Image-quality characteristics in the longitudinal direction from different image-reconstruction algorithms during single-rotation volume acquisition on head computed tomography: A phantom study

Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and streng...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Acta Radiologica Open Ročník 12; číslo 4; s. 205846012311689
Hlavní autoři: Watanabe, Ryo, Zensho, Ayako, Ohishi, Yoshitaka, Funama, Yoshinori
Médium: Journal Article
Jazyk:angličtina
Vydáno: London, England SAGE Publications 01.04.2023
Sage Publications Ltd
Témata:
ISSN:2058-4601, 2058-4601
On-line přístup:Získat plný text
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Abstract Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms. Material and methods Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction (Z-axis). Results The MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center. Conclusion Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.
AbstractList A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms. Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction ( -axis). The MBIR images showed smaller variation in CT numbers in the -axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS was lower for strong MBIR than for other algorithms. The high-frequency component at NPS was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center. Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the -axis.
A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation.BackgroundA multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation.To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms.PurposeTo investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms.Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction (Z-axis).Material and methodsSingle-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction (Z-axis).The MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center.ResultsThe MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center.Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.ConclusionModel-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.
Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms. Material and methods Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction (Z-axis). Results The MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center. Conclusion Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.
Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To investigate variations in image-quality characteristics in the longitudinal direction for different image-reconstruction algorithms and strengths with phantoms. Material and methods Single-rotation volume scans were performed on a 320-row multidetector CT volume scanner using three types of phantoms. Tube current was set to 200 mA (standard dose) and 50 mA (low dose). All images were reconstructed with filtered back projection (FBP), mild and strong levels with hybrid iterative reconstruction (HIR), and model-based IR (MBIR). Computed tomography numbers, image noise, noise power spectrum (NPS), task-based transfer function (TTF), and visual spatial resolution were used to evaluate uniformity of image quality in the longitudinal direction (Z-axis). Results The MBIR images showed smaller variation in CT numbers in the Z-axis. The difference in the highest and lowest CT numbers was smaller (5.0 Hounsfield units [HU]) for MBIR than for FBP (6.6 HU) and HIR (6.8 HU). The variations in image noise were the smallest for strong MBIR and the largest for FBP. The low-frequency component at NPS0.2 was lower for strong MBIR than for other algorithms. The high-frequency component at NPS0.8 was low in all reconstructions. For MBIR, the image resolution and TTFs were higher in the outer portion than in the center. Conclusion Model-based IR is the optimal image-reconstruction algorithm for single-volume scan of spherical subjects owing to its high in-plane resolution and uniformity of CT numbers, image noise, and NPS in the Z-axis.
Author Ayako Zensho
Ryo Watanabe
Yoshinori Funama
Yoshitaka Ohishi
Author_xml – sequence: 1
  givenname: Ryo
  orcidid: 0000-0002-2224-6277
  surname: Watanabe
  fullname: Watanabe, Ryo
  organization: Department of Medical Radiation Sciences, Faculty of Life Sciences
– sequence: 2
  givenname: Ayako
  surname: Zensho
  fullname: Zensho, Ayako
  organization: Department of Medical Radiation Sciences, Faculty of Life Sciences
– sequence: 3
  givenname: Yoshitaka
  surname: Ohishi
  fullname: Ohishi, Yoshitaka
  organization: Department of Medical Radiation Sciences, Faculty of Life Sciences
– sequence: 4
  givenname: Yoshinori
  orcidid: 0000-0003-1797-9625
  surname: Funama
  fullname: Funama, Yoshinori
  organization: Department of Medical Radiation Sciences, Faculty of Life Sciences
BackLink https://cir.nii.ac.jp/crid/1873679867275780864$$DView record in CiNii
https://www.ncbi.nlm.nih.gov/pubmed/37089818$$D View this record in MEDLINE/PubMed
BookMark eNp9ks1u1DAQxyNUREvpA3BBluDAZYsdx7HDBVUVH5UqcYFzNHWcxJVjb22n0j4eb8ZkU0opUiXLdmZ-_8mMZ14WBz54UxSvGT1lTMoPJRWqqikrOWO1alT9rDhabJvFePDgflicpHRNKWWV4LVgL4pDLikqmDoqfl1MMJjNzQzO5h3RI0TQ2USbstWJWE_yaIgLfrB57qwHRzobjc42eNLHMOFn35tofCZ2Hwqdwacc55UBN4Ro8zgl0s3R-oEk3BxyIcOeuA1ungwBfTPbZPcmXKOBjugwbedsOpLDFIYI23H3kZyR7QgeLSRhSrtXxfMeXDInd-dx8fPL5x_n3zaX379enJ9dbnTVlHlTlSX0vBJAa6lM3UDfQKO4qlgvaUWBGaUFE1edLCU3sutoCVA2WiNPuRD8uPi0xt3OV5PpNFYcwbXbiGXHXRvAtv96vB3bIdy2jC4NqhRGeH8XIYab2aTcTjZp4xx4E-bUlooKTEHIBX37CL0Oc8TX31OSc6ZEhdSbhynd5_KnvQiwFdAxpBRNf48w2i5T1P43RaiRjzTarp3Cqqx7Unm6KhPOwd-EnxK8WwXeWvzLsjMleS3RiW3Ah6Cqrvhvj3DnMw
CitedBy_id crossref_primary_10_3233_THC_248006
crossref_primary_10_1007_s12553_025_00979_3
Cites_doi 10.1148/radiol.2015132766
10.1002/mp.14319
10.1148/radiol.2020202317
10.1016/j.ejmp.2020.06.004
10.4236/ojrad.2015.54031
10.1002/mp.13763
10.1007/s00330-011-2271-7
10.1007/s00234-013-1313-z
10.1016/j.ejrad.2009.03.015
10.21037/qims-21-215
10.1007/s00330-020-06724-w
10.3174/ajnr.A4911
10.2214/AJR.14.14200
10.1007/s12194-008-0039-9
10.1259/bjr.20160634
10.1007/s00247-017-4060-1
10.1007/s12194-014-0273-2
10.1118/1.4725171
10.1007/s11604-016-0576-9
10.1007/s10072-011-0785-z
10.1016/j.acra.2016.12.017
10.1148/radiol.2019191422
10.1038/nmeth.2089
ContentType Journal Article
Copyright The Author(s) 2023
The Author(s) 2023.
The Author(s) 2023. This work is licensed under the Creative Commons Attribution – Non-Commercial License https://creativecommons.org/licenses/by-nc/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
The Author(s) 2023 2023 The Foundation Acta Radiologica
Copyright_xml – notice: The Author(s) 2023
– notice: The Author(s) 2023.
– notice: The Author(s) 2023. This work is licensed under the Creative Commons Attribution – Non-Commercial License https://creativecommons.org/licenses/by-nc/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
– notice: The Author(s) 2023 2023 The Foundation Acta Radiologica
DBID RYH
AFRWT
AAYXX
CITATION
NPM
3V.
7RV
7U7
7XB
8FI
8FJ
8FK
ABUWG
AFKRA
AZQEC
BENPR
C1K
CCPQU
DWQXO
FYUFA
GHDGH
KB0
NAPCQ
PHGZM
PHGZT
PIMPY
PKEHL
PPXIY
PQEST
PQQKQ
PQUKI
PRINS
7X8
5PM
DOI 10.1177/20584601231168986
DatabaseName CiNii Complete
Sage Journals GOLD Open Access 2024
CrossRef
PubMed
ProQuest Central (Corporate)
Nursing & Allied Health Database
Toxicology Abstracts
ProQuest Central (purchase pre-March 2016)
Hospital Premium Collection
Hospital Premium Collection (Alumni Edition)
ProQuest Central (Alumni) (purchase pre-March 2016)
ProQuest Central (Alumni)
ProQuest Central UK/Ireland
ProQuest Central Essentials
ProQuest Central - New (Subscription)
Environmental Sciences and Pollution Management
ProQuest One Community College
ProQuest Central Korea
Health Research Premium Collection
Health Research Premium Collection (Alumni)
Nursing & Allied Health Database (Alumni Edition)
Nursing & Allied Health Premium
ProQuest Central Premium
ProQuest One Academic (New)
Publicly Available Content Database
ProQuest One Academic Middle East (New)
ProQuest One Health & Nursing
ProQuest One Academic Eastern Edition (DO NOT USE)
ProQuest One Academic (retired)
ProQuest One Academic UKI Edition
ProQuest Central China
MEDLINE - Academic
PubMed Central (Full Participant titles)
DatabaseTitle CrossRef
PubMed
Publicly Available Content Database
ProQuest One Academic Middle East (New)
Toxicology Abstracts
ProQuest Central Essentials
ProQuest One Academic Eastern Edition
ProQuest Central (Alumni Edition)
ProQuest One Community College
ProQuest One Health & Nursing
ProQuest Nursing & Allied Health Source
ProQuest Hospital Collection
Health Research Premium Collection (Alumni)
ProQuest Central China
ProQuest Hospital Collection (Alumni)
Environmental Sciences and Pollution Management
ProQuest Central
Nursing & Allied Health Premium
Health Research Premium Collection
ProQuest One Academic UKI Edition
ProQuest Central Korea
ProQuest Nursing & Allied Health Source (Alumni)
ProQuest Central (New)
ProQuest One Academic
ProQuest One Academic (New)
ProQuest Central (Alumni)
MEDLINE - Academic
DatabaseTitleList PubMed
MEDLINE - Academic

Publicly Available Content Database
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7RV
  name: Nursing & Allied Health Database
  url: https://search.proquest.com/nahs
  sourceTypes: Aggregation Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
EISSN 2058-4601
ExternalDocumentID PMC10116848
37089818
10_1177_20584601231168986
10.1177_20584601231168986
Genre Journal Article
GroupedDBID 0R~
53G
5I-
5I0
5VS
7RV
8FI
8FJ
AASGM
ABUWG
ACARO
ACGFS
ACHEB
ADBBV
ADEBD
ADOGD
ADWAY
AEMJX
AEWDL
AFCOW
AFFHD
AFKRA
AFRWT
ALMA_UNASSIGNED_HOLDINGS
AOIJS
AYAKG
BAWUL
BCNDV
BENPR
BKEYQ
BPHCQ
BVXVI
CCPQU
DC.
DIK
FYUFA
GROUPED_DOAJ
H13
HYE
J8X
KQ8
M48
M~E
NAPCQ
O9-
OK1
PGMZT
PHGZM
PHGZT
PIMPY
PPXIY
PQQKQ
RPM
RYH
SAUOL
SCDPB
SCNPE
SFC
TRM
UKHRP
ALIPV
ARTOV
AAYXX
ABWRX
ADRAZ
AEWLI
AFIEG
CITATION
IPNFZ
RIG
AATBZ
ACGZU
AECGH
NPM
3V.
7U7
7XB
8FK
AZQEC
C1K
DWQXO
PKEHL
PQEST
PQUKI
PRINS
7X8
PUEGO
5PM
ID FETCH-LOGICAL-c492t-422af345a0678e69af9a983841f7040a1e8c515bd7273e7dd02aa29cc06703553
IEDL.DBID 7RV
ISICitedReferencesCount 2
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000964132700001&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 2058-4601
IngestDate Tue Nov 04 02:06:42 EST 2025
Fri Sep 05 12:44:17 EDT 2025
Sat Jul 26 03:32:33 EDT 2025
Wed Feb 19 02:24:11 EST 2025
Sat Nov 29 08:08:49 EST 2025
Tue Nov 18 22:10:26 EST 2025
Tue Jun 17 22:31:36 EDT 2025
Mon Nov 10 09:19:55 EST 2025
IsDoiOpenAccess true
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 4
Keywords Head computed tomography
reconstruction algorithm
image quality
single-rotation volume scan
longitudinal direction
Language English
License This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
The Author(s) 2023.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c492t-422af345a0678e69af9a983841f7040a1e8c515bd7273e7dd02aa29cc06703553
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ORCID 0000-0002-2224-6277
0000-0003-1797-9625
OpenAccessLink https://www.proquest.com/docview/2807331854?pq-origsite=%requestingapplication%
PMID 37089818
PQID 2807331854
PQPubID 4450824
ParticipantIDs pubmedcentral_primary_oai_pubmedcentral_nih_gov_10116848
proquest_miscellaneous_2805515578
proquest_journals_2807331854
pubmed_primary_37089818
crossref_primary_10_1177_20584601231168986
crossref_citationtrail_10_1177_20584601231168986
sage_journals_10_1177_20584601231168986
nii_cinii_1873679867275780864
PublicationCentury 2000
PublicationDate 2023-04-01
PublicationDateYYYYMMDD 2023-04-01
PublicationDate_xml – month: 04
  year: 2023
  text: 2023-04-01
  day: 01
PublicationDecade 2020
PublicationPlace London, England
PublicationPlace_xml – name: London, England
– name: England
– name: London
– name: Sage UK: London, England
PublicationTitle Acta Radiologica Open
PublicationTitleAlternate Acta Radiol Open
PublicationYear 2023
Publisher SAGE Publications
Sage Publications Ltd
Publisher_xml – name: SAGE Publications
– name: Sage Publications Ltd
References Watanabe, Ino, Yoshikawa 2015; 5
Mori, Machida 2009; 2
Solomon, Lyu, Marin 2020; 47
Brady, Trout, Somasundaram 2021; 298
Urikura, Ichikawa, Hara 2014; 7
Racine, Becce, Viry 2020; 76
Shirasaka, Hiwatashi, Yamashita 2017; 90
San Millán Ruíz, Murphy, Gailloud 2010; 74
Gervaise, Osemont, Lecocq 2012; 22
Ehman, Yu, Manduca 2014; 34
Jeon, Choi, Cheon 2018; 48
Hayashida, Hirai, Nakamura 2016; 34
Mileto, Guimaraes, McCollough 2019; 293
Biswas, Chandran, Roughley 2015; 205
Geyer, Schoepf, Meinel 2015; 276
Greffier, Dabli, Hamard 2022; 12
Richard, Husarik, Yadava 2012; 39
Schneider, Rasband, Eliceiri 2012; 9
Siebert, Diekmann, Masuhr 2012; 33
Samei, Bakalyar, Boedeker 2019; 46
Komlosi, Zhang, Leiva-Salinas 2014; 56
Radon, Chandran, Bhojak 2016; 37
Surabhi, Ashu, Arun 2020
Fletcher, Yu, Fidler 2017; 24
Greffier, Hamard, Pereira 2020; 30
bibr9-20584601231168986
Surabhi V (bibr5-20584601231168986) 2020
bibr6-20584601231168986
bibr14-20584601231168986
bibr24-20584601231168986
bibr19-20584601231168986
bibr1-20584601231168986
bibr11-20584601231168986
bibr21-20584601231168986
bibr22-20584601231168986
bibr18-20584601231168986
bibr8-20584601231168986
bibr25-20584601231168986
bibr12-20584601231168986
bibr2-20584601231168986
bibr7-20584601231168986
Ehman EC (bibr15-20584601231168986) 2014; 34
bibr4-20584601231168986
bibr23-20584601231168986
bibr10-20584601231168986
bibr17-20584601231168986
bibr20-20584601231168986
bibr26-20584601231168986
bibr13-20584601231168986
bibr16-20584601231168986
bibr3-20584601231168986
References_xml – volume: 24
  start-page: 876
  issue: 7
  year: 2017
  end-page: 890
  article-title: Estimation of observer performance for reduced radiation dose levels in CT
  publication-title: Academ Radiol
– volume: 90
  start-page: 20160634
  issue: 1070
  year: 2017
  article-title: Optimal scan timing for artery-vein separation at whole-brain CT angiography using a 320-row MDCT volume scanner
  publication-title: The British J Radiol
– volume: 276
  start-page: 339
  issue: 2
  year: 2015
  end-page: 357
  article-title: State of the art: iterative CT reconstruction techniques
  publication-title: Radiol
– volume: 22
  start-page: 295
  issue: 2
  year: 2012
  end-page: 301
  article-title: CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT
  publication-title: European Radiol
– volume: 56
  start-page: 187
  issue: 3
  year: 2014
  end-page: 193
  article-title: Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies
  publication-title: Neuroradiol
– volume: 12
  start-page: 229
  issue: 1
  year: 2022
  end-page: 243
  article-title: Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study
  publication-title: Quantitati Imag Med Surg
– volume: 39
  start-page: 4115
  issue: 7
  year: 2012
  end-page: 4122
  article-title: Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms
  publication-title: Med Phy
– volume: 2
  start-page: 22
  issue: 1
  year: 2009
  end-page: 32
  article-title: Deriving the modulation transfer function of CT from extremely noisy edge profiles
  publication-title: Radiol Phy Technol
– volume: 33
  start-page: 741
  issue: 4
  year: 2012
  end-page: 747
  article-title: Measurement of cerebral circulation times using dynamic whole-brain CT-angiography: feasibility and initial experience
  publication-title: Neurol Sci
– volume: 34
  start-page: 691
  issue: 10
  year: 2016
  end-page: 699
  article-title: Additive value of 320-section low-dose dynamic volume CT in relation to 3-T MRI for the preoperative evaluation of brain tumors
  publication-title: Japanese J Radiol
– volume: 5
  start-page: 224
  issue: 4
  year: 2015
  end-page: 234
  article-title: Discussion about improvement of stability of the scan timing by placing small ROI in cerebral 3D-CTA
  publication-title: Open J Radiol
– volume: 74
  start-page: 413
  issue: 3
  year: 2010
  end-page: 419
  article-title: 320-multidetector row whole-head dynamic subtracted CT angiography and whole-brain CT perfusion before and after carotid artery stenting: technical note
  publication-title: Eur J Radiol
– volume: 298
  start-page: 180
  issue: 1
  year: 2021
  end-page: 188
  article-title: Improving image quality and reducing radiation dose for pediatric CT by using deep learning reconstruction
  publication-title: Radiol
– volume: 205
  start-page: 1269
  issue: 6
  year: 2015
  end-page: 1275
  article-title: Cerebral CT venography using a 320-MDCT scanner with a time-density curve technique and low volume of contrast agent: comparison with fixed time-delay technique
  publication-title: American J Roentgenol
– volume: 47
  start-page: 3961
  issue: 9
  year: 2020
  end-page: 3971
  article-title: Noise and spatial resolution properties of a commercially available deep learning-based CT reconstruction algorithm
  publication-title: Medical Phy
– volume: 30
  start-page: 3951
  issue: 7
  year: 2020
  end-page: 3959
  article-title: Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
  publication-title: European Radiol
– volume: 76
  start-page: 28
  year: 2020
  end-page: 37
  article-title: Task-based characterization of a deep learning image reconstruction and comparison with filtered back-projection and a partial model-based iterative reconstruction in abdominal CT: a phantom study
  publication-title: Physica Medica, AUGUST
– year: 2020
  article-title: Diagnostic radiology: pediatric imaging
  publication-title: Tech Considerat Pediat Imag
– volume: 46
  start-page: 735
  issue: 11
  year: 2019
  end-page: 756
  article-title: Performance evaluation of computed tomography systems: summary of AAPM task group 233
  publication-title: Med Phys
– volume: 9
  start-page: 671
  issue: 7
  year: 2012
  end-page: 675
  article-title: NIH Image to ImageJ: 25 years of image analysis
  publication-title: Nat Metho
– volume: 293
  start-page: 491
  issue: 3
  year: 2019
  end-page: 503
  article-title: State of the art in abdominal CT: the limits of iterative reconstruction algorithms
  publication-title: Radiol
– volume: 7
  start-page: 358
  issue: 2
  year: 2014
  end-page: 366
  article-title: Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography
  publication-title: Radiol Phy Technol
– volume: 34
  start-page: 849
  issue: 4
  year: 2014
  end-page: 862
  article-title: Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT
  publication-title: Rad Graph
– volume: 48
  start-page: 594
  issue: 4
  year: 2018
  end-page: 601
  article-title: Unenhanced 320-row multidetector computed tomography of the brain in children: comparison of image quality and radiation dose among wide-volume, one-shot volume, and helical scan modes
  publication-title: Pediat Radiol
– volume: 37
  start-page: 2189
  issue: 12
  year: 2016
  end-page: 2194
  article-title: Radiation dose reduction in 4D cerebral CT angiography by individualized estimation of cerebral circulation time
  publication-title: American J Neuroradiol
– ident: bibr12-20584601231168986
  doi: 10.1148/radiol.2015132766
– ident: bibr13-20584601231168986
  doi: 10.1002/mp.14319
– ident: bibr19-20584601231168986
  doi: 10.1148/radiol.2020202317
– ident: bibr18-20584601231168986
  doi: 10.1016/j.ejmp.2020.06.004
– year: 2020
  ident: bibr5-20584601231168986
  publication-title: Tech Considerat Pediat Imag
– ident: bibr3-20584601231168986
  doi: 10.4236/ojrad.2015.54031
– ident: bibr21-20584601231168986
  doi: 10.1002/mp.13763
– ident: bibr10-20584601231168986
  doi: 10.1007/s00330-011-2271-7
– ident: bibr11-20584601231168986
  doi: 10.1007/s00234-013-1313-z
– ident: bibr1-20584601231168986
  doi: 10.1016/j.ejrad.2009.03.015
– ident: bibr26-20584601231168986
– ident: bibr23-20584601231168986
  doi: 10.21037/qims-21-215
– ident: bibr22-20584601231168986
  doi: 10.1007/s00330-020-06724-w
– ident: bibr8-20584601231168986
  doi: 10.3174/ajnr.A4911
– ident: bibr2-20584601231168986
  doi: 10.2214/AJR.14.14200
– ident: bibr25-20584601231168986
  doi: 10.1007/s12194-008-0039-9
– volume: 34
  start-page: 849
  issue: 4
  year: 2014
  ident: bibr15-20584601231168986
  publication-title: Rad Graph
– ident: bibr4-20584601231168986
  doi: 10.1259/bjr.20160634
– ident: bibr6-20584601231168986
  doi: 10.1007/s00247-017-4060-1
– ident: bibr14-20584601231168986
  doi: 10.1007/s12194-014-0273-2
– ident: bibr24-20584601231168986
  doi: 10.1118/1.4725171
– ident: bibr9-20584601231168986
  doi: 10.1007/s11604-016-0576-9
– ident: bibr7-20584601231168986
  doi: 10.1007/s10072-011-0785-z
– ident: bibr16-20584601231168986
  doi: 10.1016/j.acra.2016.12.017
– ident: bibr17-20584601231168986
  doi: 10.1148/radiol.2019191422
– ident: bibr20-20584601231168986
  doi: 10.1038/nmeth.2089
SSID ssj0001453651
Score 2.2292988
Snippet Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To...
A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. To investigate...
Background A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation. Purpose To...
A multi detector computed tomography (CT) scanner with wide-area coverage enables whole-brain volumetric scanning in a single rotation.BackgroundA multi...
SourceID pubmedcentral
proquest
pubmed
crossref
sage
nii
SourceType Open Access Repository
Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 205846012311689
SubjectTerms Algorithms
Computed tomography
Image acquisition
Image filters
Image processing
Image quality
Image reconstruction
Image resolution
Iterative methods
Medical imaging
Original Research
Rotation
Scanners
Spatial discrimination
Spatial resolution
Tomography
Transfer functions
Variation
Visual perception
Title Image-quality characteristics in the longitudinal direction from different image-reconstruction algorithms during single-rotation volume acquisition on head computed tomography: A phantom study
URI https://cir.nii.ac.jp/crid/1873679867275780864
https://journals.sagepub.com/doi/full/10.1177/20584601231168986
https://www.ncbi.nlm.nih.gov/pubmed/37089818
https://www.proquest.com/docview/2807331854
https://www.proquest.com/docview/2805515578
https://pubmed.ncbi.nlm.nih.gov/PMC10116848
Volume 12
WOSCitedRecordID wos000964132700001&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVHPJ
  databaseName: ROAD: Directory of Open Access Scholarly Resources
  customDbUrl:
  eissn: 2058-4601
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0001453651
  issn: 2058-4601
  databaseCode: M~E
  dateStart: 20150101
  isFulltext: true
  titleUrlDefault: https://road.issn.org
  providerName: ISSN International Centre
– providerCode: PRVPQU
  databaseName: Nursing & Allied Health Database
  customDbUrl:
  eissn: 2058-4601
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0001453651
  issn: 2058-4601
  databaseCode: 7RV
  dateStart: 20160101
  isFulltext: true
  titleUrlDefault: https://search.proquest.com/nahs
  providerName: ProQuest
– providerCode: PRVPQU
  databaseName: ProQuest Central (subscription)
  customDbUrl:
  eissn: 2058-4601
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0001453651
  issn: 2058-4601
  databaseCode: BENPR
  dateStart: 20160101
  isFulltext: true
  titleUrlDefault: https://www.proquest.com/central
  providerName: ProQuest
– providerCode: PRVPQU
  databaseName: Publicly Available Content Database
  customDbUrl:
  eissn: 2058-4601
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0001453651
  issn: 2058-4601
  databaseCode: PIMPY
  dateStart: 20160101
  isFulltext: true
  titleUrlDefault: http://search.proquest.com/publiccontent
  providerName: ProQuest
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV3di9QwEA_enYgvfn9U744RBEEotk3apL7IKXd44C3LobI-LWmadQu77Z7bFfzz_M-cabLdW07vRSjtQ9I2odOZXzIzv2HsZc4LUWQ8CbMyoQWKKMJcyyKUaF1wPWFFVHYkrp_kYKBGo3zoN9yWPqxyrRM7RV02hvbI3xBrC6dUX_FucRFS1SjyrvoSGjtsLyZsjPIsz79u9lhEyrM09s5M4llKIjK4hCPiOFM5ZVBfMkc7dVX9DWleDZi8FPXVGaKTu_87hXvsjoegcORk5j67YesH7NaZd7I_ZL9P5_ja0KVb_gKzTekMVQ0IGmHWUKWjVUlVtcAZRvzEQOkqsK660kLVPapbdfdMtaBn33FY7XS-BJclCbRhMcN-jQsMAKczQZuLVeWCygAPtBslGFeGooS2mXu67bdwBIspVUOeQ0eX-4h9OTn-_OFj6Cs9hEbkSRuKJNETLlJNttNmuZ7kOldciXgiUcvo2CqDwKsoCW1ZWZZRonWSG0NZRoiY-GO2Wze1fcogkzpNjZoUUuBK14pCWC4talHEuaYUWcCi9QcfG0-DTtU4ZuPYM59fkZGAve5vWTgOkOs6H6AU4aPpHCvJyc1F7m5Ujbh8FAHbXwvH2KuL5XgjGQF70Tfjj07eG13bZtX1Sakej1QBe-LEsR8NlxG-O8YWtSWofQciEd9uqatpRyYekydOCbz1Fcn0Zkz_nOGz62fwnN1OEAi66KZ9tovCZQ_YTfOzrZY_DrsflM4jdcj23h8Phud4HZ6eDb_9AQSaRcU
linkProvider ProQuest
linkToHtml http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMw1V1fb9MwED9tHWK88P9PYYNDAiEhRTSJEydICE3AtGpt1YchjafOcdw1Upt0tAXtS_Ed-GbcxUm7arC3PSBVfbHjxu75fmff3e8AXsV-IpLQ95ww9fiAIhInVjJxJKELnSeMaKUliWtH9nrR8XHc34BfdS4Mh1XWOrFU1Gmh-Y78HbO2-JzqKz5OzxyuGsXe1bqEhhWLQ3P-k45ssw_tz_T_vva8_S9Hnw6cqqqAo0XszR3heWroi0CxnjZhrIaxiiM_Eu5QkkQr10SaQD5JGdmNTNOWp5QXa80ZLYTOPo27CVuChb0BW_12t_9tdasjAj8M3Mp9ysxOXoshni0X1w2jmHO2LwDgZp5lf7NtL4doXogzK6Fv_87_tmh34XZlZOOe3RX3YMPk9-FmtwojeAC_2xOapmMTSs9Rr5NWY5YjmcU4LriW0yLlumFooZ-EGDkhB-u6MnPMyqHKe4UlFy-q8Sktw3w0maHNA0W-khlTv8KGPqBFBVT6bJHZsDmkDyFjitoW2khxXkwqQvH3uIfTEdd7nmBJCPwQvl7LAj6CRl7k5glgKFUQ6GiYSEFneSMSYXxpCCfIktepCJvQqgVsoCuid643Mh64Fbf7JZlswtvlI1PLcnJV512SWhqav91I-uzIY4c-KX86IIsm7NTCOKgU4mywksQmvFw2kypj_5TKTbEo-wRccUhGTXhsxX_5Nr5s0W-71BKtbYxlB6ZJX2_Js1FJl-6yrzES9Ogb3kOrd_rnDJ9ePYMXsH1w1O0MOu3e4TO45ZHZa2O5dqBBgmZ24Yb-Mc9m359X6gHh5Lo31x8DsZ1L
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Image-quality+characteristics+in+the+longitudinal+direction+from+different+image-reconstruction+algorithms+during+single-rotation+volume+acquisition+on+head+computed+tomography%3A+A+phantom+study&rft.jtitle=Acta+radiologica+open&rft.au=Watanabe%2C+Ryo&rft.au=Zensho%2C+Ayako&rft.au=Ohishi%2C+Yoshitaka&rft.au=Funama%2C+Yoshinori&rft.date=2023-04-01&rft.issn=2058-4601&rft.eissn=2058-4601&rft.volume=12&rft.issue=4&rft.spage=205846012311689&rft_id=info:doi/10.1177%2F20584601231168986&rft.externalDBID=n%2Fa&rft.externalDocID=10_1177_20584601231168986
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=2058-4601&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=2058-4601&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=2058-4601&client=summon