Portable and platform‐independent MR pulse sequence programs
Purpose To introduce a new sequence description format for vendor‐independent MR sequences that include all calculation logic portably. To introduce a new MRI sequence development approach which utilizes flexibly reusable modules. Methods The proposed sequence description contains a sequence module...
Uložené v:
| Vydané v: | Magnetic resonance in medicine Ročník 83; číslo 4; s. 1277 - 1290 |
|---|---|
| Hlavní autori: | , , , |
| Médium: | Journal Article |
| Jazyk: | English |
| Vydavateľské údaje: |
United States
Wiley Subscription Services, Inc
01.04.2020
|
| Predmet: | |
| ISSN: | 0740-3194, 1522-2594, 1522-2594 |
| On-line prístup: | Získať plný text |
| Tagy: |
Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
|
| Shrnutí: | Purpose
To introduce a new sequence description format for vendor‐independent MR sequences that include all calculation logic portably. To introduce a new MRI sequence development approach which utilizes flexibly reusable modules.
Methods
The proposed sequence description contains a sequence module hierarchy for loop and group logic, which is enhanced by a novel strategy for performing efficient parameter and pulse shape calculation. These calculations are powered by a flow graph structure. By using the flow graph, all calculations are performed with no redundancy and without requiring preprocessing. The generation of this interpretable structure is a separate step that combines MRI techniques while actively considering their context. The driver interface is slim and highly flexible through scripting support. The sequences do not require any vendor‐specific compiling or processing step. A vendor‐independent frontend for sequence configuration can be used. Tests that ensure physical feasibility of the sequence are integrated into the calculation logic.
Results
The framework was used to define a set of standard sequences. Resulting images were compared to respective images acquired with sequences provided by the device manufacturer. Images were acquired using a standard commercial MRI system.
Conclusions
The approach produces configurable, vendor‐independent sequences, whose configurability enables rapid prototyping. The transparent data structure simplifies the process of sharing reproducible sequences, modules, and techniques. |
|---|---|
| Bibliografia: | Funding information This work was supported by the FhG Internal Programs under Grant No. Attract 142‐600172. ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
| ISSN: | 0740-3194 1522-2594 1522-2594 |