Investigation of PVA cryogel Young’s modulus stability with time, controlled by a simple reliable technique
We describe a quasistatic method for mechanical characterization of tissue-mimicking material used in elastography. We demonstrate that it is possible to assess the elasticity modulus with a reasonable reproducibility using simple and easy tools and methods. Possessing a simple relevant technique wi...
Saved in:
| Published in: | Medical physics (Lancaster) Vol. 36; no. 2; pp. 656 - 661 |
|---|---|
| Main Authors: | , , , , , |
| Format: | Journal Article |
| Language: | English |
| Published: |
United States
American Association of Physicists in Medicine
01.02.2009
|
| Subjects: | |
| ISSN: | 0094-2405, 2473-4209 |
| Online Access: | Get full text |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | We describe a quasistatic method for mechanical characterization of tissue-mimicking material used in elastography. We demonstrate that it is possible to assess the elasticity modulus with a reasonable reproducibility using simple and easy tools and methods. Possessing a simple relevant technique with evaluated relative error to assess Young’s modulus of these phantoms could deeply improve the quality of the research in the field of elastography. The method was tested and validated with four samples of polyvinyl alcohol (PVA) cryogel with different elasticity values corresponding to those of stiffer soft biological tissues. Young’s moduli, varying from
70
to
180
kPa
depending on the number of freeze-thaw cycles (two to five), were measured within strict measurement conditions and found to have a reproducibility varying from 4% to 8%. Relative error, estimated as the ratio between observed and reference values, varied from 16% to 32%. Besides, measurement stability over
4
months
was evaluated. The method demonstrated good feasibility and acceptable reproducibility for mechanically characterizing and controlled over time phantoms used for validating new potential ultrasound imaging techniques in the field of elastography. Nevertheless, in this study, investigation was performed on gel possessing young’s modulus values ranging from
80
to
215
kPa
. Some tissue values of Young’modulus were reported to be lower, ranging from
0.6
to
28
kPa
as liver or glandular values. Consequently, further validation of this static method for mechanical characterization of phantom gels should be performed using softer PVA cryogel. |
|---|---|
| Bibliography: | duboeuf.francois@creatis.insa-lyon.fr Electronic mail ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 0094-2405 2473-4209 |
| DOI: | 10.1118/1.3065031 |