Assessing cardiac function from total-variation-regularized 4D C-arm CT in the presence of angular undersampling
Time-resolved tomographic cardiac imaging using an angiographic C-arm device may support clinicians during minimally invasive therapy by enabling a thorough analysis of the heart function directly in the catheter laboratory. However, clinically feasible acquisition protocols entail a highly challeng...
Gespeichert in:
| Veröffentlicht in: | Physics in medicine & biology Jg. 62; H. 7; S. 2762 - 2777 |
|---|---|
| Hauptverfasser: | , , , , , , |
| Format: | Journal Article |
| Sprache: | Englisch |
| Veröffentlicht: |
England
07.04.2017
|
| Schlagworte: | |
| ISSN: | 1361-6560 |
| Online-Zugang: | Weitere Angaben |
| Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
| Zusammenfassung: | Time-resolved tomographic cardiac imaging using an angiographic C-arm device may support clinicians during minimally invasive therapy by enabling a thorough analysis of the heart function directly in the catheter laboratory. However, clinically feasible acquisition protocols entail a highly challenging reconstruction problem which suffers from sparse angular sampling of the trajectory. Compressed sensing theory promises that useful images can be recovered despite massive undersampling by means of sparsity-based regularization. For a multitude of reasons-most notably the desired reduction of scan time, dose and contrast agent required-it is of great interest to know just how little data is actually sufficient for a certain task. In this work, we apply a convex optimization approach based on primal-dual splitting to 4D cardiac C-arm computed tomography. We examine how the quality of spatially and temporally total-variation-regularized reconstruction degrades when using as few as [Formula: see text] projection views per heart phase. First, feasible regularization weights are determined in a numerical phantom study, demonstrating the individual benefits of both regularizers. Secondly, a task-based evaluation is performed in eight clinical patients. Semi-automatic segmentation-based volume measurements of the left ventricular blood pool performed on strongly undersampled images show a correlation of close to 99% with measurements obtained from less sparsely sampled data. |
|---|---|
| Bibliographie: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 1361-6560 |
| DOI: | 10.1088/1361-6560/aa6241 |