Ultrafast LC-MS/MS analysis of 5-hydroxyindoleacetic acid (5-HIAA) in serum
A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50-100 nmol/L...
Gespeichert in:
| Veröffentlicht in: | Scandinavian journal of clinical and laboratory investigation Jg. 81; H. 5; S. 401 - 405 |
|---|---|
| Hauptverfasser: | , |
| Format: | Journal Article |
| Sprache: | Englisch |
| Veröffentlicht: |
England
Taylor & Francis
01.09.2021
|
| Schlagworte: | |
| ISSN: | 0036-5513, 1502-7686, 1502-7686 |
| Online-Zugang: | Volltext |
| Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
| Zusammenfassung: | A very quick and easy LC-MS/MS analysis method for 5-HIAA (5-hydoxyindoleacetic acid) has been developed. The method was fully validated and proved to work well in a clinical setting. Precision at the upper reference limit 123 nmol/L was 3,3% CV. Accuracy ranged from 96% at low levels (50-100 nmol/L) to 99.7% at high levels (500 nmol/L). A previously reported reference interval of 35-123 nmol/L was confirmed as valid based on analysis of 40 samples from voluntary blood donors. |
|---|---|
| Bibliographie: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 0036-5513 1502-7686 1502-7686 |
| DOI: | 10.1080/00365513.2021.1930141 |