Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia

During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10)...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Journal of applied physiology (1985) Ročník 92; číslo 1; s. 162
Hlavní autoři: Yamaya, Yoshiki, Bogaard, Harm J, Wagner, Peter D, Niizeki, Kyuichi, Hopkins, Susan R
Médium: Journal Article
Jazyk:angličtina
Vydáno: United States 01.01.2002
Témata:
ISSN:8750-7587
On-line přístup:Zjistit podrobnosti o přístupu
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Abstract During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (Sa(O(2))) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (VO(2 max)) under various conditions [normoxia, hypoxia inspired oxygen fraction (FI(O(2))) = 0.125; hyperoxia, FI(O(2)) = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, Sa(O(2)) = 73-99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices tested (y = 1.009x - 0.52, R(2) = 0.90, B+/-P = 0.3 +/- 2.5). Finger sensors had low precision and a significant negative bias (D-25: y = 1.004x - 2.327, R(2) = 0.52, B+/-P = -2.0 +/- 7.3; Ivy: y = 1.237x - 24.2, R(2) = 0.78, B+/-P = -2.0 +/- 5.2). Eliminating measurements in which heart rate differed by >10 beats/min from the electrocardiogram value improved precision minimally and did not affect bias substantially (B+/-P = 0.5 +/- 2.0, -1.8 +/- 8.4, and -1.25+/-4.33 for RS-10, D-25, and Ivy, respectively). Signal detection algorithms and pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.
AbstractList During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (Sa(O(2))) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (VO(2 max)) under various conditions [normoxia, hypoxia inspired oxygen fraction (FI(O(2))) = 0.125; hyperoxia, FI(O(2)) = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, Sa(O(2)) = 73-99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices tested (y = 1.009x - 0.52, R(2) = 0.90, B+/-P = 0.3 +/- 2.5). Finger sensors had low precision and a significant negative bias (D-25: y = 1.004x - 2.327, R(2) = 0.52, B+/-P = -2.0 +/- 7.3; Ivy: y = 1.237x - 24.2, R(2) = 0.78, B+/-P = -2.0 +/- 5.2). Eliminating measurements in which heart rate differed by >10 beats/min from the electrocardiogram value improved precision minimally and did not affect bias substantially (B+/-P = 0.5 +/- 2.0, -1.8 +/- 8.4, and -1.25+/-4.33 for RS-10, D-25, and Ivy, respectively). Signal detection algorithms and pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.
During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (Sa(O(2))) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (VO(2 max)) under various conditions [normoxia, hypoxia inspired oxygen fraction (FI(O(2))) = 0.125; hyperoxia, FI(O(2)) = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, Sa(O(2)) = 73-99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices tested (y = 1.009x - 0.52, R(2) = 0.90, B+/-P = 0.3 +/- 2.5). Finger sensors had low precision and a significant negative bias (D-25: y = 1.004x - 2.327, R(2) = 0.52, B+/-P = -2.0 +/- 7.3; Ivy: y = 1.237x - 24.2, R(2) = 0.78, B+/-P = -2.0 +/- 5.2). Eliminating measurements in which heart rate differed by >10 beats/min from the electrocardiogram value improved precision minimally and did not affect bias substantially (B+/-P = 0.5 +/- 2.0, -1.8 +/- 8.4, and -1.25+/-4.33 for RS-10, D-25, and Ivy, respectively). Signal detection algorithms and pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (Sa(O(2))) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (VO(2 max)) under various conditions [normoxia, hypoxia inspired oxygen fraction (FI(O(2))) = 0.125; hyperoxia, FI(O(2)) = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, Sa(O(2)) = 73-99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices tested (y = 1.009x - 0.52, R(2) = 0.90, B+/-P = 0.3 +/- 2.5). Finger sensors had low precision and a significant negative bias (D-25: y = 1.004x - 2.327, R(2) = 0.52, B+/-P = -2.0 +/- 7.3; Ivy: y = 1.237x - 24.2, R(2) = 0.78, B+/-P = -2.0 +/- 5.2). Eliminating measurements in which heart rate differed by >10 beats/min from the electrocardiogram value improved precision minimally and did not affect bias substantially (B+/-P = 0.5 +/- 2.0, -1.8 +/- 8.4, and -1.25+/-4.33 for RS-10, D-25, and Ivy, respectively). Signal detection algorithms and pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.
Author Bogaard, Harm J
Wagner, Peter D
Niizeki, Kyuichi
Hopkins, Susan R
Yamaya, Yoshiki
Author_xml – sequence: 1
  givenname: Yoshiki
  surname: Yamaya
  fullname: Yamaya, Yoshiki
  organization: Department of Medicine, University of California, San Diego, La Jolla, California 92093, USA
– sequence: 2
  givenname: Harm J
  surname: Bogaard
  fullname: Bogaard, Harm J
– sequence: 3
  givenname: Peter D
  surname: Wagner
  fullname: Wagner, Peter D
– sequence: 4
  givenname: Kyuichi
  surname: Niizeki
  fullname: Niizeki, Kyuichi
– sequence: 5
  givenname: Susan R
  surname: Hopkins
  fullname: Hopkins, Susan R
BackLink https://www.ncbi.nlm.nih.gov/pubmed/11744656$$D View this record in MEDLINE/PubMed
BookMark eNo1kElPwzAQhX0oogv8BfCJEwnj1I7jI6rYpEpcoNfI8UJdOQtOIjX_Hlctp5n39M2T3izRrGkbg9A9gZQQlj0dZNf5bj_1rvUpAAWRZgBkhhYFZ5BwVvA5Wvb9IZqUMnKN5oRwSnOWL9BuJ73Tbphwa3E3-t7g9uhqM4QJ6zG45gfXMhrSY3M0QbkIuAY3bagjJx_xfurOi2z0SZhwkjfoysoYdnuZK_T9-vK1eU-2n28fm-dtotaCDwm3wAuTUchVlTNLuLLrXAFkSmrBRQGVFkwInWvLjMgNsHhRaCtjwUpBla3Qwzm3C-3vaPqhrF2vjPeyMe3Yl5ysKQPOInh3AceqNrrsQuwUpvL_E9kfLvJlBw
CitedBy_id crossref_primary_10_1371_journal_pone_0191869
crossref_primary_10_1016_j_forsciint_2005_10_014
crossref_primary_10_1113_expphysiol_2014_080820
crossref_primary_10_1159_000522517
crossref_primary_10_1002_ajhb_24090
crossref_primary_10_1016_j_prrv_2024_06_001
crossref_primary_10_1007_s00421_007_0554_0
crossref_primary_10_1080_02640414_2025_2474333
crossref_primary_10_1152_japplphysiol_01106_2006
crossref_primary_10_1111_j_1399_6576_2009_02101_x
crossref_primary_10_1152_japplphysiol_00180_2007
crossref_primary_10_1088_0967_3334_25_5_009
crossref_primary_10_1249_MSS_0000000000001658
crossref_primary_10_3233_IES_200237
crossref_primary_10_3390_s21041263
crossref_primary_10_1161_CIR_0b013e3181e52e69
crossref_primary_10_1097_01_jsm_0000171287_99174_b7
crossref_primary_10_1152_japplphysiol_01088_2002
crossref_primary_10_3389_fspor_2020_00041
crossref_primary_10_1111_j_1600_0838_2006_00563_x
crossref_primary_10_1111_j_1398_9995_2005_00914_x
crossref_primary_10_1183_09031936_00024410
crossref_primary_10_3390_s22114236
crossref_primary_10_1111_cpf_12193
crossref_primary_10_1016_S1569_9048_02_00166_0
crossref_primary_10_14814_phy2_12747
crossref_primary_10_1016_j_jcf_2008_09_008
crossref_primary_10_1089_ham_2007_1058
crossref_primary_10_1016_j_chest_2016_11_001
crossref_primary_10_1371_journal_pone_0187266
crossref_primary_10_7189_jogh_12_04036
crossref_primary_10_1002_j_2040_4603_2013_tb00500_x
crossref_primary_10_1111_j_1445_5994_2006_01069_x
crossref_primary_10_1113_jphysiol_2003_040568
crossref_primary_10_1097_PEP_0b013e318208ca9e
crossref_primary_10_1111_j_1467_2995_2005_00271_x
crossref_primary_10_1016_j_jshs_2015_07_005
crossref_primary_10_1155_2012_824091
crossref_primary_10_1016_j_resp_2013_11_009
crossref_primary_10_1152_japplphysiol_00491_2009
crossref_primary_10_1016_j_nanoen_2021_106755
crossref_primary_10_1249_MSS_0000000000001473
crossref_primary_10_4103_jpcc_jpcc_26_25
crossref_primary_10_1055_s_0043_1770362
crossref_primary_10_1088_1361_6579_aacfe5
crossref_primary_10_1183_16000617_0029_2023
crossref_primary_10_1183_16000617_0101_2018
crossref_primary_10_1371_journal_pone_0146087
crossref_primary_10_1371_journal_pone_0222487
crossref_primary_10_1123_jsr_2019_0307
crossref_primary_10_1152_japplphysiol_00323_2001
crossref_primary_10_1016_j_resp_2004_06_011
crossref_primary_10_1378_chest_07_0617
crossref_primary_10_3390_bios11100391
crossref_primary_10_1007_s00421_011_2175_x
crossref_primary_10_1177_0271678X221117008
crossref_primary_10_1016_S0761_8425_07_91122_5
crossref_primary_10_4187_respcare_02494
crossref_primary_10_3390_jcm9124062
crossref_primary_10_1378_chest_08_2222
crossref_primary_10_1016_j_jcf_2010_09_006
crossref_primary_10_1186_1471_2253_11_10
crossref_primary_10_1016_j_resp_2006_07_002
crossref_primary_10_1002_ppul_10411
crossref_primary_10_3109_10903120903349747
crossref_primary_10_1136_vetreccr_2016_000320
crossref_primary_10_1016_j_chest_2020_04_017
crossref_primary_10_1177_14799731211070844
crossref_primary_10_1249_MSS_0000000000001745
ContentType Journal Article
DBID CGR
CUY
CVF
ECM
EIF
NPM
7X8
DOI 10.1152/japplphysiol.00409.2001
DatabaseName Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
DatabaseTitle MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList MEDLINE
MEDLINE - Academic
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod no_fulltext_linktorsrc
Discipline Medicine
Anatomy & Physiology
ExternalDocumentID 11744656
Genre Research Support, U.S. Gov't, P.H.S
Clinical Trial
Research Support, Non-U.S. Gov't
Journal Article
GrantInformation_xml – fundername: NCRR NIH HHS
  grantid: MO1 RR-00827
– fundername: NHLBI NIH HHS
  grantid: HL-17731
GroupedDBID ---
-~X
.55
.GJ
18M
1CY
29J
2WC
39C
3O-
4.4
53G
5VS
85S
8M5
AAFWJ
ABCQX
ABDNZ
ABHWK
ABJNI
ABKWE
ABOCM
ACBEA
ACGFO
ACGFS
ACIWK
ACKIV
ACPRK
ACYGS
ADBBV
ADFNX
AEILP
AENEX
AFOSN
AFRAH
AGCDD
AGNAY
AI.
AIDAL
AJUXI
ALMA_UNASSIGNED_HOLDINGS
BAWUL
BKKCC
BTFSW
C1A
C2-
CGR
CS3
CUY
CVF
DIK
DU5
E3Z
EBS
ECM
EIF
EJD
EMOBN
F5P
FRP
GX1
H13
H~9
ITBOX
J5H
KQ8
L7B
MVM
NEJ
NPM
OHT
OK1
P-O
P2P
P6G
PQQKQ
RAP
RHF
RHI
RPL
RPRKH
SJN
TR2
UHB
UKR
UPT
VH1
VXZ
W8F
WH7
WOQ
X7M
XOL
XSW
YBH
YCJ
YQJ
YQT
YWH
ZXP
~02
7X8
ABUFD
ADXHL
ID FETCH-LOGICAL-c397t-7f078e2406cb65f17cf36c002cad97980bd9599d6df5e96e057f08dfa200bc0b2
IEDL.DBID 7X8
ISICitedReferencesCount 85
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000173205400022&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 8750-7587
IngestDate Sun Nov 09 11:04:16 EST 2025
Wed Feb 19 01:19:42 EST 2025
IsPeerReviewed true
IsScholarly true
Issue 1
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c397t-7f078e2406cb65f17cf36c002cad97980bd9599d6df5e96e057f08dfa200bc0b2
Notes ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
PMID 11744656
PQID 71345075
PQPubID 23479
ParticipantIDs proquest_miscellaneous_71345075
pubmed_primary_11744656
PublicationCentury 2000
PublicationDate 2002-Jan
20020101
PublicationDateYYYYMMDD 2002-01-01
PublicationDate_xml – month: 01
  year: 2002
  text: 2002-Jan
PublicationDecade 2000
PublicationPlace United States
PublicationPlace_xml – name: United States
PublicationTitle Journal of applied physiology (1985)
PublicationTitleAlternate J Appl Physiol (1985)
PublicationYear 2002
SSID ssj0014451
Score 2.0122063
Snippet During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and...
SourceID proquest
pubmed
SourceType Aggregation Database
Index Database
StartPage 162
SubjectTerms Adult
Algorithms
Bicycling - physiology
Blood Gas Analysis
Electrocardiography - drug effects
Exercise - physiology
Female
Heart Failure - physiopathology
Heart Rate - drug effects
Humans
Hypoxia - blood
Male
Oximetry - standards
Oxygen - blood
Oxygen Consumption - drug effects
Physical Fitness - physiology
Pulmonary Disease, Chronic Obstructive - physiopathology
Reproducibility of Results
Title Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia
URI https://www.ncbi.nlm.nih.gov/pubmed/11744656
https://www.proquest.com/docview/71345075
Volume 92
WOSCitedRecordID wos000173205400022&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1LS8NAEB7Uinjx0fqozz2IJ2PTNJtkQZAiFg-29KClt7LZh1ZsEvsQ---d3SR4Eg9elmzYhGUymfl2ngAXTUFVS8TSkdoU1fZwiLTvO7GirkDhiHxkE4Ufw14vGg5ZfwVuylwYE1ZZykQrqGUqjI28YXIeEbvQ2-zDMT2jjG-1aKCxCpUWAhkT0BUOf3wIpvRWbt1zHUTFYRHdhQqr8Wa8w9Z0kL5fGz62KSvN31Gm1Tad7f_tcwe2CpRJ2jlb7MKKSqpQayd4wp4sySXp55tKX5ZV2OgW7vUaDAaIyiXicpJqki1Qa5L0azxR8-mS5PmMZMLxBr66bNVExglJEPfiOn5FXpdZfsETaSZqaqZ78Ny5f7p7cIrGC45AeDJ3Qo3AQRldL-KA6mYodCsQKDsFlyxkkRtLRhmTgdRUsUAh5tNuJDVHIsbCjb19WEvSRB0C4V4ziKiUHi71VSg4pR5zfcmYoqHPRR3OSzKOkLGNt4InKl3MRiUh63CQf4lRltffMPXJbZm3oz-fPYZN273FmkxOoKLxl1ansC4-5-PZ9MzyC469fvcbyzjMWQ
linkProvider ProQuest
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Validity+of+pulse+oximetry+during+maximal+exercise+in+normoxia%2C+hypoxia%2C+and+hyperoxia&rft.jtitle=Journal+of+applied+physiology+%281985%29&rft.au=Yamaya%2C+Yoshiki&rft.au=Bogaard%2C+Harm+J&rft.au=Wagner%2C+Peter+D&rft.au=Niizeki%2C+Kyuichi&rft.date=2002-01-01&rft.issn=8750-7587&rft.volume=92&rft.issue=1&rft.spage=162&rft_id=info:doi/10.1152%2Fjapplphysiol.00409.2001&rft_id=info%3Apmid%2F11744656&rft_id=info%3Apmid%2F11744656&rft.externalDocID=11744656
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=8750-7587&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=8750-7587&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=8750-7587&client=summon