Recommendations on the Collection of Comparator Measurement Data in the Performance Evaluation of Continuous Glucose Monitoring Systems
Gespeichert in:
| Titel: | Recommendations on the Collection of Comparator Measurement Data in the Performance Evaluation of Continuous Glucose Monitoring Systems |
|---|---|
| Autoren: | Freckmann, Guido, Pleus, Stefan, Eichenlaub, Manuel, Boija, Elisabet Eriksson, Fokkert, Marion, Hinzmann, Rolf, Jendle, Johan, 1963, Klonoff, David C., Makris, Konstantinos, Nichols, James H., Pemberton, John, Selvin, Elizabeth, Tran, Nam K., Witthauer, Lilian, Slingerland, Robbert J. |
| Weitere Verfasser: | Working Group on Continuous Glucose Monitoring of the IFCC Scientific Division, Contributor |
| Quelle: | EJIFCC. 36(3):225-235 |
| Schlagwörter: | Clinical performance evaluation, comparator data requirements, continuous glucose monitoring, standardization, traceability |
| Beschreibung: | While current systems for continuous glucose monitoring (CGM) are safe and effective, there is a high degree of variability between readings within and across CGM systems. In current CGM performance studies, device readings are compared to glucose concentrations obtained with a comparator ("reference") measurement procedure (usually capillary or venous glucose). However, glucose concentrations from capillary and venous samples can systematically differ, often by as much as 5 to 10%. Different comparator methods have shown biases of up to 8%, and comparator devices of the same brand can systematically differ by more than 5%. To address these issues, the Working Group on CGM of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC WG-CGM) recommends standardizing study procedures and the comparator measurement process in CGM performance studies. The majority of IFCC WG-CGM members recommend the use of capillary samples as reference, mainly because CGM readings will then be aligned better with results from self-monitoring of blood glucose (SMBG). Even with factory-calibrated CGM systems, manufacturers require CGM users to perform SMBG in some situations, e.g., manual calibration, confirmation of extreme readings, discordance between CGM readings and symptoms of hyper- or hypoglycemia, or intermittent signal loss. Comparator devices should meet defined analytical performance specifications for bias and imprecision. Comparator bias can be reduced by retrospective correction of comparator values based on measurements with a method or materials of higher metrological order. Once manufacturers align CGM readings of their systems with comparator results using standardized procedures, variability across CGM systems will be reduced. |
| Dateibeschreibung: | |
| Zugangs-URL: | https://urn.kb.se/resolve?urn=urn:nbn:se:oru:diva-124444 https://doi.org/10.1177/19322968251336221 |
| Datenbank: | SwePub |
| Abstract: | While current systems for continuous glucose monitoring (CGM) are safe and effective, there is a high degree of variability between readings within and across CGM systems. In current CGM performance studies, device readings are compared to glucose concentrations obtained with a comparator ("reference") measurement procedure (usually capillary or venous glucose). However, glucose concentrations from capillary and venous samples can systematically differ, often by as much as 5 to 10%. Different comparator methods have shown biases of up to 8%, and comparator devices of the same brand can systematically differ by more than 5%. To address these issues, the Working Group on CGM of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC WG-CGM) recommends standardizing study procedures and the comparator measurement process in CGM performance studies. The majority of IFCC WG-CGM members recommend the use of capillary samples as reference, mainly because CGM readings will then be aligned better with results from self-monitoring of blood glucose (SMBG). Even with factory-calibrated CGM systems, manufacturers require CGM users to perform SMBG in some situations, e.g., manual calibration, confirmation of extreme readings, discordance between CGM readings and symptoms of hyper- or hypoglycemia, or intermittent signal loss. Comparator devices should meet defined analytical performance specifications for bias and imprecision. Comparator bias can be reduced by retrospective correction of comparator values based on measurements with a method or materials of higher metrological order. Once manufacturers align CGM readings of their systems with comparator results using standardized procedures, variability across CGM systems will be reduced. |
|---|---|
| ISSN: | 16503414 |
| DOI: | 10.1177/19322968251336221 |