Experimental Implementation of a Pulse Compression Technique Using Coherent Plane-Wave Compounding

The axial resolution of an ultrasound imaging system is inversely proportional to the bandwidth of the emitted signal. When conventional pulsing (CP) is used, the impulse response of the transducer and the excitation signal determine together the shape of the emitted pulse and its bandwidth. A way t...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:IEEE transactions on ultrasonics, ferroelectrics, and frequency control Ročník 65; číslo 6; s. 1025 - 1036
Hlavní autoři: Benane, Yanis Mehdi, Bujoreanu, Denis, Lavarello, Roberto J., Varray, Francois, Escoffre, Jean-Michel, Novell, Anthony, Cachard, Christian, Basset, Olivier
Médium: Journal Article
Jazyk:angličtina
Vydáno: United States IEEE 01.06.2018
Institute of Electrical and Electronics Engineers
Témata:
ISSN:0885-3010, 1525-8955, 1525-8955