Accuracy of a computerized tomography-guided template-assisted implant placement system: an in vitro study
Objectives: To evaluate the accuracy of computer‐assisted 3D planning and implant insertion using computerized tomography (CT). Materials and methods: Nine implants were planned on pre‐operative CTs of six resin models, which were acquired with radiographic templates, using a planning software (E im...
Uloženo v:
| Vydáno v: | Clinical oral implants research Ročník 20; číslo 10; s. 1156 - 1162 |
|---|---|
| Hlavní autoři: | , , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
Oxford, UK
Blackwell Publishing Ltd
01.10.2009
|
| Témata: | |
| ISSN: | 0905-7161, 1600-0501, 1600-0501 |
| On-line přístup: | Získat plný text |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | Objectives: To evaluate the accuracy of computer‐assisted 3D planning and implant insertion using computerized tomography (CT).
Materials and methods: Nine implants were planned on pre‐operative CTs of six resin models, which were acquired with radiographic templates, using a planning software (E implants). Each resin model contained three pre‐existing control implants (C implants). Radiographic templates were converted into operative guides containing 4.8‐mm‐diameter titanium sleeves. A single set of insertable sleeves was used for consecutively drilling the six models, followed by implant insertion through the guide sleeves. Models were further divided into group A (the first three models) and group B (the last three models). Post‐operative CTs were used to compare implant positions with pre‐operative planned positions. Statistical analysis included the Mann–Whitney U test for E and C implants and the Wilcoxon's signed ranks test for groups A and B.
Results: The mean apex depth deviations for E and C implants [0.49 mm±0.36 standard deviation (SD) and 0.32 mm±0.21 SD, respectively], and the mean apex radial deviations (0.63 mm±0.38 SD and 0.49 mm±0.17 SD, respectively) were similar (P>0.05). The mean angulation deviations for E and C implants were 2.17±1.06°SD and 1.33±0.69°SD, P<0.05. E implant deviations of all the parameters in group A were significantly smaller than E implant deviations in group B.
Conclusions: Computer‐assisted implant planning and insertion provides good accuracy. Deviations are mainly related to system and reproducibility errors. Multiple use of drills and titanium sleeves significantly reduces system accuracy. |
|---|---|
| Bibliografie: | istex:FC1D2EC3A694374756DD4FC0EC88927827910C54 ark:/67375/WNG-TBLZ0TNG-P ArticleID:CLR1748 ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Undefined-1 ObjectType-Feature-3 |
| ISSN: | 0905-7161 1600-0501 1600-0501 |