High-intensity focused ultrasound treatment in refractory glaucoma patients: results at 1 year of prospective clinical study
To evaluate the safety and efficacy of the ultrasonic circular cyclo-coagulation procedure using high-intensity focused ultrasound by a miniaturized annular device containing 6 piezoceramic transducers in patients with refractory glaucoma. This was a prospective interventional noncomparative study o...
Uloženo v:
| Vydáno v: | European journal of ophthalmology Ročník 25; číslo 6; s. 483 |
|---|---|
| Hlavní autoři: | , , , , |
| Médium: | Journal Article |
| Jazyk: | angličtina |
| Vydáno: |
United States
01.11.2015
|
| Témata: | |
| ISSN: | 1724-6016, 1724-6016 |
| On-line přístup: | Zjistit podrobnosti o přístupu |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
|
| Shrnutí: | To evaluate the safety and efficacy of the ultrasonic circular cyclo-coagulation procedure using high-intensity focused ultrasound by a miniaturized annular device containing 6 piezoceramic transducers in patients with refractory glaucoma.
This was a prospective interventional noncomparative study of 20 eyes of 20 patients with refractory glaucoma. All eyes were treated with 6 activated transducers operating at 21 MHz. Ultrasound biomicroscopy and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, and 1, 3, 6, and 12 months after the procedure. Primary outcomes were surgical success (defined as intraocular pressure (IOP) reduction from baseline ≥20% and IOP >5 mm Hg) at the last follow-up visit. Secondary outcomes were mean IOP at each follow-up visit compared to baseline, medication use, complications, and re-interventions.
Intraocular pressure was significantly reduced (p<0.01) from a mean preoperative value of 36.4 ± 5.7 mm Hg to a mean postoperative value of 22.5 ± 10.3 mm Hg at 12 months. Four patients needed to be re-treated. The mean IOP reduction achieved was 38%. Surgical success was achieved in 13 of 20 eyes (65%). No major intraoperative or postoperative complications occurred.
Ultrasonic circular cyclo-coagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing 6 piezoceramic transducers is an effective and well-tolerated method to reduce IOP in patients with refractory glaucoma. |
|---|---|
| Bibliografie: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
| ISSN: | 1724-6016 1724-6016 |
| DOI: | 10.5301/ejo.5000620 |