Study on the Long Bone Failure Behaviors Under the Indenter Rigid-Contact by Experiment Analysis and Subject-Specific Simulation

The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and falling incidents, etc. For evaluating bone bending behaviors, most of the previous studies used an indenter in three point bending experiments w...

Celý popis

Uložené v:
Podrobná bibliografia
Vydané v:Journal of biomechanical engineering Ročník 143; číslo 2
Hlavní autori: Du, Xianping, Jiang, Binhui, Zhang, Guanjun, Chou, Clifford C, Bai, Zhonghao
Médium: Journal Article
Jazyk:English
Vydavateľské údaje: United States 01.02.2021
Predmet:
ISSN:1528-8951, 1528-8951
On-line prístup:Zistit podrobnosti o prístupe
Tagy: Pridať tag
Žiadne tagy, Buďte prvý, kto otaguje tento záznam!
Abstract The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and falling incidents, etc. For evaluating bone bending behaviors, most of the previous studies used an indenter in three point bending experiments while the effect of its rigidity was never considered. In this work, using the porcine long bones, the three point bending tests were conducted to explore the bone fracture behaviors under a rigid indenter. In addition to collecting the force applied, the bone fracture dynamic process was recorded by high-speed photography, and the fracture surface profile in mesoscale was observed by the scanning electron microscope (SEM). Based on CT scanning of long bones, the cross section properties of test specimens were calculated by a homemade matlab script for correlating with their failure strengths. Also, a subject-specific finite element (FE) model was developed to identify the outcomes induced by a rigid indenter on simulation. Findings led to conclusions as follows: (1) The tension fracture came with fracture path deflection, which was caused by the bone indentation induced mesoscale crack-opening. Due to this damage before the whole bone fracture, a bone fracture moment correction was established to compensate experimental data. (2) The plastic indentation caused the force fluctuation as suggested by correlation analysis. (3) The bone failure moment correlated with the inertial moment of the bone cross section at the fracture location higher than the traditional cross section area. (4) In the subject-specific simulation, the indentation caused compression fracture under a much lower failure force. Removing the element erosion on the indenter-contacted area only during the validation was verified as a good option to solve this issue.
AbstractList The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and falling incidents, etc. For evaluating bone bending behaviors, most of the previous studies used an indenter in three point bending experiments while the effect of its rigidity was never considered. In this work, using the porcine long bones, the three point bending tests were conducted to explore the bone fracture behaviors under a rigid indenter. In addition to collecting the force applied, the bone fracture dynamic process was recorded by high-speed photography, and the fracture surface profile in mesoscale was observed by the scanning electron microscope (SEM). Based on CT scanning of long bones, the cross section properties of test specimens were calculated by a homemade matlab script for correlating with their failure strengths. Also, a subject-specific finite element (FE) model was developed to identify the outcomes induced by a rigid indenter on simulation. Findings led to conclusions as follows: (1) The tension fracture came with fracture path deflection, which was caused by the bone indentation induced mesoscale crack-opening. Due to this damage before the whole bone fracture, a bone fracture moment correction was established to compensate experimental data. (2) The plastic indentation caused the force fluctuation as suggested by correlation analysis. (3) The bone failure moment correlated with the inertial moment of the bone cross section at the fracture location higher than the traditional cross section area. (4) In the subject-specific simulation, the indentation caused compression fracture under a much lower failure force. Removing the element erosion on the indenter-contacted area only during the validation was verified as a good option to solve this issue.
The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and falling incidents, etc. For evaluating bone bending behaviors, most of the previous studies used an indenter in three point bending experiments while the effect of its rigidity was never considered. In this work, using the porcine long bones, the three point bending tests were conducted to explore the bone fracture behaviors under a rigid indenter. In addition to collecting the force applied, the bone fracture dynamic process was recorded by high-speed photography, and the fracture surface profile in mesoscale was observed by the scanning electron microscope (SEM). Based on CT scanning of long bones, the cross section properties of test specimens were calculated by a homemade matlab script for correlating with their failure strengths. Also, a subject-specific finite element (FE) model was developed to identify the outcomes induced by a rigid indenter on simulation. Findings led to conclusions as follows: (1) The tension fracture came with fracture path deflection, which was caused by the bone indentation induced mesoscale crack-opening. Due to this damage before the whole bone fracture, a bone fracture moment correction was established to compensate experimental data. (2) The plastic indentation caused the force fluctuation as suggested by correlation analysis. (3) The bone failure moment correlated with the inertial moment of the bone cross section at the fracture location higher than the traditional cross section area. (4) In the subject-specific simulation, the indentation caused compression fracture under a much lower failure force. Removing the element erosion on the indenter-contacted area only during the validation was verified as a good option to solve this issue.The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and falling incidents, etc. For evaluating bone bending behaviors, most of the previous studies used an indenter in three point bending experiments while the effect of its rigidity was never considered. In this work, using the porcine long bones, the three point bending tests were conducted to explore the bone fracture behaviors under a rigid indenter. In addition to collecting the force applied, the bone fracture dynamic process was recorded by high-speed photography, and the fracture surface profile in mesoscale was observed by the scanning electron microscope (SEM). Based on CT scanning of long bones, the cross section properties of test specimens were calculated by a homemade matlab script for correlating with their failure strengths. Also, a subject-specific finite element (FE) model was developed to identify the outcomes induced by a rigid indenter on simulation. Findings led to conclusions as follows: (1) The tension fracture came with fracture path deflection, which was caused by the bone indentation induced mesoscale crack-opening. Due to this damage before the whole bone fracture, a bone fracture moment correction was established to compensate experimental data. (2) The plastic indentation caused the force fluctuation as suggested by correlation analysis. (3) The bone failure moment correlated with the inertial moment of the bone cross section at the fracture location higher than the traditional cross section area. (4) In the subject-specific simulation, the indentation caused compression fracture under a much lower failure force. Removing the element erosion on the indenter-contacted area only during the validation was verified as a good option to solve this issue.
Author Zhang, Guanjun
Chou, Clifford C
Du, Xianping
Jiang, Binhui
Bai, Zhonghao
Author_xml – sequence: 1
  givenname: Xianping
  surname: Du
  fullname: Du, Xianping
  organization: The State Key Laboratory of Advanced Design and Manufacturing for Vehicle Body, Hunan University, Changsha, Hunan 410082, China; Department of Mechanical and Aerospace Engineering, Rutgers University, Piscataway, NJ 08854
– sequence: 2
  givenname: Binhui
  surname: Jiang
  fullname: Jiang, Binhui
  organization: The State Key Laboratory of Advanced Design and Manufacturing for Vehicle Body, Hunan University, 2 Lushan S Rd., Yuelu District, Changsha, Hunan 410082, China
– sequence: 3
  givenname: Guanjun
  surname: Zhang
  fullname: Zhang, Guanjun
  organization: The State Key Laboratory of Advanced Design and Manufacturing for Vehicle Body, Hunan University, Changsha, Hunan 410082, China
– sequence: 4
  givenname: Clifford C
  surname: Chou
  fullname: Chou, Clifford C
  organization: Bioengineering Center, Wayne State University, Detroit, MI 48201
– sequence: 5
  givenname: Zhonghao
  surname: Bai
  fullname: Bai, Zhonghao
  organization: The State Key Laboratory of Advanced Design and Manufacturing for Vehicle Body, Hunan University, Changsha, Hunan 410082, China
BackLink https://www.ncbi.nlm.nih.gov/pubmed/32839823$$D View this record in MEDLINE/PubMed
BookMark eNpNkEtLw0AUhQep2Icu_AMySzep80gmybItrQoFwdh1mGRu2inJTM1MxOz86Qat4Oqcw_04cM8UjYw1gNAtJXNKafRA5yEJE0b4BZrQiCVBkkZ09M-P0dS5IyGUJiG5QmPOEp4mjE_QV-Y71WNrsD8A3lqzx8uhHG-krrsW8BIO8kPb1uGdUdD-UM-DM34Ir3qvVbCyxsvS46LH688TtLoZrnhhZN077bA0CmddcYTSB9kJSl3pEme66WrptTXX6LKStYObs87QbrN-Wz0F25fH59ViG0hOhA-EIkrwStKYQRpTAYVQaViEUnAouFJURVKAYqmICMSV4KJK0pIWPI6IFKVkM3T_23tq7XsHzueNdiXUtTRgO5ezkMeUkYjzAb07o13RgMpPw0uy7fO_1dg34ZxwQA
CitedBy_id crossref_primary_10_1016_j_taml_2024_100547
crossref_primary_10_1080_13588265_2023_2272818
ContentType Journal Article
Copyright Copyright © 2021 by ASME.
Copyright_xml – notice: Copyright © 2021 by ASME.
DBID NPM
7X8
DOI 10.1115/1.4048203
DatabaseName PubMed
MEDLINE - Academic
DatabaseTitle PubMed
MEDLINE - Academic
DatabaseTitleList PubMed
MEDLINE - Academic
Database_xml – sequence: 1
  dbid: NPM
  name: PubMed
  url: http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed
  sourceTypes: Index Database
– sequence: 2
  dbid: 7X8
  name: MEDLINE - Academic
  url: https://search.proquest.com/medline
  sourceTypes: Aggregation Database
DeliveryMethod no_fulltext_linktorsrc
Discipline Medicine
Engineering
Forestry
EISSN 1528-8951
ExternalDocumentID 32839823
Genre Research Support, Non-U.S. Gov't
Journal Article
GroupedDBID ---
-~X
.DC
29J
4.4
5AI
5GY
ABJNI
ACBEA
ACGFO
ACGFS
ACKMT
ADPDT
AGNGV
ALMA_UNASSIGNED_HOLDINGS
CS3
EBS
F5P
H~9
L7B
NPM
P2P
RAI
RNS
RXW
TAE
TN5
UKR
7X8
ID FETCH-LOGICAL-a306t-6d0d63fa172e9716eb6d94b4a63eb3dd1d5a6ed29650e7f636f89c1b3750a6ca2
IEDL.DBID 7X8
ISICitedReferencesCount 2
ISICitedReferencesURI http://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=Summon&SrcAuth=ProQuest&DestLinkType=CitingArticles&DestApp=WOS_CPL&KeyUT=000603631300003&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
ISSN 1528-8951
IngestDate Fri Jul 11 10:59:07 EDT 2025
Mon Jul 21 06:03:04 EDT 2025
IsPeerReviewed true
IsScholarly true
Issue 2
Keywords delamination
subject-specific modeling
indentation
long bone
three point bending
bone fracture
Language English
License Copyright © 2021 by ASME.
LinkModel DirectLink
MergedId FETCHMERGED-LOGICAL-a306t-6d0d63fa172e9716eb6d94b4a63eb3dd1d5a6ed29650e7f636f89c1b3750a6ca2
Notes ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
PMID 32839823
PQID 2437120533
PQPubID 23479
ParticipantIDs proquest_miscellaneous_2437120533
pubmed_primary_32839823
PublicationCentury 2000
PublicationDate 2021-Feb-01
20210201
PublicationDateYYYYMMDD 2021-02-01
PublicationDate_xml – month: 02
  year: 2021
  text: 2021-Feb-01
  day: 01
PublicationDecade 2020
PublicationPlace United States
PublicationPlace_xml – name: United States
PublicationTitle Journal of biomechanical engineering
PublicationTitleAlternate J Biomech Eng
PublicationYear 2021
SSID ssj0011840
Score 2.325465
Snippet The bending fracture behaviors of long bone have gained great attention due to the high bending fracture risk during sports events, traffic accidents, and...
SourceID proquest
pubmed
SourceType Aggregation Database
Index Database
Title Study on the Long Bone Failure Behaviors Under the Indenter Rigid-Contact by Experiment Analysis and Subject-Specific Simulation
URI https://www.ncbi.nlm.nih.gov/pubmed/32839823
https://www.proquest.com/docview/2437120533
Volume 143
WOSCitedRecordID wos000603631300003&url=https%3A%2F%2Fcvtisr.summon.serialssolutions.com%2F%23%21%2Fsearch%3Fho%3Df%26include.ft.matches%3Dt%26l%3Dnull%26q%3D
hasFullText
inHoldings 1
isFullTextHit
isPrint
link http://cvtisr.summon.serialssolutions.com/2.0.0/link/0/eLvHCXMwpV1LS8QwEA7qiujB92N9EcFrdJt00_YkKi4KuogP2NuS5rH0YKt2FfbmT3cmm1UvguAl9NBCm0yn32S-fh8hh7EzgmsnWGxiC4NQLHfSMHS4kqlLklx5ndnrpNtNe73sNmy41YFWOcmJPlGbSuMe-TEK50Uc_xw9eX5h6BqF3dVgoTFNGgKgDFK6kt53FwGrF6-XylOWApQIykIAgo6joxhil3uvrF-Qpf_CdJb-e2_LZDFgS3o6DoYVMmXLVbLwQ3FwlcyhFSf6u8HhTWirr5EPZBOOaFVSgIP0uioH9KwqLe2oAlnrNIgovtbU2yT5s67QOhcWhd4Vg8IwVLlSekjzEb34Mg2gE8kTqkpDIUfhpg_zlveu0PS-eArmYevksXPxcH7JgjUDU1BjDJk0LSOFUwB_LIpQ2VyaLM5jJQVU58ZEpq2kNTwDAGgTJ4V0aaajXABAUVIrvkFmSniQLUKVS1vC6chalcZatnMhIwNASXMD6F_FTXIwmfQ-hD72M1Rpq7e6_z3tTbI5Xrn-81ijoy8ANmUpF9t_uHqHzHNkqngu9i5pOHjx7R6Z1e_Don7d9zEFY_f25hN2gNe7
linkProvider ProQuest
openUrl ctx_ver=Z39.88-2004&ctx_enc=info%3Aofi%2Fenc%3AUTF-8&rfr_id=info%3Asid%2Fsummon.serialssolutions.com&rft_val_fmt=info%3Aofi%2Ffmt%3Akev%3Amtx%3Ajournal&rft.genre=article&rft.atitle=Study+on+the+Long+Bone+Failure+Behaviors+Under+the+Indenter+Rigid-Contact+by+Experiment+Analysis+and+Subject-Specific+Simulation&rft.jtitle=Journal+of+biomechanical+engineering&rft.au=Du%2C+Xianping&rft.au=Jiang%2C+Binhui&rft.au=Zhang%2C+Guanjun&rft.au=Chou%2C+Clifford+C&rft.date=2021-02-01&rft.issn=1528-8951&rft.eissn=1528-8951&rft.volume=143&rft.issue=2&rft_id=info:doi/10.1115%2F1.4048203&rft.externalDBID=NO_FULL_TEXT
thumbnail_l http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/lc.gif&issn=1528-8951&client=summon
thumbnail_m http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/mc.gif&issn=1528-8951&client=summon
thumbnail_s http://covers-cdn.summon.serialssolutions.com/index.aspx?isbn=/sc.gif&issn=1528-8951&client=summon