Podrobná bibliografie
| Název: |
Optimizing calibration of fluxgate magnetometer offsets using magnetic holes in the earth's magnetosheath. |
| Autoři: |
Gou, Xiaochen, Li, Lei, Zhang, Yiteng, Ye, Shanzhi, Wang, Jindong, Zhou, Bin, Xie, Lianghai, Feng, Yongyong, Jin, Taifeng, Yao, Shutao |
| Zdroj: |
Earth, Planets & Space; 10/16/2025, Vol. 77 Issue 1, p1-12, 12p |
| Témata: |
FLUXGATE magnetometers, CALIBRATION, MAGNETIC anomalies, SPATIAL arrangement |
| Abstrakt: |
The offset of the fluxgate magnetometer (FGM) drifts slowly with time and temperature, requiring regular in-flight calibration. Recent studies utilizing magnetic holes have optimized calibration error to ~ 0.3 nT in the solar wind, but an order larger in the Earth's magnetosheath. Previous magnetic hole calibration is based on the 1D assumption that the magnetic field of the magnetic hole varies only along its axial direction. In this study, we show that the calibration accuracy is primarily affected by the 3D configuration of magnetic holes, background magnetic field, and satellite trajectory related to magnetic holes. To address these challenges, we consider a 3D magnetic bottle-like topology of the magnetic hole and introduce an equal-weight random sampling method and a symmetry recognition algorithm to optimize the FGM calibration error to ~ 0.5 nT in the magnetosheath based on Cluster and magnetospheric multiscale (MMS) mission data, achieving an order-of-magnitude improvement in accuracy. This study offers a novel solution for improving FGM calibration in planetary magnetosheaths, such as the upcoming solar wind–magnetosphere–ionosphere link explorer (SMILE) mission, ensuring more accurate magnetic field observations. [ABSTRACT FROM AUTHOR] |
|
Copyright of Earth, Planets & Space is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
| Databáze: |