Ultrafast LC-MS/MS analysis of 5-hydroxyindoleacetic acid (5-HIAA) in serum
A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50-100 nmol/L...
Uloženo v:
| Vydáno v: | Scandinavian journal of clinical and laboratory investigation Ročník 81; číslo 5; s. 401 - 405 |
|---|---|
| Hlavní autoři: | , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
England
Taylor & Francis
01.09.2021
|
| Témata: | |
| ISSN: | 0036-5513, 1502-7686, 1502-7686 |
| On-line přístup: | Získat plný text |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50-100 nmol/L) to 99.7% at high levels (500 nmol/L). A previously reported reference interval of 35-123 nmol/L was confirmed as valid based on analysis of 40 samples from voluntary blood donors. |
|---|---|
| Bibliografie: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 0036-5513 1502-7686 1502-7686 |
| DOI: | 10.1080/00365513.2021.1930141 |