Efficient MLC quality assurance using a virtual picket fence test in an MR‐Linac.

Uložené v:
Podrobná bibliografia
Názov: Efficient MLC quality assurance using a virtual picket fence test in an MR‐Linac.
Autori: Yuan, Kai, Cheung, Matthew Manhin, Lee, Louis
Zdroj: Journal of Applied Clinical Medical Physics; Sep2025, Vol. 26 Issue 9, p1-11, 11p
Predmety: QUALITY assurance, RADIOTHERAPY, COLLIMATORS, IMAGE-guided radiation therapy
Abstrakt: Background: The Elekta Unity MR‐Linac system integrates magnetic resonance imaging (MRI) with a linear accelerator (Linac) for adaptive radiation therapy. Traditional quality assurance (QA) methods for multi‐leaf collimators (MLCs) face challenges in this system due to the magnetic field and limited field size of electronic portal imaging devices (EPID). Purpose: This study aims to develop a 'virtual picket fence' test using machine log files to evaluate MLC performance in the Elekta Unity MR‐Linac system, providing a more efficient and comprehensive QA method that overcomes the limitations of traditional approaches. Methods: A picket fence test plan with 11 segments was delivered on the Elekta Unity system. Maximum absolute error and root mean square (RMS) error for each leaf were calculated by comparing log file data with nominal values. A deliberate 1 mm error was introduced in selected MLCs to test the sensitivity of the virtual test. The results from the log file were further compared with measurements from radiochromic films. Results: The maximum deviation between log file data and nominal values was within 1 mm for all leaves. The virtual picket fence test successfully identified MLCs with deviations beyond the 0.5 mm warning threshold in the error‐introduced test. Comparisons with film‐based measurements showed good agreement, with deviations between film and log file data also within 1 mm. Conclusions: The virtual picket fence test provides an efficient and comprehensive method for MLC QA in the Elekta Unity MR‐Linac system. This method can be integrated into weekly QA workflows alongside traditional film‐based methods for thorough quality control. [ABSTRACT FROM AUTHOR]
Copyright of Journal of Applied Clinical Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Databáza: Complementary Index
Popis
Abstrakt:Background: The Elekta Unity MR‐Linac system integrates magnetic resonance imaging (MRI) with a linear accelerator (Linac) for adaptive radiation therapy. Traditional quality assurance (QA) methods for multi‐leaf collimators (MLCs) face challenges in this system due to the magnetic field and limited field size of electronic portal imaging devices (EPID). Purpose: This study aims to develop a 'virtual picket fence' test using machine log files to evaluate MLC performance in the Elekta Unity MR‐Linac system, providing a more efficient and comprehensive QA method that overcomes the limitations of traditional approaches. Methods: A picket fence test plan with 11 segments was delivered on the Elekta Unity system. Maximum absolute error and root mean square (RMS) error for each leaf were calculated by comparing log file data with nominal values. A deliberate 1 mm error was introduced in selected MLCs to test the sensitivity of the virtual test. The results from the log file were further compared with measurements from radiochromic films. Results: The maximum deviation between log file data and nominal values was within 1 mm for all leaves. The virtual picket fence test successfully identified MLCs with deviations beyond the 0.5 mm warning threshold in the error‐introduced test. Comparisons with film‐based measurements showed good agreement, with deviations between film and log file data also within 1 mm. Conclusions: The virtual picket fence test provides an efficient and comprehensive method for MLC QA in the Elekta Unity MR‐Linac system. This method can be integrated into weekly QA workflows alongside traditional film‐based methods for thorough quality control. [ABSTRACT FROM AUTHOR]
ISSN:15269914
DOI:10.1002/acm2.70232