Characterization of Compton-scatter imaging with an analytical simulation method
By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and...
Gespeichert in:
| Veröffentlicht in: | Physics in medicine & biology Jg. 63; H. 2; S. 025016 |
|---|---|
| Hauptverfasser: | , , , , , |
| Format: | Journal Article |
| Sprache: | Englisch |
| Veröffentlicht: |
England
11.01.2018
|
| Schlagworte: | |
| ISSN: | 1361-6560, 1361-6560 |
| Online-Zugang: | Weitere Angaben |
| Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
| Abstract | By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min
treatment delivery) is expected to generate an average 1000 photons per mm
at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images. |
|---|---|
| AbstractList | By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min-1 treatment delivery) is expected to generate an average 1000 photons per mm2 at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min-1 treatment delivery) is expected to generate an average 1000 photons per mm2 at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images. By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min treatment delivery) is expected to generate an average 1000 photons per mm at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images. |
| Author | Bernard, Damian Chu, James C H Redler, Gage Turian, Julius V Templeton, Alistair Jones, Kevin C |
| Author_xml | – sequence: 1 givenname: Kevin C surname: Jones fullname: Jones, Kevin C organization: Department of Radiation Oncology, Rush University Medical Center, Chicago, Illinois 60612, United States of America – sequence: 2 givenname: Gage surname: Redler fullname: Redler, Gage – sequence: 3 givenname: Alistair surname: Templeton fullname: Templeton, Alistair – sequence: 4 givenname: Damian surname: Bernard fullname: Bernard, Damian – sequence: 5 givenname: Julius V surname: Turian fullname: Turian, Julius V – sequence: 6 givenname: James C H surname: Chu fullname: Chu, James C H |
| BackLink | https://www.ncbi.nlm.nih.gov/pubmed/29243663$$D View this record in MEDLINE/PubMed |
| BookMark | eNpNUEtLxDAQDqK4D717kh691M2rSXqU4gsW9KDnMk3S3UibrE2KrL_ewq7i8MEM8z1gZoFOffAWoSuCbwlWakWYILkoBF4BAMX4BM3_Vqf_5hlaxPiBMSGK8nM0oyXlTAg2R6_VFgbQyQ7uG5ILPgttVoV-l4LPo4Y0MZnrYeP8JvtyaZuBnwDdPjkNXRZdP3YHY2_TNpgLdNZCF-3lsS_R-8P9W_WUr18en6u7da65KFKuBZNMFq21ZYNbwmVLZVnwouQclNDYGCM1yIlitqENodYQ0ZjGFFwrVrZ0iW4OubshfI42prp3UduuA2_DGGtSyqlKpegkvT5Kx6a3pt4N00HDvv79Av0Bcg5h_w |
| CitedBy_id | crossref_primary_10_1002_acm2_12615 crossref_primary_10_1088_1361_6560_abd66c crossref_primary_10_1016_j_radphyschem_2022_110712 crossref_primary_10_1016_j_radphyschem_2025_112611 crossref_primary_10_1109_TRPMS_2019_2943555 crossref_primary_10_1515_jiip_2020_0052 crossref_primary_10_1002_mp_14365 crossref_primary_10_1088_1361_6420_abf0f0 crossref_primary_10_1002_mp_12755 crossref_primary_10_1109_TCI_2020_2999672 |
| ContentType | Journal Article |
| DBID | CGR CUY CVF ECM EIF NPM 7X8 |
| DOI | 10.1088/1361-6560/aaa200 |
| DatabaseName | Medline MEDLINE MEDLINE (Ovid) MEDLINE MEDLINE PubMed MEDLINE - Academic |
| DatabaseTitle | MEDLINE Medline Complete MEDLINE with Full Text PubMed MEDLINE (Ovid) MEDLINE - Academic |
| DatabaseTitleList | MEDLINE - Academic MEDLINE |
| Database_xml | – sequence: 1 dbid: NPM name: PubMed url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed sourceTypes: Index Database – sequence: 2 dbid: 7X8 name: MEDLINE - Academic url: https://search.proquest.com/medline sourceTypes: Aggregation Database |
| DeliveryMethod | no_fulltext_linktorsrc |
| Discipline | Medicine Biology Physics |
| EISSN | 1361-6560 |
| ExternalDocumentID | 29243663 |
| Genre | Journal Article |
| GrantInformation_xml | – fundername: NCI NIH HHS grantid: R21 CA213407 |
| GroupedDBID | --- -DZ -~X 123 1JI 4.4 5B3 5RE 5VS 5ZH 7.M 7.Q AAGCD AAJIO AAJKP AALHV AATNI ABCXL ABHWH ABJNI ABLJU ABQJV ABVAM ACAFW ACGFS ACHIP AEFHF AENEX AFYNE AKPSB ALMA_UNASSIGNED_HOLDINGS AOAED ASPBG ATQHT AVWKF AZFZN CBCFC CEBXE CGR CJUJL CRLBU CS3 CUY CVF DU5 EBS ECM EDWGO EIF EJD EMSAF EPQRW EQZZN F5P HAK IHE IJHAN IOP IZVLO KOT LAP M45 N5L N9A NPM NT- NT. P2P PJBAE R4D RIN RNS RO9 ROL RPA SY9 TN5 UCJ W28 XPP 7X8 ADEQX AEINN |
| ID | FETCH-LOGICAL-c465t-c637375fee9b0f147f279545944a86c0ddd7ca70f13eb2b12ed16bdbd54c839f2 |
| IEDL.DBID | 7X8 |
| ISICitedReferencesCount | 15 |
| ISICitedReferencesURI | http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000422721900004&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D |
| ISSN | 1361-6560 |
| IngestDate | Thu Oct 02 07:54:07 EDT 2025 Thu Jan 02 23:01:10 EST 2025 |
| IsDoiOpenAccess | false |
| IsOpenAccess | true |
| IsPeerReviewed | true |
| IsScholarly | true |
| Issue | 2 |
| Language | English |
| LinkModel | DirectLink |
| MergedId | FETCHMERGED-LOGICAL-c465t-c637375fee9b0f147f279545944a86c0ddd7ca70f13eb2b12ed16bdbd54c839f2 |
| Notes | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| OpenAccessLink | http://doi.org/10.1088/1361-6560/aaa200 |
| PMID | 29243663 |
| PQID | 1977779882 |
| PQPubID | 23479 |
| ParticipantIDs | proquest_miscellaneous_1977779882 pubmed_primary_29243663 |
| PublicationCentury | 2000 |
| PublicationDate | 20180111 |
| PublicationDateYYYYMMDD | 2018-01-11 |
| PublicationDate_xml | – month: 1 year: 2018 text: 20180111 day: 11 |
| PublicationDecade | 2010 |
| PublicationPlace | England |
| PublicationPlace_xml | – name: England |
| PublicationTitle | Physics in medicine & biology |
| PublicationTitleAlternate | Phys Med Biol |
| PublicationYear | 2018 |
| SSID | ssj0011824 |
| Score | 2.3568473 |
| Snippet | By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of... |
| SourceID | proquest pubmed |
| SourceType | Aggregation Database Index Database |
| StartPage | 025016 |
| SubjectTerms | Algorithms Humans Lung Neoplasms - diagnostic imaging Lung Neoplasms - radiotherapy Monte Carlo Method Phantoms, Imaging Photons Radiographic Image Interpretation, Computer-Assisted Radiotherapy Planning, Computer-Assisted - methods Scattering, Radiation Tomography, X-Ray Computed - methods |
| Title | Characterization of Compton-scatter imaging with an analytical simulation method |
| URI | https://www.ncbi.nlm.nih.gov/pubmed/29243663 https://www.proquest.com/docview/1977779882 |
| Volume | 63 |
| WOSCitedRecordID | wos000422721900004&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D |
| hasFullText | |
| inHoldings | 1 |
| isFullTextHit | |
| isPrint | |
| link | http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1bS8MwFA7qVHzxMu83Ivga1iZp0zyJDIcPbuxBYW8jzQUGrp12Cvv3nrTZ9iQIQulDSqEkJyffOd_p-RC69yiXOg1BDpw1hOc6JUonnFDGjU4t1dLWK_0iBoNsNJLDkHCrQlnl0ifWjtqU2ufIOzEAFeGba9GH2QfxqlGeXQ0SGpuoxQDK-JIuMVqzCICdG1HbNCa-yUygKWFjdVZjHaUUjaLfAWZ90PQO_vuJh2g_QEz82NjEEdqwRRvtNKKTizba7Qc6HQbr-k9dHaNhd9W4ufkvE5cOe1_hFYYrXTfhxJNpLWmEfe4WqwIu9b6oc-G4mkyDDhhuNKlP0Fvv6bX7TILYAtE8TeZEp0wwkThrZR65mAtHhQR4JTlXWaojY4zQSsAjBsF4HlNr4jQ3uUm4BpDl6CnaKsrCniOcMRYrK53TDNBWarPcmchaKp2QylJ7ge6W8zcGY_YMhSps-VWN1zN4gc6aRRjPmq4bYwqRIgN8dPmHt6_QHgAbX6VH4vgatRxsZXuDtvX3fFJ93tZWAvfBsP8DbqTHtA |
| linkProvider | ProQuest |
| openUrl | ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Characterization+of+Compton-scatter+imaging+with+an+analytical+simulation+method&rft.jtitle=Physics+in+medicine+%26+biology&rft.au=Jones%2C+Kevin+C&rft.au=Redler%2C+Gage&rft.au=Templeton%2C+Alistair&rft.au=Bernard%2C+Damian&rft.date=2018-01-11&rft.eissn=1361-6560&rft.volume=63&rft.issue=2&rft.spage=025016&rft_id=info:doi/10.1088%2F1361-6560%2Faaa200&rft_id=info%3Apmid%2F29243663&rft_id=info%3Apmid%2F29243663&rft.externalDocID=29243663 |
| thumbnail_l | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1361-6560&client=summon |
| thumbnail_m | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1361-6560&client=summon |
| thumbnail_s | http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1361-6560&client=summon |