High levels of hemolysis do not affect measurement of PAPP-A, β-HCG and TRAb on BRAHMS KRYPTOR compact plus

To assess the impact of high levels of hemolysis on the laboratory results for free β-hCG, PAPP-A, and TRAb performed on the B·R·A·H·M·S KRYPTOR Compact PLUS. Adapted from the CLSI guidelines EP07-A2, paired difference testing was performed on serum samples from the routine laboratory workflow. Thre...

Celý popis

Uložené v:
Podrobná bibliografia
Vydané v:Scandinavian journal of clinical and laboratory investigation Ročník ahead-of-print; číslo ahead-of-print; s. 1 - 4
Hlavní autori: Pettersson Pablo, Paul, Aneskans, Cecilia, Vink, Martin
Médium: Journal Article
Jazyk:English
Vydavateľské údaje: England Taylor & Francis 03.10.2023
Predmet:
ISSN:0036-5513, 1502-7686, 1502-7686
On-line prístup:Získať plný text
Tagy: Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
Popis
Shrnutí:To assess the impact of high levels of hemolysis on the laboratory results for free β-hCG, PAPP-A, and TRAb performed on the B·R·A·H·M·S KRYPTOR Compact PLUS. Adapted from the CLSI guidelines EP07-A2, paired difference testing was performed on serum samples from the routine laboratory workflow. Three sample pools for each assessed analyte was prepared and subjected to increased levels of added hemolysate. For β-hCG and PAPP-A, the relative difference in the measured analyte concentration between the sample with 0 g/L added Hb and the samples with increasing free Hb concentrations (up to 6 g/L), was well below the pre-set acceptance criterion of 10% at all levels. The TRAb results showed greater variation than the other analytes, likely a consequence of imprecision rather than hemolysis. Hemolysis has a negligible effect on the analysis results of free beta-hCG, PAPP-A and TRAb measured on the B·R·A·H·M·S KRYPTOR Compact PLUS.
Bibliografia:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0036-5513
1502-7686
1502-7686
DOI:10.1080/00365513.2023.2233886