Measuring the photon depth dose distribution produced by a medical linear accelerator in a water-equivalent radio-fluorogenic gel

The present work describes use of a water-equivalent radio-fluorogenic gel dosimeter for measurement of a depth dose distribution from a medical linear accelerator. Relative depth dose distributions for a 6 MV photon beam were measured with a novel radio-fluorogenic gel comprised of aqueous gelatin...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Journal of radioanalytical and nuclear chemistry Ročník 307; číslo 3; s. 2505 - 2508
Hlavní autoři: Sandwall, Peter A., Spitz, Henry B., Elson, Howard R., Lamba, Michael A. S., Connick, William B., Fenichel, Henry
Médium: Journal Article
Jazyk:angličtina
Vydáno: Dordrecht Springer Netherlands 01.03.2016
Springer
Témata:
ISSN:0236-5731, 1588-2780
On-line přístup:Získat plný text
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Popis
Shrnutí:The present work describes use of a water-equivalent radio-fluorogenic gel dosimeter for measurement of a depth dose distribution from a medical linear accelerator. Relative depth dose distributions for a 6 MV photon beam were measured with a novel radio-fluorogenic gel comprised of aqueous gelatin and coumarin-3-carboxylic acid. Agreement was within 3 % of published values in most areas of electronic equilibrium. Results support continued development of radio-fluorogenic gel dosimetry systems for quality assurance of clinical photon beams.
ISSN:0236-5731