A method of predicting the image noise in paediatric multi-slice computed tomography images
The purpose of this study was to develop an equation with which to determine the tube current to be used in order to obtain a certain image noise level for differently sized children undergoing multi-slice computed tomography examination. The relationship between image noise and detector dose for di...
Saved in:
| Published in: | Radiation protection dosimetry Vol. 114; no. 1-3; p. 313 |
|---|---|
| Main Authors: | , , , , , , |
| Format: | Journal Article |
| Language: | English |
| Published: |
England
01.01.2005
|
| Subjects: | |
| ISSN: | 0144-8420 |
| Online Access: | Get more information |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | The purpose of this study was to develop an equation with which to determine the tube current to be used in order to obtain a certain image noise level for differently sized children undergoing multi-slice computed tomography examination. The relationship between image noise and detector dose for different examination protocols was established for a LightSpeed Ultra, an eight slice CT from GEMS, using homogeneous water phantoms of different sizes. Three different anatomical areas (head, thorax and abdomen) were studied in 111 patients between 0 and 17 y of age. The mean ratio between the calculated and the measured noise in patient images was established for the different areas. Head examinations showed the best correlation (measured-to-calculated noise ratio = 1.01). In the thorax, the calculated noise was generally higher than the measured noise (ratio = 0.74), and in the abdomen, the opposite result was found (ratio = 1.20). |
|---|---|
| Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 0144-8420 |
| DOI: | 10.1093/rpd/nch568 |