A disposable biosensor for urea determination in blood based on an ammonium-sensitive transducer
A potentiometric urea-sensitive biosensor using a NH + 4-sensitive disposable electrode in double matrix membrane (DMM) technology as transducer is described. The ion-sensitive polymer matrix membrane was formed in the presence of an additional electrochemical inert filter paper matrix to improve th...
Uložené v:
| Vydané v: | Biosensors & bioelectronics Ročník 14; číslo 1; s. 33 - 41 |
|---|---|
| Hlavní autori: | , , , , , , , |
| Médium: | Journal Article |
| Jazyk: | English |
| Vydavateľské údaje: |
Lausanne
Elsevier B.V
1999
Elsevier Science |
| Predmet: | |
| ISSN: | 0956-5663, 1873-4235 |
| On-line prístup: | Získať plný text |
| Tagy: |
Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
|
| Shrnutí: | A potentiometric urea-sensitive biosensor using a NH
+
4-sensitive disposable electrode in double matrix membrane (DMM) technology as transducer is described. The ion-sensitive polymer matrix membrane was formed in the presence of an additional electrochemical inert filter paper matrix to improve the reproducibility in sensor production. The electrodes were prepared from one-side silver-coated filter paper, which is encapsulated for insulation by a heat-sealing film. A defined volume of the NH
+
4-sensitive polymer matrix membrane cocktail was deposited on this filter paper. To obtain the urea-biosensor a layer of urease was cast onto the ion-sensitive membrane. Poly (carbamoylsulfonate) hydrogel, produced from a hydrophilic polyurethane prepolymer blocked with bisulfite, served as immobilisation material. The disposable urea-sensitive electrode was combined with a disposable Ag/AgCl reference electrode to obtain the disposable urea biosensor. The sensor responded rapidly and in a stable manner to changes in urea concentrations between 7.2×10
−5 and 2.1×10
−2
mol/l. The detection limit was 2×10
−5
mol/l urea and the slope in the linear range 52
mV/decade. By taking into consideration the influence of the interfering K
+- and Na
+- ions the sensor can be used for the determination of urea in human blood and serum samples (diluted or undiluted). A good correlation was found with the data obtained by the spectrophotometric routine method. |
|---|---|