1361-6560/aaa200
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList MEDLINE - Academic
MEDLINE
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod no_fulltext_linktorsrc
Discipline Medicine
Biology
Physics
EISSN 1361-6560
ExternalDocumentID 29243663
Genre Journal Article
GrantInformation_xml – fundername: NCI NIH HHS
  grantid: R21 CA213407
GroupedDBID ---
-DZ
-~X
123
1JI
4.4
5B3
5RE
5VS
5ZH
7.M
7.Q
AAGCD
AAJIO
AAJKP
AALHV
AATNI
ABCXL
ABHWH
ABJNI
ABLJU
ABQJV
ABVAM
ACAFW
ACGFS
ACHIP
AEFHF
AENEX
AFYNE
AKPSB
ALMA_UNASSIGNED_HOLDINGS
AOAED
ASPBG
ATQHT
AVWKF
AZFZN
CBCFC
CEBXE
CGR
CJUJL
CRLBU
CS3
CUY
CVF
DU5
EBS
ECM
EDWGO
EIF
EJD
EMSAF
EPQRW
EQZZN
F5P
HAK
IHE
IJHAN
IOP
IZVLO
KOT
LAP
M45
N5L
N9A
NPM
NT-
NT.
P2P
PJBAE
R4D
RIN
RNS
RO9
ROL
RPA
SY9
TN5
UCJ
W28
XPP
7X8
ADEQX
AEINN
ID FETCH-LOGICAL-c465t-c637375fee9b0f147f279545944a86c0ddd7ca70f13eb2b12ed16bdbd54c839f2
IEDL.DBID 7X8
ISICitedReferencesCount 15
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000422721900004&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 1361-6560
IngestDate Thu Oct 02 07:54:07 EDT 2025
Thu Jan 02 23:01:10 EST 2025
IsDoiOpenAccess false
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 2
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c465t-c637375fee9b0f147f279545944a86c0ddd7ca70f13eb2b12ed16bdbd54c839f2
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
OpenAccessLink http://doi.org/10.1088/1361-6560/aaa200
PMID 29243663
PQID 1977779882
PQPubID 23479
ParticipantIDs proquest_miscellaneous_1977779882
pubmed_primary_29243663
PublicationCentury 2000
PublicationDate 20180111
PublicationDateYYYYMMDD 2018-01-11
PublicationDate_xml – month: 1
  year: 2018
  text: 20180111
  day: 11
PublicationDecade 2010
PublicationPlace England
PublicationPlace_xml – name: England
PublicationTitle Physics in medicine & biology
PublicationTitleAlternate Phys Med Biol
PublicationYear 2018
SSID ssj0011824
Score 2.3569374
Snippet By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of...
SourceID proquest
pubmed
SourceType Aggregation Database
Index Database
StartPage 025016
SubjectTerms Algorithms
Humans
Lung Neoplasms - diagnostic imaging
Lung Neoplasms - radiotherapy
Monte Carlo Method
Phantoms, Imaging
Photons
Radiographic Image Interpretation, Computer-Assisted
Radiotherapy Planning, Computer-Assisted - methods
Scattering, Radiation
Tomography, X-Ray Computed - methods
Title Characterization of Compton-scatter imaging with an analytical simulation method
URI https://www.ncbi.nlm.nih.gov/pubmed/29243663
https://www.proquest.com/docview/1977779882
Volume 63
WOSCitedRecordID wos000422721900004&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1LS8NAEF7UqnjxUV_1xQpel2azm93kJFIsXlp6UOit7CtQsEk1Vei_dzbZtidBEEIOmwTCZmb22_km8yH0EMtEW8_qwlLDCBcRI9qpjChhFGV5xDQ3tdiEHA7T8TgbhYRbFcoqVzGxDtS2ND5H3qUAVKRvrhU_zj-IV43y7GqQ0NhGLQZQxpd0yfGGRQDs3IjaCkp8k5lAU4JjdddjXaVUHEW_A8x6oekf_fcVj9FhgJj4qbGJE7Tlijbaa0Qnl220Pwh0OgzW9Z-mOkWj3rpxc_NfJi5z7GOFVxiuTN2EE09ntaQR9rlbrAo41PuyzoXjajoLOmC40aQ-Q2_959feCwliC8RwkSyIEUwymeTOZTrKKZd5LDOAVxnnKhUmstZKoyRcYrAZ1zR2lgpttU24AZCVx-dopygLd4lw6pxJwO-FjjOeWp4mmaCaUrhViEzmHXS_mr8JGLNnKFThyq9qspnBDrpoPsJk3nTdmMSwU2SAj67-8PQ1OgBg46v0CKU3qJWDK7tbtGu-F9Pq8662EjgPR4MfgJrGYA
linkProvider ProQuest
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Characterization+of+Compton-scatter+imaging+with+an+analytical+simulation+method&rft.jtitle=Physics+in+medicine+%26+biology&rft.au=Jones%2C+Kevin+C&rft.au=Redler%2C+Gage&rft.au=Templeton%2C+Alistair&rft.au=Bernard%2C+Damian&rft.date=2018-01-11&rft.eissn=1361-6560&rft.volume=63&rft.issue=2&rft.spage=025016&rft_id=info:doi/10.1088%2F1361-6560%2Faaa200&rft_id=info%3Apmid%2F29243663&rft_id=info%3Apmid%2F29243663&rft.externalDocID=29243663
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1361-6560&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1361-6560&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1361-6560&client=summon