Measuring the photon depth dose distribution produced by a medical linear accelerator in a water-equivalent radio-fluorogenic gel
The present work describes use of a water-equivalent radio-fluorogenic gel dosimeter for measurement of a depth dose distribution from a medical linear accelerator. Relative depth dose distributions for a 6 MV photon beam were measured with a novel radio-fluorogenic gel comprised of aqueous gelatin...
Uloženo v:
| Vydáno v: | Journal of radioanalytical and nuclear chemistry Ročník 307; číslo 3; s. 2505 - 2508 |
|---|---|
| Hlavní autoři: | , , , , , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
Dordrecht
Springer Netherlands
01.03.2016
Springer |
| Témata: | |
| ISSN: | 0236-5731, 1588-2780 |
| On-line přístup: | Získat plný text |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | The present work describes use of a water-equivalent radio-fluorogenic gel dosimeter for measurement of a depth dose distribution from a medical linear accelerator. Relative depth dose distributions for a 6 MV photon beam were measured with a novel radio-fluorogenic gel comprised of aqueous gelatin and coumarin-3-carboxylic acid. Agreement was within 3 % of published values in most areas of electronic equilibrium. Results support continued development of radio-fluorogenic gel dosimetry systems for quality assurance of clinical photon beams. |
|---|---|