Laser Target System in Combination with an Aiming Device for Percutaneous CT-Guided Interventions - An Accuracy Study

To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments. MATERIALS AND METHODS: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additi...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Academic radiology Ročník 30; číslo 12; s. 3047
Hlavní autoři: Scharll, Yannick, Böhler, Daniel, Laimer, Gregor, Schullian, Peter, Bale, Reto
Médium: Journal Article
Jazyk:angličtina
Vydáno: United States 01.12.2023
Témata:
ISSN:1878-4046, 1878-4046
On-line přístup:Zjistit podrobnosti o přístupu
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Abstract To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments. MATERIALS AND METHODS: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additional help of an aiming device. Conically shaped targets in a plexiglass phantom were punctured. The planning CT data sets were acquired with 1.25, 2.5 and 5 mm slice thickness. Needle placement accuracy, as well as procedural time, was assessed. The Euclidean (ED) and normal distances (ND) were calculated at the target point. Using the aiming device, the accomplished mean ND at the target for the 1.25, 2.5 and 5 mm slice thickness was 1.76 mm (SD ± 0.92), 2.09 mm (SD ± 1.06) and 1.93 mm (SD ± 1.38), respectively. Without aiming device, the corresponding results were 2.55 mm (SD ± 1.42), 2.7 mm (SD ± 1.43) and 2.31 mm (SD ± 1.64). At a slice thickness of 1.25 mm and 2.5 mm, punctures with the aiming device were significantly more accurate for both the ED and ND as compared to the punctures without aiming device (p < 0.001). The mean time required to complete the procedure, including image acquisition, trajectory planning, the placement of 10 needles, and the control-CT scan was 24.8 min without and 29.8 min with the aiming device. The additional use of the aiming device in combination with the commercially available laser guidance system significantly increased the level of accuracy during this in vitro experiment compared to freehand passes.
AbstractList To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments. MATERIALS AND METHODS: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additional help of an aiming device. Conically shaped targets in a plexiglass phantom were punctured. The planning CT data sets were acquired with 1.25, 2.5 and 5 mm slice thickness. Needle placement accuracy, as well as procedural time, was assessed. The Euclidean (ED) and normal distances (ND) were calculated at the target point.RATIONALE AND OBJECTIVESTo evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments. MATERIALS AND METHODS: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additional help of an aiming device. Conically shaped targets in a plexiglass phantom were punctured. The planning CT data sets were acquired with 1.25, 2.5 and 5 mm slice thickness. Needle placement accuracy, as well as procedural time, was assessed. The Euclidean (ED) and normal distances (ND) were calculated at the target point.Using the aiming device, the accomplished mean ND at the target for the 1.25, 2.5 and 5 mm slice thickness was 1.76 mm (SD ± 0.92), 2.09 mm (SD ± 1.06) and 1.93 mm (SD ± 1.38), respectively. Without aiming device, the corresponding results were 2.55 mm (SD ± 1.42), 2.7 mm (SD ± 1.43) and 2.31 mm (SD ± 1.64). At a slice thickness of 1.25 mm and 2.5 mm, punctures with the aiming device were significantly more accurate for both the ED and ND as compared to the punctures without aiming device (p < 0.001). The mean time required to complete the procedure, including image acquisition, trajectory planning, the placement of 10 needles, and the control-CT scan was 24.8 min without and 29.8 min with the aiming device.RESULTSUsing the aiming device, the accomplished mean ND at the target for the 1.25, 2.5 and 5 mm slice thickness was 1.76 mm (SD ± 0.92), 2.09 mm (SD ± 1.06) and 1.93 mm (SD ± 1.38), respectively. Without aiming device, the corresponding results were 2.55 mm (SD ± 1.42), 2.7 mm (SD ± 1.43) and 2.31 mm (SD ± 1.64). At a slice thickness of 1.25 mm and 2.5 mm, punctures with the aiming device were significantly more accurate for both the ED and ND as compared to the punctures without aiming device (p < 0.001). The mean time required to complete the procedure, including image acquisition, trajectory planning, the placement of 10 needles, and the control-CT scan was 24.8 min without and 29.8 min with the aiming device.The additional use of the aiming device in combination with the commercially available laser guidance system significantly increased the level of accuracy during this in vitro experiment compared to freehand passes.CONCLUSIONThe additional use of the aiming device in combination with the commercially available laser guidance system significantly increased the level of accuracy during this in vitro experiment compared to freehand passes.
To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments. MATERIALS AND METHODS: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additional help of an aiming device. Conically shaped targets in a plexiglass phantom were punctured. The planning CT data sets were acquired with 1.25, 2.5 and 5 mm slice thickness. Needle placement accuracy, as well as procedural time, was assessed. The Euclidean (ED) and normal distances (ND) were calculated at the target point. Using the aiming device, the accomplished mean ND at the target for the 1.25, 2.5 and 5 mm slice thickness was 1.76 mm (SD ± 0.92), 2.09 mm (SD ± 1.06) and 1.93 mm (SD ± 1.38), respectively. Without aiming device, the corresponding results were 2.55 mm (SD ± 1.42), 2.7 mm (SD ± 1.43) and 2.31 mm (SD ± 1.64). At a slice thickness of 1.25 mm and 2.5 mm, punctures with the aiming device were significantly more accurate for both the ED and ND as compared to the punctures without aiming device (p < 0.001). The mean time required to complete the procedure, including image acquisition, trajectory planning, the placement of 10 needles, and the control-CT scan was 24.8 min without and 29.8 min with the aiming device. The additional use of the aiming device in combination with the commercially available laser guidance system significantly increased the level of accuracy during this in vitro experiment compared to freehand passes.
Author Bale, Reto
Laimer, Gregor
Scharll, Yannick
Schullian, Peter
Böhler, Daniel
Author_xml – sequence: 1
  givenname: Yannick
  surname: Scharll
  fullname: Scharll, Yannick
  organization: Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.)
– sequence: 2
  givenname: Daniel
  surname: Böhler
  fullname: Böhler, Daniel
  organization: Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.)
– sequence: 3
  givenname: Gregor
  surname: Laimer
  fullname: Laimer, Gregor
  organization: Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.)
– sequence: 4
  givenname: Peter
  surname: Schullian
  fullname: Schullian, Peter
  organization: Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.)
– sequence: 5
  givenname: Reto
  surname: Bale
  fullname: Bale, Reto
  email: reto.bale@tirol-klinken.at
  organization: Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria (Y.S., D.B., G.L., P.S., R.B.). Electronic address: reto.bale@tirol-klinken.at
BackLink https://www.ncbi.nlm.nih.gov/pubmed/37117142$$D View this record in MEDLINE/PubMed
BookMark eNpNUEtLAzEYDFKxD_0DHiRHL7sm2Vf2WKpWoaDQel7y-LamdLM1yVb237vSCsLAzGFmGGaKRra1gNAtJTElNH_YxUI5ETPCkpgMYPwCTSgveJSSNB_902M09X5HCM1ynlyhcVJQWtCUTVC3Eh4c3gi3hYDXvQ_QYGPxom2ksSKY1uJvEz6xsHhuGmO3-BGORgGuW4ffwakuCAtt5_FiEy07o0HjVxvAHcH-pj2O8HzIKtU5oXq8Dp3ur9FlLfYebs48Qx_PT5vFS7R6W74u5qtIpRkNUZ1pmmqVS00ozUDxDIQuBIe0LmUBhVZ1LXPIBKGSZpzWJIGSp0omGZMMJJuh-1PvwbVfHfhQNcYr2O9PkyvGSVEOp9BysN6drZ1sQFcHZxrh-urvKvYDqRdvXQ
CitedBy_id crossref_primary_10_1186_s13019_024_02706_x
crossref_primary_10_1055_s_0044_1786724
crossref_primary_10_1186_s12893_024_02698_4
crossref_primary_10_3390_diagnostics14131371
crossref_primary_10_3390_diagnostics14161723
crossref_primary_10_1140_epjp_s13360_024_05158_3
ContentType Journal Article
Copyright Copyright © 2023 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.
Copyright_xml – notice: Copyright © 2023 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.
DBID NPM
7X8
DOI 10.1016/j.acra.2023.03.028
DatabaseName PubMed
MEDLINE - Academic
DatabaseTitle PubMed
MEDLINE - Academic
DatabaseTitleList MEDLINE - Academic
PubMed
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod no_fulltext_linktorsrc
Discipline Medicine
EISSN 1878-4046
ExternalDocumentID 37117142
Genre Journal Article
GroupedDBID ---
--K
.1-
.FO
0R~
1B1
1P~
23M
4.4
457
53G
5GY
5RE
AAEDT
AAEDW
AALRI
AAWTL
AAXUO
ABJNI
ABMAC
ACGFS
ADBBV
AENEX
AEVXI
AFCTW
AFRHN
AFTJW
AITUG
AJUYK
ALMA_UNASSIGNED_HOLDINGS
AMRAJ
BELOY
C5W
CS3
EBS
EFJIC
F5P
FDB
G-Q
HZ~
IHE
J1W
KOM
M41
MO0
NPM
O9-
OI~
OU0
P2P
ROL
RPZ
SEL
SES
SJN
SSZ
XH2
Z5R
7X8
AFJKZ
APXCP
EFKBS
ID FETCH-LOGICAL-c451t-f5d14dc6bd0115ec85ead7a8e4f9b7e7dcffb6e5a01b1581f03e984cb352b2eb2
IEDL.DBID 7X8
ISICitedReferencesCount 6
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=001128173100001&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 1878-4046
IngestDate Sun Sep 28 07:11:42 EDT 2025
Thu Jan 02 22:53:01 EST 2025
IsDoiOpenAccess false
IsOpenAccess true
IsPeerReviewed true
IsScholarly true
Issue 12
Keywords Laser guidance
Accuracy
Percutaneous
Aiming device
CT-guidance
Language English
License Copyright © 2023 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c451t-f5d14dc6bd0115ec85ead7a8e4f9b7e7dcffb6e5a01b1581f03e984cb352b2eb2
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
OpenAccessLink http://www.academicradiology.org/article/S1076633223001642/pdf
PMID 37117142
PQID 2807915619
PQPubID 23479
ParticipantIDs proquest_miscellaneous_2807915619
pubmed_primary_37117142
PublicationCentury 2000
PublicationDate 2023-12-01
PublicationDateYYYYMMDD 2023-12-01
PublicationDate_xml – month: 12
  year: 2023
  text: 2023-12-01
  day: 01
PublicationDecade 2020
PublicationPlace United States
PublicationPlace_xml – name: United States
PublicationTitle Academic radiology
PublicationTitleAlternate Acad Radiol
PublicationYear 2023
SSID ssj0015683
Score 2.4224887
Snippet To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro...
SourceID proquest
pubmed
SourceType Aggregation Database
Index Database
StartPage 3047
Title Laser Target System in Combination with an Aiming Device for Percutaneous CT-Guided Interventions - An Accuracy Study
URI https://www.ncbi.nlm.nih.gov/pubmed/37117142
https://www.proquest.com/docview/2807915619
Volume 30
WOSCitedRecordID wos001128173100001&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1LS8QwEA7qinjx_X4wgtdg0yRtepJldVXQZQ8r7G1p85C9dLVrBf-9k6ayXgTBS2-BNjOdfDPz5RtCLosiRlwsHWU2SqmwEafKcENj55VKeC54HIZNpIOBGo-zYVtwm7e0yu-Y2ARqM9O-Rn7lVVsyTDZYdv36Rv3UKN9dbUdoLJMORyjjKV3peNFFkEkjw8kUZkoCE8H20kzgd-W68rpDMW9ETmP1O8Rsjpr-5n9fcotstCATusErtsmSLXfI2lPbRt8l9SOeXRWMGhY4BNFymJaAwQET5cZW4Au0kJfQ9WO_XuDG-pACCHFhaCtdI6a0s3oOvRG9q6fGGnj4wZ6cA4UurtW6rnL9CZ6t-LlHnvu3o949becvUC0ke6dOGiaMTgrjcaPVSqLbpbmywmVFalOjnSsSK_OIFUwq5iJuMyV0gaCuiDFl3ycr5ay0hwQQVJpY5lzmWojIaOUSxzD1SpJMCowqR-Tie0Mn6N--aRE-Y7LY0iNyEKwyeQ1CHBOeMj-_PT7-w-oTsu6NHZgop6Tj8O-2Z2RVf7xP59V54zj4HAyfvgAFqc0l
linkProvider ProQuest
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Laser+Target+System+in+Combination+with+an+Aiming+Device+for+Percutaneous+CT-Guided+Interventions+-+An+Accuracy+Study&rft.jtitle=Academic+radiology&rft.au=Scharll%2C+Yannick&rft.au=B%C3%B6hler%2C+Daniel&rft.au=Laimer%2C+Gregor&rft.au=Schullian%2C+Peter&rft.date=2023-12-01&rft.issn=1878-4046&rft.eissn=1878-4046&rft.volume=30&rft.issue=12&rft.spage=3047&rft_id=info:doi/10.1016%2Fj.acra.2023.03.028&rft.externalDBID=NO_FULL_TEXT
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1878-4046&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1878-4046&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1878-4046&client=summon