The effect of corneal crosslinking on the rigidity of the cornea estimated using a modified algorithm for the Schiøtz tonometer
Purpose: The aim of this study was to test a method for estimating corneal rigidity before and after cross-linking (CXL) using a Schiøtz tonometer. Methods: The study was performed in the Kyiv City Clinical Ophthalmological Hospital "Eye Microsurgical Center", Ukraine. This was a prospecti...
Uložené v:
| Vydané v: | Indian journal of ophthalmology Ročník 69; číslo 6; s. 1531 - 1536 |
|---|---|
| Hlavní autori: | , , , , |
| Médium: | Journal Article |
| Jazyk: | English |
| Vydavateľské údaje: |
India
Wolters Kluwer - Medknow
01.06.2021
Medknow Publications and Media Pvt. Ltd Medknow Publications & Media Pvt. Ltd Wolters Kluwer Medknow Publications |
| Vydanie: | 2 |
| Predmet: | |
| ISSN: | 0301-4738, 1998-3689, 1998-3689 |
| On-line prístup: | Získať plný text |
| Tagy: |
Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
|
| Shrnutí: | Purpose:
The aim of this study was to test a method for estimating corneal rigidity before and after cross-linking (CXL) using a Schiøtz tonometer.
Methods:
The study was performed in the Kyiv City Clinical Ophthalmological Hospital "Eye Microsurgical Center", Ukraine. This was a prospective, consecutive, randomized, masked, case-by-case, clinical study. Corneal rigidity, indicated by the gradient (G) between lg applied weight and corresponding lg scale reading during Schiøtz tonometry, were obtained by increasing (A-mode) then reducing (D-mode) weights by two operators [A] in keratoconus, post-CXL and control subjects for estimation of (i) interoperator and (ii) intersessional errors, (iii) intergroup differences; [B] before and after CXL. Central corneal thickness CCT was measured by scanning slit pachymetry. ANOVA, t tests, linear regression were the statistical tools used.
Results:
Average interoperator difference (ΔG) was -0.120 (SD = ±0.294, 95%CI = -0.175 to -0.066). A significant correlation between ΔG and the mean of each pair of G values was found (r = -0.196, n = 112, P = 0.038). Intersessional differences in mean G values were insignificant (P > 0.05). There was a significant correlation between G at first session (X1) and difference between sessions (ΔG) [Operator 1, ΔG = 0.598x1-0.461, r = 0.601, n = 27, P = 0.009]. Significant intergroup differences in G were found (Operator 1, one-way ANOVA, F = 4.489, P = 0.014). The difference (Δ) between the pre-(X2) and post-CXL treatment G values was significantly associated with the pre-CXL treatment value (Operator 1, Δ = 1.970x2-1.622, r = 0.642, n = 18, P = <.001). G values were correlated with CCT in keratoconus and post-CXL.
Conclusion:
Corneal rigidity (G) estimated using the Schiøtz tonometer can be useful for detecting changes after CXL. However, G values are linked to CCT, can vary from time-to-time and the procedure is operator dependent. |
|---|---|
| Bibliografia: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 ObjectType-Undefined-3 |
| ISSN: | 0301-4738 1998-3689 1998-3689 |
| DOI: | 10.4103/ijo.IJO_1820_20 |