Optic nerve sheath diameter measured sonographically as non-invasive estimator of intracranial pressure: a systematic review and meta-analysis
Summary Purpose Although invasive intracranial devices (IIDs) are the gold standard for intracranial pressure (ICP) measurement, ultrasonography of the optic nerve sheath diameter (ONSD) has been suggested as a potential non-invasive ICP estimator. We performed a meta-analysis to evaluate the diagno...
Uloženo v:
| Vydáno v: | Intensive care medicine Ročník 44; číslo 8; s. 1284 - 1294 |
|---|---|
| Hlavní autoři: | , , , , , , , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
Berlin/Heidelberg
Springer Berlin Heidelberg
01.08.2018
Springer Springer Nature B.V |
| Témata: | |
| ISSN: | 0342-4642, 1432-1238, 1432-1238 |
| On-line přístup: | Získat plný text |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | Summary
Purpose
Although invasive intracranial devices (IIDs) are the gold standard for intracranial pressure (ICP) measurement, ultrasonography of the optic nerve sheath diameter (ONSD) has been suggested as a potential non-invasive ICP estimator. We performed a meta-analysis to evaluate the diagnostic accuracy of sonographic ONSD measurement for assessment of intracranial hypertension (IH) in adult patients.
Methods
We searched on electronic databases (MEDLINE/PubMed
®
, Scopus
®
, Web of Science
®
, ScienceDirect
®
, Cochrane Library
®
) until 31 May 2018 for comparative studies that evaluated the efficacy of sonographic ONSD vs. ICP measurement with IID. Data were extracted independently by two authors. We used the QUADAS-2 tool for assessing the risk of bias (RB) of each study. A diagnostic meta-analysis following the bivariate approach and random-effects model was performed.
Results
Seven prospective studies (320 patients) were evaluated for IH detection (assumed with ICP > 20 mmHg or > 25 cmH
2
O). The accuracy of included studies ranged from 0.811 (95% CI 0.678‒0.847) to 0.954 (95% CI 0.853‒0.983). Three studies were at high RB. No significant heterogeneity was found for the diagnostic odds ratio (DOR), positive likelihood ratio (PLR) and negative likelihood ratio (NLR), with
I
2
< 50% for each parameter. The pooled DOR, PLR and NLR were 67.5 (95% CI 29‒135), 5.35 (95% CI 3.76‒7.53) and 0.088 (95% CI 0.046‒0.152), respectively. The area under the hierarchical summary receiver-operating characteristic curve (AUHSROC) was 0.938. In the subset of five studies (275 patients) with IH defined for ICP > 20 mmHg, the pooled DOR, PLR and NLR were 68.10 (95% CI 26.8‒144), 5.18 (95% CI 3.59‒7.37) and 0.087 (95% CI 0.041‒0.158), respectively, while the AUHSROC was 0.932.
Conclusions
Although the wide 95% CI in our pooled DOR suggests caution, ultrasonographic ONSD may be a potentially useful approach for assessing IH when IIDs are not indicated or available (CRD42018089137, PROSPERO). |
|---|---|