Automated detection of the phase III slope during inert gas washout testing

We describe a method to determine the phase III slope for the purpose of calculating indexes of ventilation heterogeneity, S(acin) and S(cond), from the multiple breath nitrogen washout test (MBNW). Our automated method applies a recursive, segmented linear regression technique to each breath of the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied physiology (1985) Jg. 112; H. 6; S. 1073
Hauptverfasser: Stuart-Andrews, Christopher R, Kelly, Vanessa J, Sands, Scott A, Lewis, Angela J, Ellis, Matthew J, Thompson, Bruce R
Format: Journal Article
Sprache:Englisch
Veröffentlicht: United States 01.03.2012
Schlagworte:
ISSN:1522-1601, 1522-1601
Online-Zugang:Weitere Angaben
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe a method to determine the phase III slope for the purpose of calculating indexes of ventilation heterogeneity, S(acin) and S(cond), from the multiple breath nitrogen washout test (MBNW). Our automated method applies a recursive, segmented linear regression technique to each breath of the MBNW test and determines the best point of transition, or breakpoint, between each phase of the washout. A sample set of 50 MBNW tests (controls, asthma, and COPD) was used to establish the conditions in which the phase III slope obtained from the automated technique best matched that obtained by two manual interpreters. We then applied our technique to a test set of 30 subjects (with an even number of subjects in each of the above groups) and compared these results against the manual analysis of a third independent manual interpreter. Indexes of ventilation heterogeneity were determined using both methods and compared. The phase III slopes determined by the automatic technique best matched the manual interpreter when the phase III slope was calculated from the phase II-III transition point plus the addition of 50% of the phase II volume to the end of the expiration. Calculation of the indexes S(acin) and S(cond) showed no overall difference between analysis methods in either S(acin) (P = 0.14) or S(cond) (P = 0.59) when the set threshold was applied to our automated analysis. Our analysis method provides an alternate means for rapid quantification of the MBNW test, removing operator dependence without alteration in either S(acin) or S(cond).
Bibliographie:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1522-1601
1522-1601
DOI:10.1152/japplphysiol.00372.2011