FEASIBILITY OF IDENTIFY (TM) OPTICAL SURFACE GUIDANCE SYSTEM FOR CRANIOSPINAL IRRADIATION

Purpose:Craniospinal irradiation (CSI) at our centre is delivered using a multi-isocentric technique. For CSI, patient alignment at each isocentre is verified using orthogonal kV imaging to ensure adherence to procedural tolerances. A retrospective review of the first three fractions from 20 adult C...

Celý popis

Uloženo v:
Podrobná bibliografie
Vydáno v:Radiotherapy and oncology Ročník 210; s. S32
Hlavní autoři: MacLennan, Ranslee, Ploquin, Nicolas, Murtha, Nathan
Médium: Journal Article
Jazyk:angličtina
Vydáno: 01.09.2025
Témata:
ISSN:0167-8140
On-line přístup:Získat plný text
Tagy: Přidat tag
Žádné tagy, Buďte první, kdo vytvoří štítek k tomuto záznamu!
Abstract Purpose:Craniospinal irradiation (CSI) at our centre is delivered using a multi-isocentric technique. For CSI, patient alignment at each isocentre is verified using orthogonal kV imaging to ensure adherence to procedural tolerances. A retrospective review of the first three fractions from 20 adult CSI treatments between 2019 and 2024 was conducted to assess the time allocation for patient alignment and treatment. On average, kV imaging accounted for 34% of the total treatment time, representing a substantial portion of the overall patient appointment duration. This study seeks to determine whether the Varian IDENTIFY TM surface guidance system can be used in place of kV orthogonal imaging for the initial alignment of CSI patients. We propose that a surface guidance system can improve CSI patient experience by reducing the alignment time while maintaining or improving alignment accuracy. Materials and Methods:The IDENTIFY TM system uses three cameras to generate a real-time surface contour of the patient. This real-time contour is continuously compared to a predefined reference contour, enabling the system to detect and quantify translational and rotational deviations. By providing precise positional data, IDENTIFY TM enables accurate patient alignment. Unique regions of interest (ROIs) are defined on the body contour at each isocentre. A prospective study was conducted using an anthropomorphic phantom to compare positioning accuracy between the IDENTIFY TM system and kV imaging. A CT scan was acquired of the phantom immobilized with an open-face five-point shell. Next, a surface contour of the phantom was generated, and a treatment plan consisting of three isocentres was created. ROIs were defined at each isocentre, selecting areas with stable, prominent anatomical features to minimize susceptibility to rapid contour changes in a clinical setting. A series of measurements were performed for each isocentre in which the phantom was initially aligned based on IDENTIFY TM-derived positioning data, followed by orthogonal kV imaging for setup verification. Results:Across all isocentres, the largest difference between IDENTIFY TM and kV imaging registrations was 0.07cm in translational shifts and 0.4 degrees in rotational adjustments. The positioning data from the phantom study shows a high level of agreement between IDENTIFY TM and kV imaging, remaining well within institutional tolerances for each isocentre. Conclusions:This study suggests that the current patient alignment procedure based on kV imaging can be substituted with the IDENTIFY TM system without compromising accuracy.
AbstractList Purpose:Craniospinal irradiation (CSI) at our centre is delivered using a multi-isocentric technique. For CSI, patient alignment at each isocentre is verified using orthogonal kV imaging to ensure adherence to procedural tolerances. A retrospective review of the first three fractions from 20 adult CSI treatments between 2019 and 2024 was conducted to assess the time allocation for patient alignment and treatment. On average, kV imaging accounted for 34% of the total treatment time, representing a substantial portion of the overall patient appointment duration. This study seeks to determine whether the Varian IDENTIFY TM surface guidance system can be used in place of kV orthogonal imaging for the initial alignment of CSI patients. We propose that a surface guidance system can improve CSI patient experience by reducing the alignment time while maintaining or improving alignment accuracy. Materials and Methods:The IDENTIFY TM system uses three cameras to generate a real-time surface contour of the patient. This real-time contour is continuously compared to a predefined reference contour, enabling the system to detect and quantify translational and rotational deviations. By providing precise positional data, IDENTIFY TM enables accurate patient alignment. Unique regions of interest (ROIs) are defined on the body contour at each isocentre. A prospective study was conducted using an anthropomorphic phantom to compare positioning accuracy between the IDENTIFY TM system and kV imaging. A CT scan was acquired of the phantom immobilized with an open-face five-point shell. Next, a surface contour of the phantom was generated, and a treatment plan consisting of three isocentres was created. ROIs were defined at each isocentre, selecting areas with stable, prominent anatomical features to minimize susceptibility to rapid contour changes in a clinical setting. A series of measurements were performed for each isocentre in which the phantom was initially aligned based on IDENTIFY TM-derived positioning data, followed by orthogonal kV imaging for setup verification. Results:Across all isocentres, the largest difference between IDENTIFY TM and kV imaging registrations was 0.07cm in translational shifts and 0.4 degrees in rotational adjustments. The positioning data from the phantom study shows a high level of agreement between IDENTIFY TM and kV imaging, remaining well within institutional tolerances for each isocentre. Conclusions:This study suggests that the current patient alignment procedure based on kV imaging can be substituted with the IDENTIFY TM system without compromising accuracy.
Author Murtha, Nathan
Ploquin, Nicolas
MacLennan, Ranslee
Author_xml – sequence: 1
  fullname: MacLennan, Ranslee
– sequence: 2
  fullname: Ploquin, Nicolas
– sequence: 3
  fullname: Murtha, Nathan
BookMark eNo9kEFrg0AQhfeQQpO0P6Gwx-RguzvqGi8FazRdSDSoOXhaVt0F09QUpYX8-65J6WVmGOZ9vHkzNOnOnULoiZJnSih7yU3xrBV1yALcJXE8GyyYoOn_-h7NhuFICAFie1NUxlGQ8ze-5UWJ0xjzdZQUPC7xotgtcboveBhscX7I4iCM8ObA10FihrzMi2iH4zTDYRYkPM33PDGHPMuCNQ8KniYP6E7L06Ae__ocHeKoCN-tbboZoVZtDIMlXdUwLZmnGtDQ2DZroPapXvlU-Y5DtU-aqmKeZlq7ZCWlairl21qCrhjUxJ4j98at-_Mw9EqLr779lP1FUCLGTMQ1EzE-L8AV10wEGN3rTaeMuZ9W9aI-tV1by9OHuqjheP7uO-NbUDGAIDfIyAD3SgD7F5rOZkA
ContentType Journal Article
Copyright Elsevier Ireland Ltd
Copyright_xml – notice: Elsevier Ireland Ltd
DBID AAYXX
CITATION
DOI 10.1016/S0167-8140(25)04732-2
DatabaseName CrossRef
DatabaseTitle CrossRef
DatabaseTitleList
DeliveryMethod fulltext_linktorsrc
Discipline Medicine
EndPage S32
ExternalDocumentID 10_1016_S0167_8140_25_04732_2
1_s2_0_S0167814025047322
GroupedDBID ---
--K
--M
.1-
.55
.FO
.GJ
.~1
0R~
123
1B1
1P~
1RT
1~.
1~5
29P
4.4
457
4G.
53G
5RE
5VS
7-5
71M
8P~
9JM
AABNK
AAEDT
AAEDW
AAIKJ
AAKOC
AALRI
AAOAW
AAQFI
AAQXK
AATTM
AAXKI
AAXUO
AAYWO
ABBQC
ABFNM
ABJNI
ABMAC
ABMZM
ABOCM
ABUDA
ABWVN
ABXDB
ACDAQ
ACGFS
ACIEU
ACIUM
ACLOT
ACRLP
ACRPL
ACVFH
ADBBV
ADCNI
ADEZE
ADMUD
ADNMO
ADVLN
AEBSH
AEHWI
AEIPS
AEKER
AENEX
AEUPX
AEVXI
AFJKZ
AFPUW
AFRHN
AFTJW
AFXIZ
AGHFR
AGQPQ
AGUBO
AGYEJ
AHHHB
AIEXJ
AIGII
AIIUN
AIKHN
AITUG
AJRQY
AJUYK
AKBMS
AKRWK
AKYEP
ALMA_UNASSIGNED_HOLDINGS
AMRAJ
ANKPU
ANZVX
APXCP
ASPBG
AVWKF
AXJTR
AZFZN
BKOJK
BLXMC
BNPGV
CS3
DU5
E3Z
EBS
EFJIC
EFKBS
EFLBG
EJD
EO8
EO9
EP2
EP3
F5P
FDB
FEDTE
FGOYB
FIRID
FNPLU
FYGXN
G-2
G-Q
GBLVA
HED
HMK
HMO
HVGLF
HZ~
IHE
J1W
J5H
KOM
M27
M41
MO0
N9A
O-L
O9-
OAUVE
OC~
OO-
OZT
P-8
P-9
P2P
PC.
Q38
R2-
ROL
RPZ
SAE
SDF
SDG
SEL
SES
SEW
SPCBC
SSH
SSU
SSZ
T5K
UV1
WUQ
X7M
Z5R
ZGI
ZXP
~G-
~HD
9DU
AAYXX
CITATION
ID FETCH-LOGICAL-c1012-a5ed6fa67ed2f2d336d2c91f891e9441f90dbb67f6ff508aaedbe93fa2fb62c03
ISSN 0167-8140
IngestDate Sat Nov 29 07:26:17 EST 2025
Sun Oct 19 01:22:09 EDT 2025
IsPeerReviewed true
IsScholarly true
Language English
LinkModel OpenURL
MergedId FETCHMERGED-LOGICAL-c1012-a5ed6fa67ed2f2d336d2c91f891e9441f90dbb67f6ff508aaedbe93fa2fb62c03
ParticipantIDs crossref_primary_10_1016_S0167_8140_25_04732_2
elsevier_clinicalkeyesjournals_1_s2_0_S0167814025047322
PublicationCentury 2000
PublicationDate 2025-09-00
PublicationDateYYYYMMDD 2025-09-01
PublicationDate_xml – month: 09
  year: 2025
  text: 2025-09-00
PublicationDecade 2020
PublicationTitle Radiotherapy and oncology
PublicationYear 2025
SSID ssj0002037
Score 2.4714704
Snippet Purpose:Craniospinal irradiation (CSI) at our centre is delivered using a multi-isocentric technique. For CSI, patient alignment at each isocentre is verified...
SourceID crossref
elsevier
SourceType Index Database
Publisher
StartPage S32
SubjectTerms Hematology, Oncology, and Palliative Medicine
Title FEASIBILITY OF IDENTIFY (TM) OPTICAL SURFACE GUIDANCE SYSTEM FOR CRANIOSPINAL IRRADIATION
URI https://www.clinicalkey.es/playcontent/1-s2.0-S0167814025047322
Volume 210
hasFullText 1
inHoldings 1
isFullTextHit
isPrint
journalDatabaseRights – providerCode: PRVESC
  databaseName: Elsevier SD Freedom Collection Journals 2021
  issn: 0167-8140
  databaseCode: AIEXJ
  dateStart: 20201001
  customDbUrl:
  isFulltext: true
  dateEnd: 99991231
  titleUrlDefault: https://www.sciencedirect.com
  omitProxy: false
  ssIdentifier: ssj0002037
  providerName: Elsevier
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwtV3db5swELemdpr2Mu1Ta_chP2xSq4gNDBh4ZAlsTAuJApGSJ2SwkSpVtCpp1T9_Z2wI3aZpfdgLilBskO_nO99x9zuEPljUZa7DSsPijBiSUM5gIGrweTwqqFlVts27ZhNemvqbTbDUUaW2ayfgNY1_extc_ldRwz0QtiydvYe4h0nhBvwGocMVxA7XfxJ8HIVZ8iX5keRbmdKTzKI0T2LJm-3ncxkEWCxzRYKwXsXhNJp8XScz2aJmkm2zPJpPwC-cTFdhmiwy8Pnhj8lqFc6SfTCr5_Nm_EzXbykSp4umuhOk7_KKG90AeQU28VwMMFqeg0VS_AUdGFk7Ev1OfYVKu7j-OC5B3CHxaghVggqWfFpjXUt0DqvSlpkKbf6mxVVAIRtm-CgLbGGBTMezia6WvMOc_YtFG_IMRyls1CvkVAVxi26aAuz2IQFsgio8DJNo830w4MRUVKv94_eFX5_373RC3FP9Pn8-0oyOKflT9ET7FzhUuHiGHojmOXo01xkUL9B2BA-8iHEPD3ySz0-xhgbW0MA9NLCCBgZo4DE08AgaL9E6jvLpN0O31zAqSeoGe1JwWjPqCU5qwm2bclIFVu0HlgjglFwHJi9L6tW0ruEYz5jgpQjsmpG6pKQy7VfooLloxGuEJaum4zvgHpemY1fMr6oANrlTOjwIasaP0Kd-hYpLxaJS_FU2R8jr17HoS4TBqIlWb7S2sIqWFKYaLQdLKj4YSo7v-6g36LGCLzUs-hYd7K6uxTv0sLrZnbVX7zU2fgJ1Lmol
linkProvider Elsevier
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=FEASIBILITY+OF+IDENTIFY+%28TM%29+OPTICAL+SURFACE+GUIDANCE+SYSTEM+FOR+CRANIOSPINAL+IRRADIATION&rft.jtitle=Radiotherapy+and+oncology&rft.au=MacLennan%2C+Ranslee&rft.au=Ploquin%2C+Nicolas&rft.au=Murtha%2C+Nathan&rft.date=2025-09-01&rft.issn=0167-8140&rft.volume=210&rft.spage=S32&rft_id=info:doi/10.1016%2FS0167-8140%2825%2904732-2&rft.externalDBID=n%2Fa&rft.externalDocID=10_1016_S0167_8140_25_04732_2
thumbnail_m http://cvtisr.summon.serialssolutions.com/2.0.0/image/custom?url=https%3A%2F%2Fcdn.clinicalkey.com%2Fck-thumbnails%2F01678140%2FS0167814025X00104%2Fcov150h.gif