A strategy for OCT estimation of the optic nerve head pigment epithelium central limit‐inner limit of the retina minimal distance, PIMD‐2π
Purpose To develop a semi‐automatic algorithm for estimation of pigment epithelium central limit‐inner limit of the retina minimal distance averaged over 2π radians (PIMD‐2π) and to estimate the precision of the algorithm. Further, the variances in estimates of PIMD‐2π were to be estimated in a pilo...
Uloženo v:
| Vydáno v: | Acta ophthalmologica (Oxford, England) Ročník 97; číslo 2; s. 208 - 213 |
|---|---|
| Hlavní autoři: | , , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
England
Wiley Subscription Services, Inc
01.03.2019
|
| Témata: | |
| ISSN: | 1755-375X, 1755-3768, 1755-3768 |
| On-line přístup: | Získat plný text |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | Purpose
To develop a semi‐automatic algorithm for estimation of pigment epithelium central limit‐inner limit of the retina minimal distance averaged over 2π radians (PIMD‐2π) and to estimate the precision of the algorithm. Further, the variances in estimates of PIMD‐2π were to be estimated in a pilot sample of glaucomatous eyes.
Methods
Three‐dimensional cubes of the optic nerve head (ONH) were captured with a commercial SD‐OCT device. Raw cube data were exported for semi‐automatic segmentation. The inner limit of the retina was automatically detected. Custom software aided the delineation of the ONH pigment epithelium central limit resolved in 500 evenly distributed radii. Sources of variation in PIMD estimates were analysed with an analysis of variance.
Results
The estimated variance for segmentations and angles was 130 μm2 and 1280 μm2, respectively. Considering averaging eight segmentations, a 95 % confidence interval for mean PIMD‐2π was estimated to 212 ± 10 μm (df = 7). The coefficient of variation for segmentation was estimated at 0.05. In the glaucomatous eyes, the within‐subject variance for captured volumes and for segmentations within volumes was 10 μm2 and 50 μm2, respectively.
Conclusion
The developed semi‐automatic algorithm enables estimation of PIMD‐2π in glaucomatous eyes with relevant precision using few segmentations of each captured volume. |
|---|---|
| Bibliografie: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| ISSN: | 1755-375X 1755-3768 1755-3768 |
| DOI: | 10.1111/aos.13908 |