A dosimetric comparison of helical tomotherapy treatment delivery with real-time adaption and no motion correction

[Display omitted] •Testing a helical tomotherapy system equipped with kV imaging and surface guidance.•A unified testing framework compared real-time adaption versus no motion correction.•Average reduction in 2 %/2 mm γ-fail rate was 17.3 % across 5 lung traces.•Average reduction in 2 %/2 mm γ-fail...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics and imaging in radiation oncology Jg. 34; S. 100741
Hauptverfasser: Hindmarsh, Jonathan, Crowe, Scott, Johnson, Julia, Sengupta, Chandrima, Walsh, Jemma, Dieterich, Sonja, Booth, Jeremy, Keall, Paul
Format: Journal Article
Sprache:Englisch
Veröffentlicht: Netherlands Elsevier B.V 01.04.2025
Elsevier
Schlagworte:
ISSN:2405-6316, 2405-6316
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:[Display omitted] •Testing a helical tomotherapy system equipped with kV imaging and surface guidance.•A unified testing framework compared real-time adaption versus no motion correction.•Average reduction in 2 %/2 mm γ-fail rate was 17.3 % across 5 lung traces.•Average reduction in 2 %/2 mm γ-fail rate was 11.8 % across 3 prostate traces.•Helical tomotherapy system performed comparably to other real-time adaptive methods. This study assesses the ability of a helical tomotherapy system equipped with kV imaging and optical surface guidance to adapt to motion traces in real-time. To assess the delivery accuracy with motion, a unified testing framework was used. The average 2 %/2 mm γ-fail rates across all lung traces were 0.1 % for motion adapted and 17.4 % for no motion correction. Average 2 %/2 mm γ-fail rates across all prostate traces were 0.4 % for motion adapted and 12.2 % for no motion correction. Real-time motion adaption was shown to improve the accuracy of dose delivered to a moving phantom compared with no motion adaption. MeSH Keywords: Radiotherapy, image-guided; Radiation therapy, targeted.
Bibliographie:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:2405-6316
2405-6316
DOI:10.1016/j.phro.2025.100741