The effects of magnetic field distortion on the accuracy of passive device localization frames in MR imaging

Purpose: The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensit...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Medical physics (Lancaster) Ročník 41; číslo 5; s. 052301 - n/a
Hlavní autoři: Cepek, Jeremy, Chronik, Blaine A., Fenster, Aaron
Médium: Journal Article
Jazyk:angličtina
Vydáno: United States American Association of Physicists in Medicine 01.05.2014
Témata:
ISSN:0094-2405, 2473-4209, 2473-4209
On-line přístup:Získat plný text
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Abstract Purpose: The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR. Methods: Three localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scanner's center frequency and gradient shim values. Results: Plots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1–100μT and static field distortion gradients of 0.01–1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging. Conclusions: The error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frame's design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice-select error (such as the z-frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in-plane measurements of marker positions, (b) the in-plane marker positions in images are not sensitive to slice-select error, and (c) methods of correcting in-plane error in the frequency-encoded direction are employed.
AbstractList Purpose: The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR. Methods: Three localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scannerˈs center frequency and gradient shim values. Results: Plots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1–100μT and static field distortion gradients of 0.01–1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging. Conclusions: The error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frameˈs design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice‐select error (such as the z‐frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in‐plane measurements of marker positions, (b) the in‐plane marker positions in images are not sensitive to slice‐select error, and (c) methods of correcting in‐plane error in the frequency‐encoded direction are employed.
The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR.PURPOSEThe interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR.Three localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scanner's center frequency and gradient shim values.METHODSThree localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scanner's center frequency and gradient shim values.Plots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1-100 μT and static field distortion gradients of 0.01-1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging.RESULTSPlots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1-100 μT and static field distortion gradients of 0.01-1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging.The error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frame's design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice-select error (such as the z-frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in-plane measurements of marker positions, (b) the in-plane marker positions in images are not sensitive to slice-select error, and (c) methods of correcting in-plane error in the frequency-encoded direction are employed.CONCLUSIONSThe error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frame's design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice-select error (such as the z-frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in-plane measurements of marker positions, (b) the in-plane marker positions in images are not sensitive to slice-select error, and (c) methods of correcting in-plane error in the frequency-encoded direction are employed.
The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR. Three localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scanner's center frequency and gradient shim values. Plots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1-100 μT and static field distortion gradients of 0.01-1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging. The error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frame's design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice-select error (such as the z-frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in-plane measurements of marker positions, (b) the in-plane marker positions in images are not sensitive to slice-select error, and (c) methods of correcting in-plane error in the frequency-encoded direction are employed.
Purpose: The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular, geometric distortion of the static magnetic field may be both appreciable and unpredictable. This paper aims to quantify the sensitivity of localization error of various passive device localization frames to static magnetic field distortion in MR. Methods: Three localization frames were considered based on having distinctly different methods of encoding position and orientation in MR images. For each frame, the effects of static field distortion were modeled, allowing rotational and translational errors to be computed as functions of the level of distortion, which was modeled using a first order approximation. Validation of the model was performed by imaging the localization frames in a 3T clinical MR scanner, and simulating the effects of static field distortion by varying the scanner's center frequency and gradient shim values. Results: Plots of the rotational and translational components of error in localization frame position and orientation estimates are provided for ranges of uniform static field distortions of 1–100μT and static field distortion gradients of 0.01–1 mT/m in all three directions. The theoretical estimates are in good agreement with the results obtained by imaging. Conclusions: The error in position and orientation estimation of passive localization frames in MR can be sensitive to static magnetic field distortions. The level of sensitivity, the type of error (i.e., rotational or translational), and the direction of error are dependent on the frame's design and the method used to image it. If 2D gradient echo imaging is employed, frames with position and orientation estimate sensitivity to slice-select error (such as the z-frame) should be avoided, since this source of error is not easily correctable. Accurate frame position and orientation estimates that are insensitive to static field distortion can be achieved using 2D gradient echo imaging if: (a) the method of determining position and orientation only uses in-plane measurements of marker positions, (b) the in-plane marker positions in images are not sensitive to slice-select error, and (c) methods of correcting in-plane error in the frequency-encoded direction are employed.
Author Fenster, Aaron
Chronik, Blaine A.
Cepek, Jeremy
Author_xml – sequence: 1
  givenname: Jeremy
  surname: Cepek
  fullname: Cepek, Jeremy
  email: jcepek@robarts.ca
  organization: Robarts Research Institute, The University of Western Ontario, London, Ontario N6A 5B7, Canada and Biomedical Engineering, The University of Western Ontario, London, Ontario N6A 5B9, Canada
– sequence: 2
  givenname: Blaine A.
  surname: Chronik
  fullname: Chronik, Blaine A.
  organization: Robarts Research Institute, The University of Western Ontario, London, Ontario N6A 5B7, Canada and Department of Physics and Astronomy, The University of Western Ontario, London, Ontario N6A 3K7, Canada
– sequence: 3
  givenname: Aaron
  surname: Fenster
  fullname: Fenster, Aaron
  organization: Robarts Research Institute, The University of Western Ontario, London, Ontario N6A 5B7, Canada and Biomedical Engineering, The University of Western Ontario, London, Ontario N6A 5B9, Canada
BackLink https://www.ncbi.nlm.nih.gov/pubmed/24784394$$D View this record in MEDLINE/PubMed
BookMark eNp90FtLwzAUB_AgE92mD34ByaMInTlN0sujDG8wUWQ-lzQ5mZGurU03mZ_e7uIQBCFwHvI7F_4D0iurEgk5AzYCgOQKRiKJWRrBAemHIuaBCFnaI33GUhGEgsljMvD-nTEWccmOyHGHEsFT0SfF9A0pWou69bSydK5mJbZOU-uwMNQ431ZN66qSdq_trNJ60Si9WuNaee-WSA0unUZaVFoV7kttuG3UHD11JX18oa4b68rZCTm0qvB4uqtD8np7Mx3fB5Onu4fx9SSoOaQQ8DyXWkfGotFS56EwPBexNjyRuY2MYAYiZAIhlmmqrMx5mDLQcZKEuRUQ8SG52M6tm-pjgb7N5s5rLApVYrXwGcgQuGAgZUfPd3SRz9FkddPd2qyyn4Q6EGzBpytwtf8Hlq2jzyDbRZ89Pq9L5y-33mvXbqLY9yyr5pevjf0P_1nAvwHxGZNJ
CODEN MPHYA6
ContentType Journal Article
Copyright American Association of Physicists in Medicine
2014 American Association of Physicists in Medicine
Copyright_xml – notice: American Association of Physicists in Medicine
– notice: 2014 American Association of Physicists in Medicine
DBID CGR
CUY
CVF
ECM
EIF
NPM
7X8
DOI 10.1118/1.4870961
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList
MEDLINE - Academic
MEDLINE

Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
Physics
EISSN 2473-4209
EndPage n/a
ExternalDocumentID 24784394
MP0961
Genre article
Validation Studies
Research Support, Non-U.S. Gov't
Journal Article
GrantInformation_xml – fundername: Canadian Institutes of Health Research
GroupedDBID ---
--Z
-DZ
.GJ
0R~
1OB
1OC
29M
2WC
33P
36B
3O-
4.4
476
53G
5GY
5RE
5VS
AAHHS
AANLZ
AAQQT
AASGY
AAXRX
AAZKR
ABCUV
ABEFU
ABFTF
ABJNI
ABLJU
ABQWH
ABTAH
ABXGK
ACAHQ
ACBEA
ACCFJ
ACCZN
ACGFO
ACGFS
ACGOF
ACPOU
ACSMX
ACXBN
ACXQS
ADBBV
ADBTR
ADKYN
ADOZA
ADXAS
ADZMN
AEEZP
AEGXH
AEIGN
AENEX
AEQDE
AEUYR
AFBPY
AFFPM
AHBTC
AIACR
AIAGR
AIURR
AIWBW
AJBDE
ALMA_UNASSIGNED_HOLDINGS
ALUQN
AMYDB
ASPBG
BFHJK
C45
CS3
DCZOG
DRFUL
DRMAN
DRSTM
DU5
EBD
EBS
EJD
EMB
EMOBN
F5P
G8K
HDBZQ
HGLYW
I-F
KBYEO
LATKE
LEEKS
LOXES
LUTES
LYRES
MEWTI
O9-
OVD
P2P
P2W
PALCI
PHY
RJQFR
RNS
ROL
SAMSI
SUPJJ
SV3
TEORI
TN5
TWZ
USG
WOHZO
WXSBR
XJT
ZGI
ZVN
ZXP
ZY4
ZZTAW
AAHQN
AAIPD
AAMNL
AAYCA
ABDPE
AFWVQ
AITYG
ALVPJ
ADMLS
CGR
CUY
CVF
ECM
EIF
NPM
7X8
AAMMB
ABUFD
AEFGJ
AEYWJ
AGXDD
AGYGG
AIDQK
AIDYY
LH4
ID FETCH-LOGICAL-p3191-3bb5cc6dfedc5cb24d3b47cd385bf6d40d16e04e17599af5b32901c7882bf4163
IEDL.DBID DRFUL
ISICitedReferencesCount 2
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000336053100031&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 0094-2405
2473-4209
IngestDate Sun Nov 09 11:58:18 EST 2025
Wed Feb 19 01:57:00 EST 2025
Wed Jan 22 16:48:22 EST 2025
Fri Jun 21 00:28:34 EDT 2024
Sun Jul 14 10:05:21 EDT 2019
IsPeerReviewed true
IsScholarly true
Issue 5
Keywords MRI-guided
localization frame
stereotactic
Language English
License 0094-2405/2014/41(5)/052301/12/$30.00
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-p3191-3bb5cc6dfedc5cb24d3b47cd385bf6d40d16e04e17599af5b32901c7882bf4163
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ObjectType-Undefined-3
PMID 24784394
PQID 1521340155
PQPubID 23479
PageCount 12
ParticipantIDs wiley_primary_10_1118_1_4870961_MP0961
proquest_miscellaneous_1521340155
scitation_primary_10_1118_1_4870961
pubmed_primary_24784394
PublicationCentury 2000
PublicationDate May 2014
PublicationDateYYYYMMDD 2014-05-01
PublicationDate_xml – month: 05
  year: 2014
  text: May 2014
PublicationDecade 2010
PublicationPlace United States
PublicationPlace_xml – name: United States
PublicationTitle Medical physics (Lancaster)
PublicationTitleAlternate Med Phys
PublicationYear 2014
Publisher American Association of Physicists in Medicine
Publisher_xml – name: American Association of Physicists in Medicine
References Goldenberg, Trachtenberg, Yi, Weersink, Sussman, Haider, Ma, Kucharczyk (c4) 2010; 28
Cepek, Chronik, Lindner, Trachtenberg, Davidson, Bax, Fenster (c10) 2013; 40
Bricault, Zemiti, Jouniaux, Fouard, Taillant, Dorandeu, Cinquin (c24) 2008; 27
Chang, Fitzpatrick (c32) 1992; 11
Schouten, Ansems, Renema, Bosboom, Scheenen, Futterer (c7) 2010; 37
Fiedler, Schwarzmaier, Eickmeyer, Müller, Schoepp, Verreet (c12) 2001; 13
Sutherland, Lama, Gan, Wolfsberger, Zareinia (c17) 2013; 118
Anders, SanGiorgio, Deligianni, Santini, Scheffler, Boero (c21) 2012; 67
de Oliveira, Rauschenberg, Beyersdorff, Semmler, Bock (c22) 2008; 59
Seifabadi, Song, Krieger, Cho, Tokuda, Fichtinger, Iordachita (c28) 2012; 7
Schenck (c36) 1996; 23
Krieger, Iordachita, Guion, Singh, Kaushal, Ménard, Pinto, Camphausen, Fichtinger, Whitcomb (c26) 2011; 58
Ménard, Susil, Choyke, Gustafson, Kammerer, Ning, Miller, Ullman, Sears Crouse, Smith (c6) 2004; 59
Brown (c33) 1979; 14
Song, Tokuda, Tuncali, Tempany, Zhang, Hata (c8) 2013; 60
Dumoulin, Souza, Darrow (c20) 1993; 29
Krieger, Susil, Ménard, Coleman, Fichtinger, Atalar, Whitcomb (c25) 2005; 52
Mouraviev, Verma, Kalyanaraman, Zhai, Gaitonde, Pugnale, Donovan (c18) 2013; 27
Melzer, Gutmann, Remmele, Wolf, Lukoscheck, Bock, Bardenheuer, Fischer (c19) 2008; 27
Bourgeois, Magnin, Morel, Sartoretti, Huisman, Tuncdogan, Meier, Jeanmonod (c14) 1999; 41
Lee, Fichtinger, Chirikjian (c34) 2002; 29
Muntener, Patriciu, Petrisor, Mazilu, Bagga, Kavoussi, Cleary, Stoianovici (c27) 2006; 68
Vogl, Straub, Eichler, Woitaschek, Mack (c13) 2002; 225
Goldenberg, Trachtenberg, Kucharczyk, Yi, Haider, Ma, Weersink, Raoufi (c30) 2008; 13
Hall, Galicich, Bergman, Truwit (c16) 2006; 77
Dogan, Le-Petross, Stafford, Atkinson, Whitman (c11) 2012; 199
2010; 37
1993; 29
2013; 27
1979; 14
2012
2011
2010
2006; 77
2013; 40
2009
2008; 59
2007
2006
2008; 13
1999; 41
2011; 58
1992; 11
1999
2012; 199
2002; 29
2000
2006; 68
2010; 28
2004; 59
2013; 118
2008; 27
2005; 52
2013; 60
2002; 225
2013
2012; 7
2001; 13
2012; 67
1996; 23
References_xml – volume: 28
  start-page: 215
  year: 2010
  ident: c4
  article-title: Robot-assisted MRI-guided prostatic interventions
  publication-title: Robotica
– volume: 23
  start-page: 815
  year: 1996
  ident: c36
  article-title: The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds
  publication-title: Med. Phys.
– volume: 58
  start-page: 3049
  year: 2011
  ident: c26
  article-title: An MRI-compatible robotic system with hybrid tracking for MRI-guided prostate intervention
  publication-title: IEEE Trans. Biomed. Eng.
– volume: 225
  start-page: 367
  year: 2002
  ident: c13
  article-title: Malignant liver tumors treated with MR imaging–guided laser-induced thermotherapy: Experience with complications in 899 patients (2,520 lesions)
  publication-title: Radiology
– volume: 52
  start-page: 306
  year: 2005
  ident: c25
  article-title: Design of a novel MRI compatible manipulator for image guided prostate interventions
  publication-title: IEEE Trans. Biomed. Eng.
– volume: 199
  start-page: W651
  year: 2012
  ident: c11
  article-title: MRI-guided vacuum-assisted breast biopsy performed at 3 T with a 9-gauge needle: Preliminary experience
  publication-title: Am. J. Roentgenol.
– volume: 29
  start-page: 1881
  year: 2002
  ident: c34
  article-title: Numerical algorithms for spatial registration of line fiducials from cross-sectional images
  publication-title: Med. Phys.
– volume: 59
  start-page: 1414
  year: 2004
  ident: c6
  article-title: MRI-guided HDR prostate brachytherapy in standard 1.5 T scanner
  publication-title: Int. J. Radiat. Oncol.
– volume: 118
  start-page: 521
  year: 2013
  ident: c17
  article-title: Merging machines with microsurgery: Clinical experience with neuroArm: Clinical article
  publication-title: J. Neurosurg.
– volume: 13
  start-page: 374
  year: 2008
  ident: c30
  article-title: Robotic system for closed-bore MRI-guided prostatic interventions
  publication-title: IEEE/ASME Trans. Mechatronics
– volume: 40
  start-page: 012304
  year: 2013
  ident: c10
  article-title: A system for MRI-guided transperineal delivery of needles to the prostate for focal therapy
  publication-title: Med. Phys.
– volume: 11
  start-page: 319
  year: 1992
  ident: c32
  article-title: A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities
  publication-title: IEEE Trans. Med. Imaging
– volume: 77
  start-page: 297
  year: 2006
  ident: c16
  article-title: 3-Tesla intraoperative MR imaging for neurosurgery
  publication-title: J. Neuro-Oncol.
– volume: 14
  start-page: 300
  year: 1979
  ident: c33
  article-title: A stereotactic head frame for use with CT body scanners
  publication-title: Invest. Radiol.
– volume: 13
  start-page: 729
  year: 2001
  ident: c12
  article-title: Laser-induced interstitial thermotherapy of liver metastases in an interventional 0.5 Tesla MRI system: Technique and first clinical experiences
  publication-title: J. Magn. Reson. Imaging
– volume: 60
  start-page: 3019
  year: 2013
  ident: c8
  article-title: Development and preliminary evaluation of a motorized needle guide template for MRI-guided targeted prostate biopsy
  publication-title: IEEE Trans. Biomed. Eng.
– volume: 68
  start-page: 1313
  year: 2006
  ident: c27
  article-title: Magnetic resonance imaging compatible robotic system for fully automated brachytherapy seed placement
  publication-title: Urology
– volume: 67
  start-page: 290
  year: 2012
  ident: c21
  article-title: Integrated active tracking detector for MRI-guided interventions
  publication-title: Magn. Reson. Med.
– volume: 59
  start-page: 1043
  year: 2008
  ident: c22
  article-title: Automatic passive tracking of an endorectal prostate biopsy device using phase-only cross-correlation
  publication-title: Magn. Reson. Med.
– volume: 29
  start-page: 411
  year: 1993
  ident: c20
  article-title: Real-time position monitoring of invasive devices using magnetic resonance
  publication-title: Magn. Reson. Med.
– volume: 27
  start-page: 42
  year: 2008
  ident: c24
  article-title: Light puncture robot for CT and MRI interventions
  publication-title: IEEE Eng. Med. Biol.
– volume: 37
  start-page: 4744
  year: 2010
  ident: c7
  article-title: The accuracy and safety aspects of a novel robotic needle guide manipulator to perform transrectal prostate biopsies
  publication-title: Med. Phys.
– volume: 41
  start-page: 636
  year: 1999
  ident: c14
  article-title: Accuracy of MRI-guided stereotactic thalamic functional neurosurgery
  publication-title: Neuroradiology
– volume: 7
  start-page: 181
  year: 2012
  ident: c28
  article-title: Robotic system for MRI-guided prostate biopsy: Feasibility of teleoperated needle insertion and ex vivo phantom study
  publication-title: Int. J. Comput. Assist. Radiol. Surg.
– volume: 27
  start-page: 820
  year: 2013
  ident: c18
  article-title: The feasibility of multiparametric magnetic resonance imaging for targeted biopsy using novel navigation systems to detect early stage prostate cancer: The preliminary experience
  publication-title: J. Endourol.
– volume: 27
  start-page: 66
  year: 2008
  ident: c19
  article-title: Innomotion for percutaneous image-guided interventions
  publication-title: IEEE Eng. Med. Biol.
– volume: 13
  start-page: 374
  year: 2008
  end-page: 379
  article-title: Robotic system for closed‐bore MRI‐guided prostatic interventions
  publication-title: IEEE/ASME Trans. Mechatronics
– volume: 67
  start-page: 290
  year: 2012
  end-page: 296
  article-title: Integrated active tracking detector for MRI‐guided interventions
  publication-title: Magn. Reson. Med.
– volume: 52
  start-page: 306
  year: 2005
  end-page: 313
  article-title: Design of a novel MRI compatible manipulator for image guided prostate interventions
  publication-title: IEEE Trans. Biomed. Eng.
– volume: 28
  start-page: 215
  year: 2010
  end-page: 234
  article-title: Robot‐assisted MRI‐guided prostatic interventions
  publication-title: Robotica
– volume: 29
  start-page: 1881
  year: 2002
  end-page: 1891
  article-title: Numerical algorithms for spatial registration of line fiducials from cross‐sectional images
  publication-title: Med. Phys.
– volume: 13
  start-page: 729
  year: 2001
  end-page: 737
  article-title: Laser‐induced interstitial thermotherapy of liver metastases in an interventional 0.5 Tesla MRI system: Technique and first clinical experiences
  publication-title: J. Magn. Reson. Imaging
– volume: 41
  start-page: 636
  year: 1999
  end-page: 645
  article-title: Accuracy of MRI‐guided stereotactic thalamic functional neurosurgery
  publication-title: Neuroradiology
– volume: 58
  start-page: 3049
  year: 2011
  end-page: 3060
  article-title: An MRI‐compatible robotic system with hybrid tracking for MRI‐guided prostate intervention
  publication-title: IEEE Trans. Biomed. Eng.
– volume: 77
  start-page: 297
  year: 2006
  end-page: 303
  article-title: 3‐Tesla intraoperative MR imaging for neurosurgery
  publication-title: J. Neuro‐Oncol.
– volume: 59
  start-page: 1414
  year: 2004
  end-page: 1423
  article-title: MRI‐guided HDR prostate brachytherapy in standard 1.5 T scanner
  publication-title: Int. J. Radiat. Oncol.
– start-page: 477
  year: 2011
  end-page: 501
  article-title: MR‐guided interventions and surgery
– start-page: 50
  year: 2007
  end-page: 58
  article-title: Dynamic MRI scan plane control for passive tracking of instruments and devices
– volume: 27
  start-page: 820
  year: 2013
  end-page: 825
  article-title: The feasibility of multiparametric magnetic resonance imaging for targeted biopsy using novel navigation systems to detect early stage prostate cancer: The preliminary experience
  publication-title: J. Endourol.
– volume: 27
  start-page: 42
  year: 2008
  end-page: 50
  article-title: Light puncture robot for CT and MRI interventions
  publication-title: IEEE Eng. Med. Biol.
– start-page: 1939
  year: 2012
  end-page: 1945
  article-title: A MRI‐guided concentric tube continuum robot with piezoelectric actuation: A feasibility study
  publication-title: Proceedings of IEEE International Conference on Robotics and Automation
– volume: 14
  start-page: 300
  year: 1979
  end-page: 304
  article-title: A stereotactic head frame for use with CT body scanners
  publication-title: Invest. Radiol.
– start-page: 1066
  year: 2010
  end-page: 1073
  article-title: Development and preliminary evaluation of an actuated MRI‐compatible robotic device for MRI‐guided prostate intervention
– start-page: 348
  year: 2013
  end-page: 355
  article-title: Errors in device localization in MRI using Z‐Frames
– volume: 225
  start-page: 367
  year: 2002
  end-page: 377
  article-title: Malignant liver tumors treated with MR imaging–guided laser‐induced thermotherapy: Experience with complications in 899 patients (2,520 lesions)
  publication-title: Radiology
– volume: 68
  start-page: 1313
  year: 2006
  end-page: 1317
  article-title: Magnetic resonance imaging compatible robotic system for fully automated brachytherapy seed placement
  publication-title: Urology
– volume: 199
  start-page: W651
  year: 2012
  end-page: W653
  article-title: MRI‐guided vacuum‐assisted breast biopsy performed at 3 T with a 9‐gauge needle: Preliminary experience
  publication-title: Am. J. Roentgenol.
– volume: 40
  start-page: 012304
  year: 2013
  article-title: A system for MRI‐guided transperineal delivery of needles to the prostate for focal therapy
  publication-title: Med. Phys.
– start-page: 68
  year: 2006
  end-page: 73
  article-title: A system for MRI‐guided prostate interventions
  publication-title: Proceedings of IEEE Conference on Biomedical Robotics and Biomechatronics
– volume: 60
  start-page: 3019
  year: 2013
  end-page: 3027
  article-title: Development and preliminary evaluation of a motorized needle guide template for MRI‐guided targeted prostate biopsy
  publication-title: IEEE Trans. Biomed. Eng.
– start-page: 425
  year: 2007
  end-page: 433
  article-title: Robotic assistant for transperineal prostate interventions in 3T closed MRI
– start-page: 921
  year: 2000
  end-page: 930
  article-title: MR compatible surgical assist robot: System integration and preliminary feasibility study
– volume: 29
  start-page: 411
  year: 1993
  end-page: 415
  article-title: Real‐time position monitoring of invasive devices using magnetic resonance
  publication-title: Magn. Reson. Med.
– volume: 27
  start-page: 66
  year: 2008
  end-page: 73
  article-title: Innomotion for percutaneous image‐guided interventions
  publication-title: IEEE Eng. Med. Biol.
– volume: 59
  start-page: 1043
  year: 2008
  end-page: 1050
  article-title: Automatic passive tracking of an endorectal prostate biopsy device using phase‐only cross‐correlation
  publication-title: Magn. Reson. Med.
– volume: 23
  start-page: 815
  year: 1996
  end-page: 850
  article-title: The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds
  publication-title: Med. Phys.
– volume: 37
  start-page: 4744
  year: 2010
  end-page: 4750
  article-title: The accuracy and safety aspects of a novel robotic needle guide manipulator to perform transrectal prostate biopsies
  publication-title: Med. Phys.
– volume: 118
  start-page: 521
  year: 2013
  end-page: 529
  article-title: Merging machines with microsurgery: Clinical experience with neuroArm: Clinical article
  publication-title: J. Neurosurg.
– volume: 11
  start-page: 319
  year: 1992
  end-page: 329
  article-title: A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities
  publication-title: IEEE Trans. Med. Imaging
– volume: 7
  start-page: 181
  year: 2012
  end-page: 190
  article-title: Robotic system for MRI‐guided prostate biopsy: Feasibility of teleoperated needle insertion and ex vivo phantom study
  publication-title: Int. J. Comput. Assist. Radiol. Surg.
– start-page: 4450
  year: 2009
  end-page: 4456
  article-title: Design of a robotic system for mri‐guided deep brain stimulation electrode placement
– year: 1999
SSID ssj0006350
Score 2.1018367
Snippet Purpose: The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In...
The interventional magnetic resonance (MR) imaging environment presents many challenges for the accurate localization of interventional devices. In particular,...
SourceID proquest
pubmed
wiley
scitation
SourceType Aggregation Database
Index Database
Publisher
Enrichment Source
StartPage 052301
SubjectTerms Algorithms
Biological material, e.g. blood, urine; Haemocytometers
biomedical MRI
Calibration
Clinical applications
Digital computing or data processing equipment or methods, specially adapted for specific applications
Encoding
estimation theory
Image analysis
image coding
Image coding, e.g. from bit‐mapped to non bit‐mapped
Image data processing or generation, in general
Image scanners
Instrumentation
Involving electronic [emr] or nuclear [nmr] magnetic resonance, e.g. magnetic resonance imaging
Localization effects
localization frame
Magnetic Fields
Magnetic resonance
Magnetic Resonance Imaging - methods
Magnetoresistance
medical image processing
Medical imaging
Medical magnetic resonance imaging
Models, Theoretical
MRI‐guided
Rotation
sensitivity
Static magnetic fields
stereotactic
Title The effects of magnetic field distortion on the accuracy of passive device localization frames in MR imaging
URI http://dx.doi.org/10.1118/1.4870961
https://onlinelibrary.wiley.com/doi/abs/10.1118%2F1.4870961
https://www.ncbi.nlm.nih.gov/pubmed/24784394
https://www.proquest.com/docview/1521340155
Volume 41
WOSCitedRecordID wos000336053100031&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVWIB
  databaseName: Wiley Online Library Full Collection 2020
  customDbUrl:
  eissn: 2473-4209
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0006350
  issn: 0094-2405
  databaseCode: DRFUL
  dateStart: 19970101
  isFulltext: true
  titleUrlDefault: https://onlinelibrary.wiley.com
  providerName: Wiley-Blackwell
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1bS9xAFD7oai8-qNVW1xsjLX1L3SSTyYQ-ldbFB1dEFPYtzLUs1Oyy6wr99z1nMqYULBQKIXnImTBwbt_k3AA-VE4J7q1Iyky7hFdOJnqgRaIzaQXeJC98GDZRXl3J8bi6XoHPT7UwbX-I7ocbaUaw16TgSscpJCklrqefEGvTwJJVWMtQbnkP1r7dDO8uO0OMvrStQKk4BRGK2FgIl591i58DlhvwCt1PGwn_E7MGpzPc-q_tbsNmxJrsSyscb2DFNTvwchSj6TvwIqR_msUu_EBpYTG1g009u1ffG6puZCHBjdnQS4Q2yvBCyMiUMcu5Mj-JeIb4G20ms46sDgveMVZ3Mk-5Xws2adjohk3uw0ikt3A3PL_9epHEOQzJDBU0TXKtC2OE9c6awuiM21zz0thcFtoLywc2FW7AHSKRqlK-0DkFZw0erjPtCfC9g14zbdw-sFyXwuQGBaQYcKO0rJRETIWYLveirEwfTp_YUaOcU_BCNW66XNSEM3JOCK8Pey2f6lnbkKPOeCmpwrcP7zvGdS_bU46s0zoy4Fmqx-n8N0U9s74PHwM7__6denRNj4N_JTyE14i2eJsteQS9h_nSHcO6eXyYLOYnsFqO5UkU4V9mT_BY
linkProvider Wiley-Blackwell
linkToHtml http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV3dS9xAEB_0bNU-2GqrXqt2i-Jb2rtks0mgL6XtoXh3iCj4tmS_5EBzx50n-N87s1lTBIWCEJKHzIYl87G_3fkCOChsKbgzIspiZSNe2DxSHSUiFedG4C3nqfPNJrLhML-8LE4X4OdjLkxdH6I5cCPN8PaaFJwOpIOWU-R69zuCbepYsghLHMUobcHSn7PeRb-xxLiY1ikoBScvQhoqC-HwH83g55DlO1jB9ad2hT8FrX7V6b1_3Xw_wFpAm-xXLR7rsGCrDVgeBH_6Brz1AaB69hGuUV5YCO5gY8duyquK8huZD3FjxlcToZkyvBA0slLr-bTU90Q8QQSOVpMZS3aH-fUx5HcyR9FfMzaq2OCMjW58U6RPcNH7e_77KAqdGKIJqmg3SpRKtRbGWaNTrWJuEsUzbfCnKycM75iusB1uEYsURelSlZB7VuP2OlaOIN8mtKpxZbeBJSoTOtEoImmH61LlRZkjqkJUlziRFboN3x75IVHSyX1RVnY8n0lCGgknjNeGrZpRclKX5JAxz3LK8W3DfsO55mW9z8llVwYGPEt1N57-o5AT49pw6Pn58nfk4JQen_-X8CusHJ0P-rJ_PDz5AquIvXgdO7kDrdvp3O7CG313O5pN94IkPwAn2PNg
linkToPdf http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1Za9tAEB5Sp03bhx5J2rrpsaEhb2psabVaQV9KE9OQ2JiQQN4W7VUMiWzs2NB_35nVWqWQQqEgpAfNioW5vtVcAAelqwT3ViRFql3CSycT3dMi0am0Am-S5z4MmyhGI3l9XY434Mu6FqbpD9H-cCPNCPaaFNzNrI9aTpnr_c8ItmliyQPY5Dk-O7B5fDG4Om8tMTrTpgSl5BRFyGNnIVx-1C6-D1k-hcfof5pQ-J-gNXidwfP_2-8LeBbRJvvaiMdL2HD1NmwNYzx9Gx6FBFCz2IEblBcWkzvY1LPb6kdN9Y0spLgxG7qJ0E4ZXggaWWXMcl6Zn0Q8QwSOVpNZR3aHBf8Y6zuZp-yvBZvUbHjBJrdhKNIuXA1OLr99T-IkhmSGKtpPMq1zY4T1zprc6JTbTPPC2Ezm2gvLe7YvXI87xCJlWflcZxSeNXi8TrUnyPcKOvW0dm-AZboQJjMoInmPm0rLspKIqhDVZV4UpenC_pofCiWdwhdV7abLhSKkkXHCeF143TBKzZqWHCrlhaQa3y58ajnXvmzOOVL1VWTAvVSr6fw3hULWdeEw8PPv31HDMT3e_ivhR9gaHw_U-enobA-eIPTiTerkO-jczZfuPTw0q7vJYv4hCvIvxKHy2w
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=The+effects+of+magnetic+field+distortion+on+the+accuracy+of+passive+device+localization+frames+in+MR+imaging&rft.jtitle=Medical+physics+%28Lancaster%29&rft.au=Cepek%2C+Jeremy&rft.au=Chronik%2C+Blaine+A&rft.au=Fenster%2C+Aaron&rft.date=2014-05-01&rft.eissn=2473-4209&rft.volume=41&rft.issue=5&rft.spage=052301&rft_id=info:doi/10.1118%2F1.4870961&rft_id=info%3Apmid%2F24784394&rft.externalDocID=24784394
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=0094-2405&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=0094-2405&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=0094-2405&client=summon