Design and Performance Evaluation of SiPM-based High-resolution Dedicated Brain Positron Emission Tomography Scanner: A Simulation Study. |
| Autoři: |
Zare, Tahereh, Sheikhzadeh, Peyman, Teimourian Fard, Behnoosh, Ghafarian, Pardis, Ay, Mohammad Reza |
| Zdroj: |
Journal of Medical Physics; Oct-Dec2024, Vol. 49 Issue 4, p631-641, 11p |
| Témata: |
POSITRON emission tomography, MONTE Carlo method, BRAIN diseases, SCANNING systems, BASIC needs |
| Abstrakt: |
Purpose/Aim: The increasing population age highlights the critical need for early brain disease diagnosis, especially in disorders such as dementia. Consequently, a notable focus has been on developing dedicated brain positron emission tomography (PET) scanners, which offer higher resolution and sensitivity than whole-body PET scanners. This study aims to design and performance evaluation of an LYSO-based dedicated brain PET scanner. Materials and Methods: We developed a dedicated brain PET using Monte Carlo simulation based on cylindrical geometry. Each detector block consisted of a 23 × 23 array of 2 mm × 2 mm × 15 mm LYSO crystals coupled with SiPM. The performance of this scanner was evaluated based on the NEMA NU-2-2018 standard, focusing on analyzing various energy windows and coincidence time windows (CTWs). Results: The results demonstrated that the noise equivalent count rate (NECR) peaked at each CTW in the 408–680 keV energy window. In addition, increasing the CTWs from 3 ns to 10 ns resulted in a decrease of 9% in sensitivity and an increase of 63% in NECR. Furthermore, the study findings highlight that using a time-of-flight (TOF) resolution of 250 ps can substantially improve image contrast relative to non-TOF reconstruction. Conclusions: We conclude that employing a broader energy window and a narrower CTW can significantly enhance the scanner's performance regarding sensitivity and NECR. Furthermore, incorporating LYSO pixelated crystals with TOF information will facilitate the generation of high-resolution and high-contrast images. [ABSTRACT FROM AUTHOR] |
|
Copyright of Journal of Medical Physics is the property of Wolters Kluwer India Pvt Ltd and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
| Databáze: |
Complementary Index |