Fuzzy Visual Hull Algorithm for Three-Dimensional Shape Reconstruction of TKA Implants from X-Ray Cone-Beam Images

Three-Dimensional (3-D) shape reconstruction of total knee arthroplasty (TKA) implants in vivo plays a key role to investigate implanted knee kinematics. TKA implants typically consist of metal femoral and tibial components and a polyethylene tibial insert. X-ray computed tomography (CT) causes seve...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of advanced computational intelligence and intelligent informatics Jg. 14; H. 2; S. 122 - 127
Hauptverfasser: Kobashi, Syoji, Shibanuma, Nao, Hata, Yutaka
Format: Journal Article
Sprache:Englisch
Veröffentlicht: 20.03.2010
ISSN:1343-0130, 1883-8014
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Three-Dimensional (3-D) shape reconstruction of total knee arthroplasty (TKA) implants in vivo plays a key role to investigate implanted knee kinematics. TKA implants typically consist of metal femoral and tibial components and a polyethylene tibial insert. X-ray computed tomography (CT) causes severe metal artifacts, making the 3-D shape in reconstructed images extremely difficult to understand. This article proposes a new method of 3-D reconstruction from X-ray cone-beam images. Called a fuzzy visual hull, it introduces fuzzy logic in recognizing X-ray images. X-ray cone-beam images are fuzzified and back-projected into a fuzzy voxel space. Defuzzifying the fuzzy voxel space enables the 3-D TKA implant shape to be reconstructed. The results of evaluation using TKA implants in vitro and computer-synthesized images demonstrated that the fuzzy visual hull provides high robustness against noise added to X-ray cone-beam images. The new approach also reconstructed the 3-D polyethylene insert despite the difficulty of recognizing the region in conventional X-ray CT.
ISSN:1343-0130
1883-8014
DOI:10.20965/jaciii.2010.p0122