High-intensity Focused Ultrasound Treatment in Moderate Glaucoma Patients: Results of a 2-Year Prospective Clinical Trial
Ultrasound Cyclo Plasty (UCP) treatment using high-intensity focused ultrasound (HIFU) is an effective and safe therapy to reduce intraocular pressure (IOP) in moderate glaucoma patients as was measured during a 2-year follow-up period. The purpose of this study was to evaluate the long-term safety...
Saved in:
| Published in: | Journal of glaucoma Vol. 29; no. 7; p. 556 |
|---|---|
| Main Authors: | , , , , , , |
| Format: | Journal Article |
| Language: | English |
| Published: |
United States
01.07.2020
|
| Subjects: | |
| ISSN: | 1536-481X, 1536-481X |
| Online Access: | Get more information |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | Ultrasound Cyclo Plasty (UCP) treatment using high-intensity focused ultrasound (HIFU) is an effective and safe therapy to reduce intraocular pressure (IOP) in moderate glaucoma patients as was measured during a 2-year follow-up period.
The purpose of this study was to evaluate the long-term safety and efficacy of the UCP procedure using HIFU in moderate glaucoma patients.
A prospective interventional noncomparative study was carried out. Fifteen patients (15 eyes) with moderate open-angle glaucoma were enrolled. All eyes were treated with UCP-HIFU. A thorough ophthalmic examination and IOP measurements were performed before the UCP-HIFU procedure and at 1 day, 1 week, 4 weeks, 3 months, 6 months, 1 year, and 2 years after the procedure. The primary outcome was defined as a surgical success (IOP reduction of 20% or ≥5 mm Hg) at the last follow-up visit. The secondary outcomes were the mean IOP at each follow-up visit, number of medications used, complications profile, and reinterventions.
The mean preoperative IOP at baseline was 26.8±5.0 mm Hg. All patients had a positive response and a lower IOP after treatment, with a relatively stable 31% reduction in IOP during the follow-up period. A significant reduction in IOP was observed at all postprocedure examination points (P<0.01), with a mean 17.6±4.4 mm Hg at 2 years after the procedure (P=0.005). Surgical success was achieved in 87% of the patients at their last follow-up visit. There was a nonsignificant decrease in the mean number of glaucoma medications from 2.5±0.8 to 2.0±1.0 at 2 years (P=0.48). No major intraoperative or postoperative complications were noted.
UCP-HIFU treatment is an effective, safe, and well-tolerated method to reduce IOP in patients with moderate glaucoma. |
|---|---|
| Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 1536-481X 1536-481X |
| DOI: | 10.1097/IJG.0000000000001497 |