A clinical study of electrophysiological correlates of behavioural comfort levels in cochlear implantees

Objectives Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Per...

Celý popis

Uložené v:
Podrobná bibliografia
Vydané v:Cochlear implants international Ročník 15; číslo 3; s. 145 - 160
Hlavní autori: Raghunandhan, S, Ravikumar, A, Kameswaran, Mohan, Mandke, Kalyani, Ranjith, R
Médium: Journal Article
Jazyk:English
Vydavateľské údaje: England Taylor & Francis 01.05.2014
Predmet:
ISSN:1467-0100, 1754-7628, 1754-7628
On-line prístup:Získať plný text
Tagy: Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
Abstract Objectives Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Percept (MAP) and appropriate MAP becomes the key issue in the habilitation program. In 'Difficult to MAP' scenarios, objective measures become paramount to predict optimal current levels to be set in the MAP. We aimed to (a) study the trends in multi-modal electrophysiological tests and behavioural responses sequentially over the first year of implant use; (b) generate normative data from the above; (c) correlate the multi-modal electrophysiological thresholds levels with behavioural comfort levels; and (d) create predictive formulae for deriving optimal comfort levels (if unknown), using linear and multiple regression analysis. Methods This prospective study included 10 profoundly hearing impaired children aged between 2 and 7 years with normal inner ear anatomy and no additional handicaps. They received the Advanced Bionics HiRes 90 K Implant with Harmony Speech processor and used HiRes-P with Fidelity 120 strategy. They underwent, impedance telemetry, neural response imaging, electrically evoked stapedial response telemetry (ESRT), and electrically evoked auditory brainstem response (EABR) tests at 1, 4, 8, and 12 months of implant use, in conjunction with behavioural mapping. Trends in electrophysiological and behavioural responses were analyzed using paired t-test. By Karl Pearson's correlation method, electrode-wise correlations were derived for neural response imaging (NRI) thresholds versus most comfortable level (M-levels) and offset based (apical, mid-array, and basal array) correlations for EABR and ESRT thresholds versus M-levels were calculated over time. These were used to derive predictive formulae by linear and multiple regression analysis. Such statistically predicted M-levels were compared with the behaviourally recorded M-levels among the cohort, using Cronbach's alpha reliability test method for confirming the efficacy of this method. Results NRI, ESRT, and EABR thresholds showed statistically significant positive correlations with behavioural M-levels, which improved with implant use over time. These correlations were used to derive predicted M-levels using regression analysis. On an average, predicted M-levels were found to be statistically reliable and they were a fair match to the actual behavioural M-levels. When applied in clinical practice, the predicted values were found to be useful for programming members of the study group. However, individuals showed considerable deviations in behavioural M-levels, above and below the electrophysiologically predicted values, due to various factors. While the current method appears helpful as a reference to predict initial maps in 'difficult to Map' subjects, it is recommended that behavioural measures are mandatory to further optimize the maps for these individuals. Conclusion The study explores the trends, correlations and individual variabilities that occur between electrophysiological tests and behavioural responses, recorded over time among a cohort of cochlear implantees. The statistical method shown may be used as a guideline to predict optimal behavioural levels in difficult situations among future implantees, bearing in mind that optimal M-levels for individuals can vary from predicted values. In 'Difficult to MAP' scenarios, following a protocol of sequential behavioural programming, in conjunction with electrophysiological correlates will provide the best outcomes.
AbstractList Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Percept (MAP) and appropriate MAP becomes the key issue in the habilitation program. In 'Difficult to MAP' scenarios, objective measures become paramount to predict optimal current levels to be set in the MAP. We aimed to (a) study the trends in multi-modal electrophysiological tests and behavioural responses sequentially over the first year of implant use; (b) generate normative data from the above; (c) correlate the multi-modal electrophysiological thresholds levels with behavioural comfort levels; and (d) create predictive formulae for deriving optimal comfort levels (if unknown), using linear and multiple regression analysis. This prospective study included 10 profoundly hearing impaired children aged between 2 and 7 years with normal inner ear anatomy and no additional handicaps. They received the Advanced Bionics HiRes 90 K Implant with Harmony Speech processor and used HiRes-P with Fidelity 120 strategy. They underwent, impedance telemetry, neural response imaging, electrically evoked stapedial response telemetry (ESRT), and electrically evoked auditory brainstem response (EABR) tests at 1, 4, 8, and 12 months of implant use, in conjunction with behavioural mapping. Trends in electrophysiological and behavioural responses were analyzed using paired t-test. By Karl Pearson's correlation method, electrode-wise correlations were derived for neural response imaging (NRI) thresholds versus most comfortable level (M-levels) and offset based (apical, mid-array, and basal array) correlations for EABR and ESRT thresholds versus M-levels were calculated over time. These were used to derive predictive formulae by linear and multiple regression analysis. Such statistically predicted M-levels were compared with the behaviourally recorded M-levels among the cohort, using Cronbach's alpha reliability test method for confirming the efficacy of this method. NRI, ESRT, and EABR thresholds showed statistically significant positive correlations with behavioural M-levels, which improved with implant use over time. These correlations were used to derive predicted M-levels using regression analysis. On an average, predicted M-levels were found to be statistically reliable and they were a fair match to the actual behavioural M-levels. When applied in clinical practice, the predicted values were found to be useful for programming members of the study group. However, individuals showed considerable deviations in behavioural M-levels, above and below the electrophysiologically predicted values, due to various factors. While the current method appears helpful as a reference to predict initial maps in 'difficult to Map' subjects, it is recommended that behavioural measures are mandatory to further optimize the maps for these individuals. The study explores the trends, correlations and individual variabilities that occur between electrophysiological tests and behavioural responses, recorded over time among a cohort of cochlear implantees. The statistical method shown may be used as a guideline to predict optimal behavioural levels in difficult situations among future implantees, bearing in mind that optimal M-levels for individuals can vary from predicted values. In 'Difficult to MAP' scenarios, following a protocol of sequential behavioural programming, in conjunction with electrophysiological correlates will provide the best outcomes.
Objectives Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Percept (MAP) and appropriate MAP becomes the key issue in the habilitation program. In 'Difficult to MAP' scenarios, objective measures become paramount to predict optimal current levels to be set in the MAP. We aimed to (a) study the trends in multi-modal electrophysiological tests and behavioural responses sequentially over the first year of implant use; (b) generate normative data from the above; (c) correlate the multi-modal electrophysiological thresholds levels with behavioural comfort levels; and (d) create predictive formulae for deriving optimal comfort levels (if unknown), using linear and multiple regression analysis. Methods This prospective study included 10 profoundly hearing impaired children aged between 2 and 7 years with normal inner ear anatomy and no additional handicaps. They received the Advanced Bionics HiRes 90 K Implant with Harmony Speech processor and used HiRes-P with Fidelity 120 strategy. They underwent, impedance telemetry, neural response imaging, electrically evoked stapedial response telemetry (ESRT), and electrically evoked auditory brainstem response (EABR) tests at 1, 4, 8, and 12 months of implant use, in conjunction with behavioural mapping. Trends in electrophysiological and behavioural responses were analyzed using paired t-test. By Karl Pearson's correlation method, electrode-wise correlations were derived for neural response imaging (NRI) thresholds versus most comfortable level (M-levels) and offset based (apical, mid-array, and basal array) correlations for EABR and ESRT thresholds versus M-levels were calculated over time. These were used to derive predictive formulae by linear and multiple regression analysis. Such statistically predicted M-levels were compared with the behaviourally recorded M-levels among the cohort, using Cronbach's alpha reliability test method for confirming the efficacy of this method. Results NRI, ESRT, and EABR thresholds showed statistically significant positive correlations with behavioural M-levels, which improved with implant use over time. These correlations were used to derive predicted M-levels using regression analysis. On an average, predicted M-levels were found to be statistically reliable and they were a fair match to the actual behavioural M-levels. When applied in clinical practice, the predicted values were found to be useful for programming members of the study group. However, individuals showed considerable deviations in behavioural M-levels, above and below the electrophysiologically predicted values, due to various factors. While the current method appears helpful as a reference to predict initial maps in 'difficult to Map' subjects, it is recommended that behavioural measures are mandatory to further optimize the maps for these individuals. Conclusion The study explores the trends, correlations and individual variabilities that occur between electrophysiological tests and behavioural responses, recorded over time among a cohort of cochlear implantees. The statistical method shown may be used as a guideline to predict optimal behavioural levels in difficult situations among future implantees, bearing in mind that optimal M-levels for individuals can vary from predicted values. In 'Difficult to MAP' scenarios, following a protocol of sequential behavioural programming, in conjunction with electrophysiological correlates will provide the best outcomes.
Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Percept (MAP) and appropriate MAP becomes the key issue in the habilitation program. In 'Difficult to MAP' scenarios, objective measures become paramount to predict optimal current levels to be set in the MAP. We aimed to (a) study the trends in multi-modal electrophysiological tests and behavioural responses sequentially over the first year of implant use; (b) generate normative data from the above; (c) correlate the multi-modal electrophysiological thresholds levels with behavioural comfort levels; and (d) create predictive formulae for deriving optimal comfort levels (if unknown), using linear and multiple regression analysis.OBJECTIVESIndications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant based on behavioural responses may be tedious for audiologists in such cases, wherein matching an effective Measurable Auditory Percept (MAP) and appropriate MAP becomes the key issue in the habilitation program. In 'Difficult to MAP' scenarios, objective measures become paramount to predict optimal current levels to be set in the MAP. We aimed to (a) study the trends in multi-modal electrophysiological tests and behavioural responses sequentially over the first year of implant use; (b) generate normative data from the above; (c) correlate the multi-modal electrophysiological thresholds levels with behavioural comfort levels; and (d) create predictive formulae for deriving optimal comfort levels (if unknown), using linear and multiple regression analysis.This prospective study included 10 profoundly hearing impaired children aged between 2 and 7 years with normal inner ear anatomy and no additional handicaps. They received the Advanced Bionics HiRes 90 K Implant with Harmony Speech processor and used HiRes-P with Fidelity 120 strategy. They underwent, impedance telemetry, neural response imaging, electrically evoked stapedial response telemetry (ESRT), and electrically evoked auditory brainstem response (EABR) tests at 1, 4, 8, and 12 months of implant use, in conjunction with behavioural mapping. Trends in electrophysiological and behavioural responses were analyzed using paired t-test. By Karl Pearson's correlation method, electrode-wise correlations were derived for neural response imaging (NRI) thresholds versus most comfortable level (M-levels) and offset based (apical, mid-array, and basal array) correlations for EABR and ESRT thresholds versus M-levels were calculated over time. These were used to derive predictive formulae by linear and multiple regression analysis. Such statistically predicted M-levels were compared with the behaviourally recorded M-levels among the cohort, using Cronbach's alpha reliability test method for confirming the efficacy of this method.METHODSThis prospective study included 10 profoundly hearing impaired children aged between 2 and 7 years with normal inner ear anatomy and no additional handicaps. They received the Advanced Bionics HiRes 90 K Implant with Harmony Speech processor and used HiRes-P with Fidelity 120 strategy. They underwent, impedance telemetry, neural response imaging, electrically evoked stapedial response telemetry (ESRT), and electrically evoked auditory brainstem response (EABR) tests at 1, 4, 8, and 12 months of implant use, in conjunction with behavioural mapping. Trends in electrophysiological and behavioural responses were analyzed using paired t-test. By Karl Pearson's correlation method, electrode-wise correlations were derived for neural response imaging (NRI) thresholds versus most comfortable level (M-levels) and offset based (apical, mid-array, and basal array) correlations for EABR and ESRT thresholds versus M-levels were calculated over time. These were used to derive predictive formulae by linear and multiple regression analysis. Such statistically predicted M-levels were compared with the behaviourally recorded M-levels among the cohort, using Cronbach's alpha reliability test method for confirming the efficacy of this method.NRI, ESRT, and EABR thresholds showed statistically significant positive correlations with behavioural M-levels, which improved with implant use over time. These correlations were used to derive predicted M-levels using regression analysis. On an average, predicted M-levels were found to be statistically reliable and they were a fair match to the actual behavioural M-levels. When applied in clinical practice, the predicted values were found to be useful for programming members of the study group. However, individuals showed considerable deviations in behavioural M-levels, above and below the electrophysiologically predicted values, due to various factors. While the current method appears helpful as a reference to predict initial maps in 'difficult to Map' subjects, it is recommended that behavioural measures are mandatory to further optimize the maps for these individuals.RESULTSNRI, ESRT, and EABR thresholds showed statistically significant positive correlations with behavioural M-levels, which improved with implant use over time. These correlations were used to derive predicted M-levels using regression analysis. On an average, predicted M-levels were found to be statistically reliable and they were a fair match to the actual behavioural M-levels. When applied in clinical practice, the predicted values were found to be useful for programming members of the study group. However, individuals showed considerable deviations in behavioural M-levels, above and below the electrophysiologically predicted values, due to various factors. While the current method appears helpful as a reference to predict initial maps in 'difficult to Map' subjects, it is recommended that behavioural measures are mandatory to further optimize the maps for these individuals.The study explores the trends, correlations and individual variabilities that occur between electrophysiological tests and behavioural responses, recorded over time among a cohort of cochlear implantees. The statistical method shown may be used as a guideline to predict optimal behavioural levels in difficult situations among future implantees, bearing in mind that optimal M-levels for individuals can vary from predicted values. In 'Difficult to MAP' scenarios, following a protocol of sequential behavioural programming, in conjunction with electrophysiological correlates will provide the best outcomes.CONCLUSIONThe study explores the trends, correlations and individual variabilities that occur between electrophysiological tests and behavioural responses, recorded over time among a cohort of cochlear implantees. The statistical method shown may be used as a guideline to predict optimal behavioural levels in difficult situations among future implantees, bearing in mind that optimal M-levels for individuals can vary from predicted values. In 'Difficult to MAP' scenarios, following a protocol of sequential behavioural programming, in conjunction with electrophysiological correlates will provide the best outcomes.
Author Raghunandhan, S
Kameswaran, Mohan
Ranjith, R
Ravikumar, A
Mandke, Kalyani
Author_xml – sequence: 1
  givenname: S
  surname: Raghunandhan
  fullname: Raghunandhan, S
  email: raghunandhansampath@gmail.com
  organization: Consultant ENT SurgeonsMadras ENT Research Foundation, Chennai
– sequence: 2
  givenname: A
  surname: Ravikumar
  fullname: Ravikumar, A
  organization: Department of ENT and H & N Surgery, Sri Ramachandra University
– sequence: 3
  givenname: Mohan
  surname: Kameswaran
  fullname: Kameswaran, Mohan
  organization: Consultant ENT SurgeonsMadras ENT Research Foundation, Chennai
– sequence: 4
  givenname: Kalyani
  surname: Mandke
  fullname: Mandke, Kalyani
  organization: Mandke Hearing ServicesPune
– sequence: 5
  givenname: R
  surname: Ranjith
  fullname: Ranjith, R
  organization: MERF- Institute of Speech & HearingChennai
BackLink https://www.ncbi.nlm.nih.gov/pubmed/24606544$$D View this record in MEDLINE/PubMed
BookMark eNqFkctOHDEQRa2IKDx_AfWSTYNfY7clNiMUSCQkNrDIynLb5Ywjd3tiu4nm7-lhBgWxwZuyfO4tl-oeo4MxjYDQOcGXhEh1ReSCS0E7wn9d4v0R_As62oJ2Sw7mOxeyxQTjQ3Rcyh-MGeEL_g0dUi6wWHB-hFbLxsYwBmtiU-rkNk3yDUSwNaf1alNCiun3K7UpZ4imQtlKeliZ55Cm_EoGn3JtIjxDLE0Y5xe7imByE4Z1NGMFKKfoqzexwNm-nqCn2--PNz_a-4e7nzfL-9bSjtfWG8U7ZT0TPbMCnOqVxFICc8xjx61yFjz01FGrvJoJcVZISrvOY-E7yU7Qxa7vOqe_E5Sqh1AsxHkMSFPRZMEwo5hiNUvP99KpH8DpdQ6DyRv9tp1ZcL0T2JxKyeC1DdXUkMaaTYiaYL0NQ78LQ_8PY7aLD_a3Hz41LnfGMM6LHcy_lKPT1Wxiyj6b0Yai2Sc9XgBJwKKh
CitedBy_id crossref_primary_10_1016_j_ijporl_2017_08_026
crossref_primary_10_1186_s12938_020_00844_6
crossref_primary_10_1080_14992027_2021_2015543
crossref_primary_10_1371_journal_pone_0272943
crossref_primary_10_1007_s00405_024_08890_9
crossref_primary_10_1017_S0022215124000227
crossref_primary_10_1017_S0022215116009026
crossref_primary_10_1097_AUD_0000000000000495
crossref_primary_10_1007_s00106_024_01509_9
crossref_primary_10_1007_s00405_025_09532_4
crossref_primary_10_1080_14992027_2017_1370139
crossref_primary_10_1177_23312165211031751
Cites_doi 10.1097/mao.0b013e3180577919
10.1159/000066154
10.1097/00003446-199404000-00008
10.1097/00003446-200004000-00009
10.1016/j.heares.2008.06.011
10.1097/01.aud.0000146178.84065.b3
10.1097/00003446-200112000-00004
10.1080/00016480510026890
10.1097/00020840-200310000-00013
10.1097/00003446-200204000-00001
10.1016/S0165-5876(01)00426-8
ContentType Journal Article
Copyright W. S. Maney & Son Ltd 2014 2014
Copyright_xml – notice: W. S. Maney & Son Ltd 2014 2014
DBID AAYXX
CITATION
CGR
CUY
CVF
ECM
EIF
NPM
7X8
DOI 10.1179/1754762814Y.0000000064
DatabaseName CrossRef
Medline
MEDLINE
MEDLINE (Ovid)
MEDLINE
MEDLINE
PubMed
MEDLINE - Academic
DatabaseTitle CrossRef
MEDLINE
Medline Complete
MEDLINE with Full Text
PubMed
MEDLINE (Ovid)
MEDLINE - Academic
DatabaseTitleList MEDLINE

MEDLINE - Academic
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
EISSN 1754-7628
EndPage 160
ExternalDocumentID 24606544
10_1179_1754762814Y_0000000064
11694208
Genre Research Article
Journal Article
GroupedDBID ---
002
0R~
1OC
1~B
36B
4.4
53G
5GY
6PF
AAGDL
AALUX
AAMIU
AAPUL
AAQRR
AAWTL
ABBKH
ABDBF
ABEIZ
ABIJN
ABIVO
ABLIJ
ABUPF
ABWVI
ABXYU
ACENM
ACIEZ
ACUHS
ADCVX
AECIN
AFRVT
AGDLA
AGFJD
AGRBW
AGYJP
AIJEM
AIRBT
AKBVH
ALMA_UNASSIGNED_HOLDINGS
ALQZU
ALYBC
AMDAE
ARJSQ
BLEHA
BOHLJ
CCCUG
CS3
E01
EBD
EBS
EJD
EMB
EMOBN
ESX
F5P
H13
HCLVR
HZ~
KYCEM
LAW
M4Z
OIG
P2P
P76
P7A
P7B
QRW
RNANH
RVRKI
RYL
SV3
TASJS
TBQAZ
TERGH
TFL
TFW
TUROJ
TUS
UEQFS
XV2
29F
31~
8-1
AAJKZ
AAYXX
ACKZS
ACOPL
ACXQS
ACYZI
ADFOM
ADFZZ
AEIIZ
AFBPY
AJAOE
CITATION
CZDIS
DRXRE
DWTOO
IPNFZ
LJTGL
M46
RIG
ADYSH
CGR
CUY
CVF
ECM
EIF
NPM
7X8
ID FETCH-LOGICAL-c284t-fa9489cf36b3c6ed9b97077e3d3f0d4c9dcefeb2d2c9f90771dc672288f06f873
IEDL.DBID TFW
ISSN 1467-0100
1754-7628
IngestDate Thu Sep 04 18:44:16 EDT 2025
Mon Jul 21 06:01:52 EDT 2025
Tue Nov 18 21:59:19 EST 2025
Sat Nov 29 05:56:57 EST 2025
Mon Oct 20 23:46:51 EDT 2025
IsPeerReviewed true
IsScholarly true
Issue 3
Keywords measurable auditory percept
electrically evoked stapedial response telemetry
Cochlear implant
Impedance telemetry
most comfortable level
neural response imaging
Evoked compound action potential
electrically evoked auditory brainstem response
Clinical unit
Language English
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-c284t-fa9489cf36b3c6ed9b97077e3d3f0d4c9dcefeb2d2c9f90771dc672288f06f873
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
PMID 24606544
PQID 1530320209
PQPubID 23479
PageCount 16
ParticipantIDs proquest_miscellaneous_1530320209
crossref_citationtrail_10_1179_1754762814Y_0000000064
pubmed_primary_24606544
crossref_primary_10_1179_1754762814Y_0000000064
informaworld_taylorfrancis_310_1179_1754762814Y_0000000064
PublicationCentury 2000
PublicationDate 2014-May
PublicationDateYYYYMMDD 2014-05-01
PublicationDate_xml – month: 05
  year: 2014
  text: 2014-May
PublicationDecade 2010
PublicationPlace England
PublicationPlace_xml – name: England
PublicationTitle Cochlear implants international
PublicationTitleAlternate Cochlear Implants Int
PublicationYear 2014
Publisher Taylor & Francis
Publisher_xml – name: Taylor & Francis
References CIT0001
Shallop J.K. (CIT0011) 1995; 166
CIT0012
Di Nardo W. (CIT0006) 2003; 23
Mason S. (CIT0010) 2004; 43
Hodges A.V. (CIT0009) 2003
CIT0003
CIT0014
CIT0002
CIT0013
CIT0005
CIT0004
CIT0015
CIT0007
CIT0008
References_xml – ident: CIT0004
  doi: 10.1097/mao.0b013e3180577919
– ident: CIT0012
  doi: 10.1159/000066154
– ident: CIT0013
  doi: 10.1097/00003446-199404000-00008
– volume: 23
  start-page: 352
  year: 2003
  ident: CIT0006
  publication-title: Acta Otolaryngol Italica
– ident: CIT0001
  doi: 10.1097/00003446-200004000-00009
– start-page: S81
  volume-title: Cochlear Implants - Objective Measures
  year: 2003
  ident: CIT0009
– ident: CIT0005
  doi: 10.1016/j.heares.2008.06.011
– ident: CIT0007
  doi: 10.1097/01.aud.0000146178.84065.b3
– ident: CIT0002
  doi: 10.1097/00003446-200112000-00004
– volume: 43
  start-page: S33
  year: 2004
  ident: CIT0010
  publication-title: International Journal of Audiology
– ident: CIT0008
  doi: 10.1080/00016480510026890
– ident: CIT0003
  doi: 10.1097/00020840-200310000-00013
– volume: 166
  start-page: 175
  year: 1995
  ident: CIT0011
  publication-title: Annals of Otology, Rhinology and Laryngology Supplement
– ident: CIT0015
  doi: 10.1097/00003446-200204000-00001
– ident: CIT0014
  doi: 10.1016/S0165-5876(01)00426-8
SSID ssj0031454
Score 1.9792236
Snippet Objectives Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming...
Indications for cochlear implantation have expanded today to include very young children and those with syndromes/multiple handicaps. Programming the implant...
SourceID proquest
pubmed
crossref
informaworld
SourceType Aggregation Database
Index Database
Enrichment Source
Publisher
StartPage 145
SubjectTerms Auditory Threshold - physiology
Behavior
Child
Child, Preschool
Clinical unit
Cochlear implant
Cochlear Implantation
Cochlear Implants
Deafness - physiopathology
Deafness - therapy
electrically evoked auditory brainstem response
electrically evoked stapedial response telemetry
Electrophysiological Phenomena
Evoked compound action potential
Evoked Potentials, Auditory, Brain Stem
Female
Humans
Impedance telemetry
Male
measurable auditory percept
most comfortable level
neural response imaging
Prospective Studies
Telemetry
Title A clinical study of electrophysiological correlates of behavioural comfort levels in cochlear implantees
URI https://www.tandfonline.com/doi/abs/10.1179/1754762814Y.0000000064
https://www.ncbi.nlm.nih.gov/pubmed/24606544
https://www.proquest.com/docview/1530320209
Volume 15
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVAWR
  databaseName: Taylor & Francis
  customDbUrl:
  eissn: 1754-7628
  dateEnd: 99991231
  omitProxy: false
  ssIdentifier: ssj0031454
  issn: 1467-0100
  databaseCode: TFW
  dateStart: 20000301
  isFulltext: true
  titleUrlDefault: https://www.tandfonline.com
  providerName: Taylor & Francis
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV3LT8IwGG-UGOPF9wMfpCZeF8Za19YbMRIPygkjnpauj0ACg8Dw7_drtyEcCAe9buu29uv3-LX9vh9CD4IYHlNDgzDlAFDACQRSKB0AdiY2JSpl2kv6jXW7vN8Xq7kw7lilw9C2KBThbbVTbpmWDCRMNMHjUdBh3qJfRflB71jBDIPvd7rZ63xWxpi0qCdC8_YAoEdYJglvfs2af1qrXro5BvW-qHP0H704RodlJIrbxdQ5QTsmO0X77-Ve-xkatHGVNol9EVo8sbhkzfHrIZXZxMoxfIxc0OoeKRP_XTkPuDOGnuV45I4mzfEwgytq4Jgq8HA8HTmpmvk5-ui89J5fg5KXIVDgzPLASkG5UJbEIMzYaJEKFjJmiCY21FQJrYwFxK4jJSyAb9bSKmZRxLkNY8sZuUC1bJKZK4RDm2qICKVMAZcB-JIsNjxMI2qlalkZ19FjJY1ElUXLHXfGKPHghYlkZRiT32Gso-ay3bQo27G1xdOqsJPcL5jYgt0kIdsa31dTIwH1dHsuMjOTxTwBh-Ip6kNRR5fFnFn-UERjl9pLr__y6Rt0ADEcLc5g3qJaPluYO7SnvvPhfNZAu6zPG14lfgDACQXW
linkProvider Taylor & Francis
linkToHtml http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV3JTsMwEB2xCbiwL2U1EteINDaxzQ0hKhCl4lAEnKzEi6hUUkQD38_YSVgOiANckziJPZ5582zPDMChpFakzLIozgUSFASBKJPaRMidqcupzrkJku7yXk_c38ubCThrYmH8sUrPoV2VKCLYaq_cfjG60nAujxDyGCqxaLOHKv9gQNZJmD5GsPUJ9Pudu8Yc0zYLpdCCRUDyEddhwj-_5xtCfctf-rMXGtCos_gv_ViChdoZJafV7FmGCVuswOx1vd2-Co-npImcJCEPLRk5UhfOCUsijeUk2hf5GHq_1T9Sx_77jB545wm7VpKhP500JoMCr-hHX6yCDJ6eh16wdrwGt53z_tlFVJdmiDTiWRm5TDIhtaMpyjO1RuaSx5xbaqiLDdPSaOuQtJtES4f8m7eNTnmSCOHi1AlO12GqGBV2E0jscoNOYZblSM2Qf2U8tSLOE-Yy3XZZ2oLjRhxK13nLffmMoQr8hUv1ZRjV5zC24Oij3XOVuePXFidfpa3KsGbiqgIniv7W-KCZGwo11G-7ZIUdvY4VYkqoUh_LFmxUk-bjhxKW-uhetvWXT-_D3EX_uqu6l72rbZhHl45VRzJ3YKp8ebW7MKPfysH4ZS9oxjsn0wkY
linkToPdf http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwrV1Lb9QwEB5BQSsuPAqFhQKuxDXaJDZ-cKtoIxBl1cNWLCcr8UNdaZtdNSm_n7GTlN3Dqgd6TeIk9sx45rM98wF8UtRJzhxL0koiQEEnkJTK2ASxM_UVNZWwUdJnYjqV87nazIUJxyoDhvZdoYg4VwfjXlvfGbhQE_R4DG1YZux3V34wOtaH8AhjZx60fFb8GmZjmrHIhBYnBMQeaZ8lvPs9Ww5qq3zp7iA0OqPi2X104zk87UNRctzpzgt44Op9GP3sN9tfwuUxGfImSaxCS1ae9LQ5cUFkmDeJCRQfyxC1hkf6zP9QzwPvXGHPWrIMZ5MasqjxirkMVBVkcbVeBrG65hVcFKezr9-SnpghMejN2sSXikllPOUoTe6sqpRIhXDUUp9aZpQ1ziNkt7lRHtG3yKzhIs-l9Cn3UtAD2KtXtXsDJPWVxZCwLCsEZoi-SsGdTKuc-dJkvuRj-DxIQ5u-ankgz1jqiF6E0hvDqP8N4xgmt-3WXd2OO1t82RS2buOKie_oTTS9q_HRoBoa7TNsupS1W900Gj1K5KhP1Rhedzpz-0M54yG3l739n09_hNH5SaHPvk9_vIMnGM-x7jzmIey11zfuPTw2f9pFc_0h2sVf-wMHyg
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=A+clinical+study+of+electrophysiological+correlates+of+behavioural+comfort+levels+in+cochlear+implantees&rft.jtitle=Cochlear+implants+international&rft.au=Raghunandhan%2C+S&rft.au=Ravikumar%2C+A&rft.au=Kameswaran%2C+Mohan&rft.au=Mandke%2C+Kalyani&rft.date=2014-05-01&rft.pub=Taylor+%26+Francis&rft.issn=1467-0100&rft.eissn=1754-7628&rft.volume=15&rft.issue=3&rft.spage=145&rft.epage=160&rft_id=info:doi/10.1179%2F1754762814Y.0000000064&rft.externalDocID=11694208
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1467-0100&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1467-0100&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1467-0100&client=summon